Metabolic biofouling of glucose sensors in vivo: role of tissue microhemorrhages.

Ulrike Klueh1, Zenghe Liu, Ben Feldman, Timothy P Henning, Brian Cho, Tianmei Ouyang, Don Kreutzer.   

Abstract

OBJECTIVE: Based on our in vitro study that demonstrated the adverse effects of blood clots on glucose sensor function, we hypothesized that in vivo local tissue hemorrhages, induced as a consequence of sensor implantation or sensor movement post-implantation, are responsible for unreliable readings or an unexplained loss of functionality shortly after implantation. RESEARCH DESIGN AND METHODS: To investigate this issue, we utilized real-time continuous monitoring of blood glucose levels in a mouse model. Direct injection of blood at the tissue site of sensor implantation was utilized to mimic sensor-induced local tissue hemorrhages.
RESULTS: It was found that blood injections, proximal to the sensor, consistently caused lowered sensor glucose readings, designated temporary signal reduction, in vivo in our mouse model, while injections of plasma or saline did not have this effect.
CONCLUSION: These results support our hypothesis that tissue hemorrhage and resulting blood clots near the sensor can result in lowered local blood glucose concentrations due to metabolism of glucose by the clot. The lowered local blood glucose concentration led to low glucose readings from the still functioning sensor that did not reflect the systemic glucose level.
© 2011 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21722574      PMCID: PMC3192625          DOI: 10.1177/193229681100500313

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  15 in total

1.  Influence of inflammatory cells and serum on the performance of implantable glucose sensors.

Authors:  M Gerritsen; J A Jansen; A Kros; D M Vriezema; N A Sommerdijk; R J Nolte; J A Lutterman; S W Van Hövell; A Van der Gaag
Journal:  J Biomed Mater Res       Date:  2001-01

2.  Enhancement of implantable glucose sensor function in vivo using gene transfer-induced neovascularization.

Authors:  Ulrike Klueh; David I Dorsky; Don L Kreutzer
Journal:  Biomaterials       Date:  2005-04       Impact factor: 12.479

3.  Assessment of chronically implanted subcutaneous glucose sensors in dogs: the effect of surrounding fluid masses.

Authors:  W K Ward; J E Troupe
Journal:  ASAIO J       Date:  1999 Nov-Dec       Impact factor: 2.872

4.  Protein interactions with subcutaneously implanted biosensors.

Authors:  Raeann Gifford; Joseph J Kehoe; Sandra L Barnes; Boris A Kornilayev; Michail A Alterman; George S Wilson
Journal:  Biomaterials       Date:  2005-12-20       Impact factor: 12.479

5.  Saphenous vein puncture for blood sampling of the mouse, rat, hamster, gerbil, guinea pig, ferret and mink.

Authors:  A Hem; A J Smith; P Solberg
Journal:  Lab Anim       Date:  1998-10       Impact factor: 2.471

6.  Continuous subcutaneous glucose monitoring in diabetic patients: a multicenter analysis.

Authors:  Alberto Maran; Cristina Crepaldi; Antonio Tiengo; Giorgio Grassi; Emanuela Vitali; Gianfranco Pagano; Sergio Bistoni; Giuseppe Calabrese; Fausto Santeusanio; Frida Leonetti; Maria Ribaudo; Umberto Di Mario; Giovanni Annuzzi; Salvatore Genovese; Gabriele Riccardi; Marcello Previti; Domenico Cucinotta; Francesco Giorgino; Aurelia Bellomo; Riccardo Giorgino; Alessandro Poscia; Maurizio Varalli
Journal:  Diabetes Care       Date:  2002-02       Impact factor: 19.112

7.  Critical role of tissue mast cells in controlling long-term glucose sensor function in vivo.

Authors:  Ulrike Klueh; Manjot Kaur; Yi Qiao; Donald L Kreutzer
Journal:  Biomaterials       Date:  2010-03-11       Impact factor: 12.479

8.  Introductory experience with the continuous glucose monitoring system (CGMS; Medtronic Minimed) in detecting hyperglycemia by comparing the self-monitoring of blood glucose (SMBG) in non-pregnant women and in pregnant women with impaired glucose tolerance and gestational diabetes.

