Accuracy of toric intraocular lens implantation in cataract and refractive surgery.

Nienke Visser1, Tos T J M Berendschot, Noël J C Bauer, Jessica Jurich, Oliver Kersting, Rudy M M A Nuijts.   

Abstract

PURPOSE: To determine the accuracy of a commonly used 3-step procedure for toric pseudophakic and phakic intraocular lens (IOL) implantation.
SETTING: University Eye Clinic, Maastricht University Medical Centre, Maastricht, The Netherlands.
DESIGN: Cohort study.
METHOD: In this analysis of toric IOL implantation, 6 preoperative images of the eye per patient and the surgery video were obtained using a digital imaging system. All 3 steps for toric IOL implantation were analyzed as follows: reference axis marking, alignment axis marking, and IOL alignment. In addition, vector analysis was used to calculate the errors in toric IOL alignment.
RESULTS: Forty eyes (26 pseudophakic, 14 phakic) were analyzed. The mean errors in reference axis marking, alignment axis marking, and toric IOL alignment were 2.4 degrees ± 0.8 (SD), 3.3 ± 2.0 degrees, and 2.6 ± 2.6 degrees, respectively. Together, these 3 errors led to a mean total error in toric IOL alignment of 4.9 ± 2.1 degrees. Subgroup analysis showed no significant difference in mean error between pseudophakic IOL and phakic toric IOL alignment (P=.501). Vector analysis showed a mean angle or error of -2 ± 8 degrees (pseudophakic IOLs) and 6 ± 14 degrees (phakic IOLs).
CONCLUSIONS: A commonly used 3-step ink-marker procedure to implant toric IOLs led to a mean error in IOL placement of approximately 5 degrees. The error was especially relevant in cases in which higher cylinder power IOLs were implanted. Orienting the toric IOL with great accuracy is necessary in all patients to achieve the best cylinder correction.
Copyright © 2011. Published by Elsevier Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21782085     DOI: 10.1016/j.jcrs.2011.02.024

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  21 in total

1.  The effect of head tilt on keratometric measurement using the IOLMaster.

Authors:  C Y Park; J R Do; S H Kim; C-Y Lim; R S Chuck
Journal:  Eye (Lond)       Date:  2013-09-20       Impact factor: 3.775

2.  Comparison of two different methods of preoperative marking for toric intraocular lens implantation: bubble marker versus pendulum marker.

Authors:  Javed Hussain Farooqui; Archana Koul; Ranjan Dutta; Noshir Minoo Shroff
Journal:  Int J Ophthalmol       Date:  2016-05-18       Impact factor: 1.779

3.  Comparison of clinical outcomes among 3 marking methods for toric intraocular lens implantation.

Authors:  Hideyuki Onishi; Hidemasa Torii; Kazuhiro Watanabe; Kazuo Tsubota; Kazuno Negishi
Journal:  Jpn J Ophthalmol       Date:  2016-03-09       Impact factor: 2.447

4.  Toric IOL positioning with a no-touch head-up display axis alignment.

Authors:  J Luebke; D Boehringer; P Maier; T Reinhard; P Eberwein
Journal:  Int Ophthalmol       Date:  2019-11-23       Impact factor: 2.031

5.  Toric markers-assisted implantation of the scleral-fixated intraocular lens.

Authors:  Hu-Ping Song; Bing-Yu Tian; Jing Peng
Journal:  Int J Ophthalmol       Date:  2016-09-18       Impact factor: 1.779

6.  Evaluation of axis alignment and refractive results of toric phakic IOL using image-guided system.

Authors:  Sherif Emerah
Journal:  Int J Ophthalmol       Date:  2020-04-18       Impact factor: 1.779

7.  Introduction of a Toric Intraocular Lens to a Non-Refractive Cataract Practice: Challenges and Outcomes.

Authors:  Clare Kirwan; John M Nolan; Jim Stack; Ian Dooley; Johnny Moore; Tara Cb Moore; Stephen Beatty
Journal:  Int J Ophthalmol Clin Res       Date:  2016-05-25

8.  Astigmatism management in cataract surgery with Precizon(®) toric intraocular lens: a prospective study.

Authors:  Carolina Vale; Carlos Menezes; J Firmino-Machado; Pedro Rodrigues; Miguel Lume; Paula Tenedório; Pedro Menéres; Maria do Céu Brochado
Journal:  Clin Ophthalmol       Date:  2016-01-19

9.  Investigation of a real-time location system of corneal astigmatic axis.

Authors:  Jian-Guo Zhao; An-Peng Pan; Ke Zheng; A-Yong Yu
Journal:  Eye Vis (Lond)       Date:  2017-09-14

10.  Comparison Between Digital and Manual Marking for Toric Intraocular Lenses: A Randomized Trial.

Authors:  Abdel Hamid Elhofi; Hany Ahmed Helaly
Journal:  Medicine (Baltimore)       Date:  2015-09       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.