A pulsatile simulator for the in vitro analysis of the mitral valve with tri-axial papillary muscle displacement.

Riccardo Vismara1, Andrea Pavesi, Emiliano Votta, Maurizio Taramasso, Francesco Maisano, Gianfranco B Fiore.   

Abstract

PURPOSE: We developed a new pulsatile hydrodynamic simulator for the in vitro testing of mitral valve (MV) samples. The required specifications included a 3D positioning system for the papillary muscles (PMs) that is accurate and simple to manage; measurement of the force exerted by the chordae tendineae on the PMs; and the possibility to visually inspect the MV for kinematic analysis.
METHODS: An atrial/ventricular chamber system was developed. The ventricular chamber housed a tri-axial actuator system that was aligned to a morphometric Cartesian frame, allowing for PM positioning even while tests are running. Each PM holder had an embedded load cell for force measurement. The atrial chamber was designed so as to permit MV visual inspection, maintaining a non-disturbed flow at the sample inlet. The setup was subjected to trials with fresh porcine MVs. Flow and pressure difference across the MVs and PM forces were measured in different MV configurations, with different PM spatial dislocations. High speed video recordings were acquired.
RESULTS: The positioning accuracy was assessed. Tests with MVs showed good usability, even by the non-engineering personnel. The effects of PM displacement on valve function (valve competence and PM forces) was consistent with previously published data, thus confirming the general soundness of the design principles.
CONCLUSIONS: The developed simulator is a promising instrument for performing MV in vitro tests in a precise, well-repeatable manner. The ability to completely adjust the PM position while a test is running boosts the simulator's potential for detailed investigations of the pathological and surgically treated MV.

Entities:  

Mesh:

Year:  2011        PMID: 21534249     DOI: 10.5301/IJAO.2011.7729

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  6 in total

1.  Biomimetic six-axis robots replicate human cardiac papillary muscle motion: pioneering the next generation of biomechanical heart simulator technology.

Authors:  Annabel M Imbrie-Moore; Matthew H Park; Michael J Paulsen; Mark Sellke; Rohun Kulkami; Hanjay Wang; Yuanjia Zhu; Justin M Farry; Alexandra T Bourdillon; Christine Callinan; Haley J Lucian; Camille E Hironaka; Daniela Deschamps; Y Joseph Woo
Journal:  J R Soc Interface       Date:  2020-12-02       Impact factor: 4.118

2.  Pre-clinical In Vitro and In Vivo Models for Heart Valve Therapies.

Authors:  Maurizio Taramasso; Maximilian Y Emmert; Diana Reser; Andrea Guidotti; Nikola Cesarovic; Marino Campagnol; Alessandro Addis; Fabian Nietlispach; Simon P Hoerstrup; Francesco Maisano
Journal:  J Cardiovasc Transl Res       Date:  2015-05-13       Impact factor: 4.132

3.  In vitro mitral valve simulator mimics systolic valvular function of chronic ischemic mitral regurgitation ovine model.

Authors:  Andrew W Siefert; Jean Pierre M Rabbah; Kevin J Koomalsingh; Steven A Touchton; Neelakantan Saikrishnan; Jeremy R McGarvey; Robert C Gorman; Joseph H Gorman; Ajit P Yoganathan
Journal:  Ann Thorac Surg       Date:  2013-01-29       Impact factor: 4.330

4.  A novel cross-species model of Barlow's disease to biomechanically analyze repair techniques in an ex vivo left heart simulator.

Authors:  Annabel M Imbrie-Moore; Michael J Paulsen; Yuanjia Zhu; Hanjay Wang; Haley J Lucian; Justin M Farry; John W MacArthur; Michael Ma; Y Joseph Woo
Journal:  J Thorac Cardiovasc Surg       Date:  2020-02-19       Impact factor: 5.209

5.  Hydrodynamic Noise of Pulsating Jets through Bileaflet Mechanical Mitral Valve.

Authors:  Vladimir Voskoboinick; Oleksandr Voskoboinyk; Oleg Chertov; Andrey Voskoboinick; Lidiia Tereshchenko
Journal:  Biomed Res Int       Date:  2020-05-30       Impact factor: 3.411

6.  A Single-opening&closing Valve Tester for Direct Measurement of Closing Volume of the Heart Valve.

Authors:  Hao Wang; Zhan Cui; Zhongxi Zhou; Zhaoming He
Journal:  Cardiovasc Eng Technol       Date:  2021-06-25       Impact factor: 2.495

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.