Authors:  K J Bühling; B Kurzidim; C Wolf; K Wohlfarth; M Mahmoudi; C Wäscher; G Siebert; J W Dudenhausen
Journal:  Exp Clin Endocrinol Diabetes       Date:  2004-11       Impact factor: 2.949

9.  Continuous glucose monitoring: reliable measurements for up to 4 days with the SCGM1 system.

Authors:  Christoph Kapitza; Volker Lodwig; Karin Obermaier; Klaas Jan C Wientjes; Klaas Hoogenberg; Karsten Jungheim; Lutz Heinemann
Journal:  Diabetes Technol Ther       Date:  2003       Impact factor: 6.118

10.  Accuracy of the continuous glucose monitoring system in inpatient and outpatient conditions.

Authors:  C Djakouré-Platonoff; R Radermercker; G Reach; G Slama; J I Selam
Journal:  Diabetes Metab       Date:  2003-04       Impact factor: 6.041

View more
  29 in total

1.  Tissue Response to Subcutaneous Infusion Catheter.

Authors:  Ershuai Zhang; Zhiqiang Cao
Journal:  J Diabetes Sci Technol       Date:  2019-03-31

Review 2.  In Vivo Chemical Sensors: Role of Biocompatibility on Performance and Utility.

Authors:  Robert J Soto; Jackson R Hall; Micah D Brown; James B Taylor; Mark H Schoenfisch
Journal:  Anal Chem       Date:  2016-11-21       Impact factor: 6.986

3.  Glucose Sensing in the Subcutaneous Tissue: Attempting to Correlate the Immune Response with Continuous Glucose Monitoring Accuracy.

Authors:  Jeffrey I Joseph; Gabriella Eisler; David Diaz; Abdurizzagh Khalf; Channy Loeum; Marc C Torjman
Journal:  Diabetes Technol Ther       Date:  2018-05       Impact factor: 6.118

4.  Interstitial fluid physiology as it relates to glucose monitoring technologies: symposium introduction.

Authors:  Natalie A Wisniewski; Uli Klueh; Julie Stenken
Journal:  J Diabetes Sci Technol       Date:  2011-05-01

5.  Analysis: on the path to overcoming glucose-sensor-induced foreign body reactions.

Authors:  Ulrike Klueh
Journal:  J Diabetes Sci Technol       Date:  2013-03-01

6.  A promising solution to enhance the sensocompatibility of biosensors in continuous glucose monitoring systems.

Authors:  Edith E M van den Bosch; Nik H M de Bont; Jun Qiu; Onko-Jan Gelling
Journal:  J Diabetes Sci Technol       Date:  2013-03-01

Review 7.  Advances in management of type 1 diabetes mellitus.

Authors:  Ravindranath Aathira; Vandana Jain
Journal:  World J Diabetes       Date:  2014-10-15

Review 8.  Biocompatible materials for continuous glucose monitoring devices.

Authors:  Scott P Nichols; Ahyeon Koh; Wesley L Storm; Jae Ho Shin; Mark H Schoenfisch
Journal:  Chem Rev       Date:  2013-02-07       Impact factor: 60.622

9.  Cell based metabolic barriers to glucose diffusion: macrophages and continuous glucose monitoring.

Authors:  Ulrike Klueh; Jackman T Frailey; Yi Qiao; Omar Antar; Donald L Kreutzer
Journal:  Biomaterials       Date:  2014-01-22       Impact factor: 12.479

10.  Redundancy in Glucose Sensing: Enhanced Accuracy and Reliability of an Electrochemical Redundant Sensor for Continuous Glucose Monitoring.

Authors:  Amin Sharifi; Andrea Varsavsky; Johanna Ulloa; Jodie C Horsburgh; Sybil A McAuley; Balasubramanian Krishnamurthy; Alicia J Jenkins; Peter G Colman; Glenn M Ward; Richard J MacIsaac; Rajiv Shah; David N O'Neal
Journal:  J Diabetes Sci Technol       Date:  2016-05-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.