| Literature DB >> 21674146 |
Xiaolei Chen1, Bai-nan Xu, Xianghui Meng, Jun Zhang, Xingguang Yu, Dingbiao Zhou.
Abstract
We hereby report our initial clinical experience of a dual-room intraoperative magnetic resonance imaging (iMRI) suite with a movable 1.5-T magnet for both neurosurgical and independent diagnostic uses. The findings from the first 45 patients who underwent scheduled neurosurgical procedures with iMRI in this suite (mean age, 41.3 ± 12.0 years; intracranial tumors, 39 patients; cerebral vascular lesions, 5 patients; epilepsy surgery, 1 patient) were reported. The extent of resection depicted at intraoperative imaging, the surgical consequences of iMRI, and the clinical practicability of the suite were analyzed. Fourteen resections with a trans-sphenoidal/transoral approach and 31 craniotomies were performed. Eighty-two iMRI examinations were performed in the operating room, while during the same period of time, 430 diagnostic scans were finished in the diagnostic room. In 22 (48.9%) of 45 patients, iMRI revealed accessible residual tumors leading to further resection. No iMRI-related adverse event occurred. Complete lesion removal was achieved in 36 (80%) of all 45 cases. It is concluded that the dual-room 1.5-T iMRI suite can be successfully integrated into standard neurosurgical workflow. The layout of the dual-room suite can enable the maximum use of the system and save costs by sharing use of the 1.5-T magnet between neurosurgical and diagnostic use. Intraoperative MR imaging may provide valuable information that allows intraoperative modification of the surgical strategy.Entities:
Mesh:
Year: 2011 PMID: 21674146 DOI: 10.1007/s10143-011-0336-3
Source DB: PubMed Journal: Neurosurg Rev ISSN: 0344-5607 Impact factor: 3.042