Literature DB >> 21522024

Optic disc parameters from optovue optical coherence tomography: comparison of manual versus automated disc rim determination.

Nisreen K Mesiwala1, Melike Pekmezci, Travis C Porco, Shan C Lin.   

Abstract

PURPOSE: To compare optic disc parameters when using manual disc detection versus automated software on RTVue-100 Fourier-domain optical coherence tomography.
METHODS: Nonrandomized comparative study using images obtained from a tertiary-care Glaucoma Clinic from August, 2008 to May, 2009. Four hundred thirty-four patients with mean age of 67.1 years (±13.4) underwent OCT imaging with RTVue-100. Optic disc margins were manually defined by a single, trained operator and redefined with automated software that estimates the position of the retinal pigment epithelium edges. Bland-Altman plots and intraclass correlations measured agreement between the 2 detection methods for various optic disc parameters.
RESULTS: Bland-Altman plots/intraclass correlations showed good agreement in the following optic disc parameters: optic disc area, cup area, rim area, cup-to-disc area ratio, cup-to-disc ratio, and retinal nerve fiber layer thickness. There was no significant difference in measurements when comparing hand versus automated disc contour determination (P>0.05).
CONCLUSIONS: Automated determination of the disc margin was in high agreement with subjective, manual assessment and should be used in clinical practice as it can make optical coherence tomography scanning more efficient and standardized, as disc detection would only be done by the software and not biased by multiple operators.

Entities:  

Mesh:

Year:  2012        PMID: 21522024     DOI: 10.1097/IJG.0b013e31821206e8

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  4 in total

1.  The comparison of manual vs automated disc margin delineation using spectral-domain optical coherence tomography.

Authors:  S M Iverson; M Sehi
Journal:  Eye (Lond)       Date:  2013-08-02       Impact factor: 3.775

2.  A method for age-matched OCT angiography deviation mapping in the assessment of disease- related changes to the radial peripapillary capillaries.

Authors:  Alexander Pinhas; Rachel Linderman; Shelley Mo; Brian D Krawitz; Lawrence S Geyman; Joseph Carroll; Richard B Rosen; Toco Y Chui
Journal:  PLoS One       Date:  2018-05-24       Impact factor: 3.240

3.  Correlation between the optic nerve pial diameter and radial peripapillary vascular changes in primary open-angle glaucoma.

Authors:  Daniela Montorio; Gilda Cennamo; Maria Angelica Breve; Feliciana Menna; Michele Reibaldi; Giovanni Cennamo; Vincenzo Brescia Morra
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-10-21       Impact factor: 3.117

4.  Software-Assisted Depth Analysis of Optic Nerve Stereoscopic Images in Telemedicine.

Authors:  Tian Xia; Shriji N Patel; Ben C Szirth; Anton M Kolomeyer; Albert S Khouri
Journal:  Int J Telemed Appl       Date:  2016-04-14
  4 in total