Literature DB >> 21769166

Polymer-grafted QCM chemical sensor and application to heavy metalions real time detection.

Luciana Sartore1, Marzia Barbaglio, Laura Borgese, Elza Bontempi.   

Abstract

A flow type quartz crystal microbalance (QCM) chemical sensor was developed for monitoring of heavy metal ions in aqueous solutions (that is suitable for environmental monitoring). The sensor is based upon surface chelation of the metal ions at multifunctional polymer modified gold electrodes on 9 MHz AT-cut quartz resonators, functioning as a QCM. New processes have been developed which enable to obtain surface-modified gold electrodes with high heavy metal ions complexing ability. These polymer grafted QCM sensors can selectively adsorb heavy metal ions, such as copper lead chrome and cadmium, from solution over a wide range from 0.01 to 1000 ppm concentration by complexation with functional groups in the polymers. Cations typically present in natural water did not interfere with the detection of heavy metals. X-Ray Reflectivity (XRR) and Total Reflection X-ray Fluorescence (TXRF) were carried out to characterise the unmodified and modified gold surfaces as well as to verify the possibility to selectively bond and remove metal ions.

Entities:  

Year:  2011        PMID: 21769166      PMCID: PMC3137375          DOI: 10.1016/j.snb.2011.01.003

Source DB:  PubMed          Journal:  Sens Actuators B Chem        ISSN: 0925-4005            Impact factor:   7.460


  8 in total

1.  Use of total reflection X-ray fluorescence (TXRF) for the evaluation of heavy metal poisoning due to the improper use of a traditional ayurvedic drug.

Authors:  L Borgese; A Zacco; E Bontempi; M Pellegatta; L Vigna; L Patrini; L Riboldi; F M Rubino; L E Depero
Journal:  J Pharm Biomed Anal       Date:  2010-03-01       Impact factor: 3.935

2.  Polymer-coated quartz crystal microbalance chemical sensor for heavy cations in water.

Authors:  L Sartore; M Barbaglio; M Penco; P Bergese; E Bontempi; P Colombi; L E Depero
Journal:  J Nanosci Nanotechnol       Date:  2009-02

3.  A review on solid phase microextraction-high performance liquid chromatography as a novel tool for the analysis of toxic metal ions.

Authors:  Ashok Kumar Malik; Varinder Kaur; Neelam Verma
Journal:  Talanta       Date:  2005-07-22       Impact factor: 6.057

4.  Comparison of different protein immobilization methods on quartz crystal microbalance surface in flow injection immunoassay.

Authors:  Y C Liu; C M Wang; K P Hsiung
Journal:  Anal Biochem       Date:  2001-12-15       Impact factor: 3.365

5.  Polymer chemical structure is a key determinant of physicochemical and colloidal properties of polymer-DNA complexes for gene delivery.

Authors:  N A Jones; I R Hill; S Stolnik; F Bignotti; S S Davis; M C Garnett
Journal:  Biochim Biophys Acta       Date:  2000-12-15

6.  Piezoelectric sensor functionalised by a self-assembled bipyridinium derivative: characterisation and preliminary applications in the detection of heavy metal ions.

Authors:  Serena Casilli; Cosimino Malitesta; Sabrina Conoci; Salvatore Petralia; Salvatore Sortino; Ludovico Valli
Journal:  Biosens Bioelectron       Date:  2004-12-15       Impact factor: 10.618

Review 7.  Recent results on functional polymers and macromonomers of interest as biomaterials or for biomaterial modification.

Authors:  P Ferruti; E Ranucci; L Sartore; F Bignotti; M A Marchisio; P Bianciardi; F M Veronese
Journal:  Biomaterials       Date:  1994-12       Impact factor: 12.479

8.  A comparison of boron-doped diamond thin-film and Hg-coated glassy carbon electrodes for anodic stripping voltammetric determination of heavy metal ions in aqueous media.

Authors:  Elizabeth A McGaw; Greg M Swain
Journal:  Anal Chim Acta       Date:  2006-06-06       Impact factor: 6.558

  8 in total
  6 in total

1.  TXRF analysis of soils and sediments to assess environmental contamination.

Authors:  Fabjola Bilo; Laura Borgese; Davide Cazzago; Annalisa Zacco; Elza Bontempi; Rita Guarneri; Marco Bernardello; Silvia Attuati; Pranvera Lazo; Laura E Depero
Journal:  Environ Sci Pollut Res Int       Date:  2014-12       Impact factor: 4.223

2.  Improving the binding efficiency of quartz crystal microbalance biosensors by applying the electrothermal effect.

Authors:  Yao-Hung Huang; Jeng-Shian Chang; Sheng D Chao; Kuang-Chong Wu; Long-Sun Huang
Journal:  Biomicrofluidics       Date:  2014-10-15       Impact factor: 2.800

3.  Piezoelectric arsenite aptasensor based on the use of a self-assembled mercaptoethylamine monolayer and gold nanoparticles.

Authors:  Min Yuan; Qinqin Zhang; Zhihong Song; Tai Ye; Jinsong Yu; Hui Cao; Fei Xu
Journal:  Mikrochim Acta       Date:  2019-04-05       Impact factor: 5.833

4.  Dew Point Calibration System Using a Quartz Crystal Sensor with a Differential Frequency Method.

Authors:  Ningning Lin; Xiaofeng Meng; Jing Nie
Journal:  Sensors (Basel)       Date:  2016-11-18       Impact factor: 3.576

5.  Evaluation of complexation ability using a sensor electrode chip equipped with a wireless screening system.

Authors:  Takaaki Isoda; Ikuko Urushibara; Hikaru Sato; Noriyoshi Yamauchi
Journal:  Sensors (Basel)       Date:  2012-06-19       Impact factor: 3.576

6.  Semi-Crystalline Hydrophobic Polyamidoamines: A New Family of Technological Materials?

Authors:  Massimo Marcioni; Jenny Alongi; Elisabetta Ranucci; Mario Malinconico; Paola Laurienzo; Paolo Ferruti; Amedea Manfredi
Journal:  Polymers (Basel)       Date:  2021-03-25       Impact factor: 4.329

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.