Literature DB >> 21563250

In vitro and in vivo comparison of wrist MR imaging at 3.0 and 7.0 tesla using a gradient echo sequence and identical eight-channel coil array designs.

Jurek A Nordmeyer-Massner1, Michael Wyss, Gustav Andreisek, Klaas P Pruessmann, Juerg Hodler.   

Abstract

PURPOSE: To evaluate in vivo MR imaging of the wrist at 3.0 Tesla (T) and 7.0T quantitatively and qualitatively.
MATERIALS AND METHODS: To enable unbiased signal-to-noise ratio (SNR) comparisons, geometrically identical eight-channel receiver arrays were used at both field strengths. First, in vitro images of a phantom bottle were acquired at 3.0T and 7.0T to obtain an estimate of the maximum SNR gain that can be expected. MR images of the dominant wrist of 10 healthy volunteers were acquired at both field strengths. All measurements were done using the same sequence parameters. Quantitative SNR maps were calculated on a pixel-by-pixel basis and analyzed in several regions-of-interest. Furthermore, the images were qualitatively evaluated by two independent radiologists.
RESULTS: The quantitative analysis showed SNR increases of up to 100% at 7.0T compared with 3.0T, with considerable variation between different anatomical structures. The qualitative analysis revealed no significant difference in the visualization of anatomical structures comparing 3.0T and 7.0T MR images (P>0.05).
CONCLUSION: The presented results establish the SNR benefits of the transition from 3.0T to 7.0T for wrist imaging without bias by different array designs and based on exact, algebraic SNR quantification. The observed SNR increase nearly reaches expected values but varies greatly between different tissues. It does not necessarily improve the visibility of anatomic structures but adds valuable latitude for sequence optimization.
Copyright © 2011 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21563250     DOI: 10.1002/jmri.22419

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  9 in total

1.  Quantitative and qualitative comparison of MR imaging of the temporomandibular joint at 1.5 and 3.0 T using an optimized high-resolution protocol.

Authors:  Andrei Manoliu; Georg Spinner; Michael Wyss; Stefan Erni; Dominik A Ettlin; Daniel Nanz; Erika J Ulbrich; Luigi M Gallo; Gustav Andreisek
Journal:  Dentomaxillofac Radiol       Date:  2015-09-15       Impact factor: 2.419

2.  Comparison of a 32-channel head coil and a 2-channel surface coil for MR imaging of the temporomandibular joint at 3.0 T.

Authors:  Andrei Manoliu; Georg Spinner; Michael Wyss; Lukas Filli; Stefan Erni; Dominik A Ettlin; Erika J Ulbrich; Felix P Kuhn; Luigi M Gallo; Gustav Andreisek
Journal:  Dentomaxillofac Radiol       Date:  2016-02-03       Impact factor: 2.419

Review 3.  Quantitative MRI in the evaluation of articular cartilage health: reproducibility and variability with a focus on T2 mapping.

Authors:  Rachel K Surowiec; Erin P Lucas; Charles P Ho
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-10-30       Impact factor: 4.342

4.  MR imaging of the temporomandibular joint: comparison between acquisitions at 7.0 T using dielectric pads and 3.0 T.

Authors:  Felix P Kuhn; Georg Spinner; Filippo Del Grande; Michael Wyss; Marco Piccirelli; Stefan Erni; Pascal Pfister; Michael Ho; Bert-Ram Sah; Lukas Filli; Dominik A Ettlin; Luigi M Gallo; Gustav Andreisek; Andrei Manoliu
Journal:  Dentomaxillofac Radiol       Date:  2016-12-18       Impact factor: 2.419

5.  Hip imaging of avascular necrosis at 7 Tesla compared with 3 Tesla.

Authors:  J M Theysohn; O Kraff; N Theysohn; S Orzada; S Landgraeber; M E Ladd; T C Lauenstein
Journal:  Skeletal Radiol       Date:  2014-02-05       Impact factor: 2.199

6.  Determining the optimal magnetic resonance imaging sequences for the efficient diagnosis of temporomandibular joint disorders.

Authors:  Minjun Dong; Qi Sun; Qiang Yu; Xiaofeng Tao; Chi Yang; Weiliu Qiu
Journal:  Quant Imaging Med Surg       Date:  2021-04

7.  Diffusion-weighted 7.0T Magnetic Resonance Imaging in Assessment of Intervertebral Disc Degeneration in Rats.

Authors:  Long-Yang Li; Xiao-Lin Wu; Richard J Roman; Fan Fan; Chen-Sheng Qiu; Bo-Hua Chen
Journal:  Chin Med J (Engl)       Date:  2018-01-05       Impact factor: 2.628

8.  Selection and application of coils in temporomandibular joint MRI.

Authors:  Qi Sun; Min-Jun Dong; Xiao-Feng Tao; Meng-da Jiang; Chi Yang
Journal:  Dentomaxillofac Radiol       Date:  2019-09-27       Impact factor: 2.419

9.  Visualization of wrist anatomy-a comparison between 7T and 3T MRI.

Authors:  Simon Götestrand; Anders Björkman; Isabella M Björkman-Burtscher; Ingvar Kristiansson; Elenya Aksyuk; Pawel Szaro; Karin Markenroth Bloch; Mats Geijer
Journal:  Eur Radiol       Date:  2021-08-11       Impact factor: 5.315

  9 in total