| Literature DB >> 21592916 |
Ann M Simon, Levi J Hargrove, Blair A Lock, Todd A Kuiken.
Abstract
Real-time pattern recognition control is frequently affected by misclassifications. This study investigated the use of a decision-based velocity ramp that attenuated movement speed after a change in classifier decision. The goal was to improve prosthesis positioning by minimizing the effect of unintended movements. Non-amputee and amputee subjects controlled a prosthesis in real-time using pattern recognition. While performing a target achievement test in a virtual environment, subjects had a significantly higher completion rate (p < 0.05) and a more direct path (p < 0.05) to the target with the velocity ramp than without it. Using a physical prosthesis, subjects stacked a greater average number of 1 cubes (p < 0.05) in three minutes with the velocity ramp than without it (76% more blocks for non-amputees; 89% more blocks for amputees). Real-time control using the velocity ramp also showed significant performance improvements above using majority vote. Eighty-three percent of subjects preferred to control the prosthesis using the velocity ramp. These results suggest that using a decision-based velocity ramp with pattern recognition may improve user performance. Since the velocity ramp is a post-processing step, it has the potential to be used with a variety of classifiers for many applications.Entities:
Mesh:
Year: 2011 PMID: 21592916 PMCID: PMC4269322 DOI: 10.1109/TBME.2011.2155063
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538