| Literature DB >> 21544189 |
Malgorzata Szafarz, Joanna Szymura-Oleksiak, Dorota Lazewska, Katarzyna Kiec-Kononowicz.
Abstract
A sensitive and specific liquid chromatography electrospray ionisation-tandem mass spectrometry method for determination of new non-imidazole histamine H(3) receptor antagonist 1-[3-(4-tert-butylphenoxy)propyl]piperidine (DL76) in rat serum has been developed and validated. Chromatography was performed on a XBridge™ C18 analytical column (2.1 × 30 mm, 3.5 µm, Waters, Ireland) with gradient elution using a mobile phase containing acetonitrile and water with an addition of 0.1% of formic acid. Detection was achieved by an Applied Biosystems MDS Sciex (Concord, Ontario, Canada) API 2000 triple quadrupole mass spectrometer. Electrospray ionization (ESI) was used for ion production. The limit of detection in the SRM mode was found to be 0.5 ng mL(-1). The limit of quantification was 1 ng mL(-1). The precision and accuracy for both intra- and inter-day determination of DL76 ranged from 1.65 to 15.09% and from 88.74 to 113.43%. The results of this analytical method validation allow to carry out pharmacokinetic studies in rats. The method was used for the pilot study of the pharmacokinetic behavior of DL76 in rats after intravenous administration.Entities:
Year: 2011 PMID: 21544189 PMCID: PMC3073091 DOI: 10.1007/s10337-011-1983-9
Source DB: PubMed Journal: Chromatographia ISSN: 0009-5893 Impact factor: 2.044
Fig. 1Product ion mass spectra of [M+H+]+ ions of PTX (a) and DL76 (b) and proposed fragmentation paths
Fig. 2The representative extracted ion chromatograms (XIC) of blank serum (a), serum sample spiked with the concentration of 1 ng mL−1 of DL76 (b), rat serum sample after i.v. administration of DL76 in the dose of 3 mg kg−1, with the concentration of DL76 determined at 67.82 ng mL−1 (c) and IS (d)
Back-calculated concentrations of calibrators for DL76
| Nominal concentration of DL76 (ng mL−1) | Back calculated concentration (mean ± SD) | RSD (%) | Accuracy (%) | Bias (%) |
|---|---|---|---|---|
| 1 | 1.00 ± 0.15 | 14.81 | 100 | 0 |
| 10 | 9.92 ± 1.06 | 10.66 | 99.2 | −0.8 |
| 50 | 50.58 ± 1.59 | 3.15 | 101.16 | 1.16 |
| 100 | 102.11 ± 7.30 | 7.15 | 102.11 | 2.11 |
| 250 | 250.46 ± 10.82 | 4.32 | 100.18 | 0.18 |
| 500 | 499.36 ± 14.72 | 2.95 | 99.87 | −0.13 |
| 1,000 | 1,018.52 ± 39.34 | 3.86 | 101.85 | 1.85 |
| 2,000 | 1,992.28 ± 23.40 | 1.17 | 99.61 | −0.39 |
Within-day and between days accuracy (% of nominal concentration) and precision (%RSD) of DL76 in rat serum
| QC i.d | QC nominal concentration (ng mL−1) | One day | Different days | ||
|---|---|---|---|---|---|
| Precision (RSD, %) | Mean accuracy (%) | Precision (RSD, %) | Mean accuracy (%) | ||
| LLOQ | 1 | 15.09 | 112.53 | 12.8 | 113.43 |
| Low | 40 | 5.95 | 88.75 | 9.75 | 91.27 |
| Medium | 400 | 4.99 | 95.74 | 7.81 | 92.56 |
| High | 1,800 | 1.65 | 88.74 | 2.14 | 89.79 |
Bench top and postpreparative stability of DL 76
| QC i.d. | QC nominal concentration (ng mL−1) | Stability after 2 h in 20 °C (bench top stability) | Stability after 24 h in 10 °C (post preparative stability) | ||||
|---|---|---|---|---|---|---|---|
| Mean concentration found (%) | Accuracy (%) | CV (%) | Mean concentration found (%) | Accuracy (%) | CV (%) | ||
| LLOQ | 1 | 0.91 | 91.4 | 3.0 | 1.09 | 85.71 | 9.52 |
| Low | 40 | 43.62 | 109.1 | 6.7 | 38.57 | 102.68 | 6.67 |
| Medium | 400 | 395.52 | 98.9 | 11.2 | 382.4 | 85.49 | 3.09 |
| High | 1,800 | 1,800.91 | 100.1 | 14.8 | 1,843.02 | 106.93 | 2.43 |
Long term and freeze/thaw stability of DL 76
| QC i.d. | QC nominal concentration (ng mL−1) | Stability after three freeze/thaw cycles | Stability after 2 months in −30 °C (long term stability) | ||||
|---|---|---|---|---|---|---|---|
| Mean concentration found (%) | Accuracy (%) | CV (%) | Mean concentration found (%) | Accuracy (%) | CV (%) | ||
| LLOQ | 1 | 0.98 | 98.4 | 5.5 | 0.88 | 87.7 | 13.3 |
| Low | 40 | 42.63 | 106.6 | 4.5 | 44.79 | 112.0 | 6.5 |
| Medium | 400 | 374.98 | 93.7 | 3.8 | 401.95 | 100.5 | 1.9 |
| High | 1,800 | 1,822.41 | 101.2 | 0.8 | 1,794.94 | 99.7 | 1.3 |
Fig. 3Mean plasma concentrations ±SD (n = 5) versus time profile obtained after a single i.v. dose (3 mg kg−1) of DL76 compound