Postanesthesia patients with large upper arm circumference: is use of an "extra-long" adult cuff or forearm cuff placement accurate?

Sheri Watson1, Marita Aguas, Tracy Bienapfl, Pat Colegrove, Nancy Foisy, Bonnie Jondahl, Mary Beth Yosses, Larissa Yu, Zoe Anastas.   

Abstract

The purpose of this study was to determine if blood pressure (BP) measured in the forearm or with an extra-long BP cuff in the upper arm accurately reflects BP measured in the upper arm with an appropriately sized BP cuff in patients with large upper arm circumference. A method-comparison design was used with a convenience sample of 49 PACU patients. Noninvasive blood pressures were obtained in two different locations (forearm; upper arm) and in the upper arm with an extra-long adult and recommended large adult cuff sizes. Data were analyzed by calculating bias and precision for the BP cuff size and location and Student's t-tests, with P < .0125 considered significant. Significantly higher forearm systolic (P < .0001) and diastolic (P < .0002) BP measurements were found compared to BP obtained in the upper arm with the reference standard BP cuff. Significantly higher systolic (t(48df) = 5.38, P < .0001), but not diastolic (t(48df) = 4.11, P < .019), BP differences were found for BP measured with the extra-long cuff at the upper arm site compared to the upper arm, reference standard BP. Findings suggest that the clinical practice of using the forearm or an extra-long cuff in the upper arm for BP measurement in post anesthesia patients with large upper arm circumferences may result in inaccurate BP values.
Copyright © 2011 American Society of PeriAnesthesia Nurses. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21641528     DOI: 10.1016/j.jopan.2011.02.001

Source DB:  PubMed          Journal:  J Perianesth Nurs        ISSN: 1089-9472            Impact factor:   1.084


  4 in total

Review 1.  Accuracy of non-invasive and minimally invasive hemodynamic monitoring: where do we stand?

Authors:  Issa Pour-Ghaz; Theodore Manolukas; Nathalie Foray; Joel Raja; Aranyak Rawal; Uzoma N Ibebuogu; Rami N Khouzam
Journal:  Ann Transl Med       Date:  2019-09

Review 2.  Techniques for self-measurement of blood pressure: limitations and needs for future research.

Authors:  Paolo Palatini; Gerhard N Frick
Journal:  J Clin Hypertens (Greenwich)       Date:  2012-01-23       Impact factor: 3.738

3.  Is it reliable to measure the forearm blood pressure in children?

Authors:  Amar M Taksande; Aishwarya Jadhav; Jyoti Nair
Journal:  J Family Community Med       Date:  2015 May-Aug

4.  The Effect of Cuff Size on Blood Pressure Measurement in Obese Surgical Patients: A Prospective Crossover Clinical Trial.

Authors:  Serpil Yüksel; Gülay Altun-Uğraş; Nurhan Altınok; Necla Demir
Journal:  Florence Nightingale J Nurs       Date:  2020-07-03
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.