DOI-based reconstruction algorithms for a compact breast PET scanner.

Kyle M Champley1, Lawrence R MacDonald, Thomas K Lewellen, Robert S Miyaoka, Paul E Kinahan.   

Abstract

PURPOSE: The authors discuss the design and evaluate the performance of combined event estimation and image reconstruction algorithms designed for a proposed high-resolution rectangular breast PET scanner (PETX). The PETX scanner will be capable of measuring the depth of interaction by utilizing detector modules composed of depth-of-interaction microcrystal element (dMiCE) crystal pairs. This design allows a unique combination of event estimation and fast projection methods.
METHODS: The authors implemented a Monte Carlo simulator to model the PETX system using only true coincident events. The performance of the dMiCE crystal pairs was determined experimentally over a range of depths of interaction. This distribution was used to simulate the noisy dMiCE detector signals and to estimate the line of response for each decay. Three different statistical methods were implemented to determine photon event positioning. Images were reconstructed from these line of response estimators with the exact planogram frequency distance rebinning algorithm, which is an exact analytical reconstruction algorithm for planar systems. Reconstructed images were analyzed with contrast, noise, and spatial resolution metrics.
RESULTS: The authors' simulations demonstrate the ability for the PETX system to produce quantitatively accurate images from true coincident events with a contrast recovery coefficient of greater than 0.8 for 5 mm spheres at the axial center of the scanner and a spatial resolution (FWHM) of 3 mm throughout most of the imaging field of view.
CONCLUSIONS: The authors' proposed positioning and reconstruction algorithms for the PETX system show the potential for creating high-quality, high-resolution, and quantitatively accurate images within a clinically feasible reconstruction time.

Mesh:

Year:  2011        PMID: 21520879      PMCID: PMC3064687          DOI: 10.1118/1.3547712

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  17 in total

Review 1.  Positron-emission tomography and assessment of cancer therapy.

Authors:  Malik E Juweid; Bruce D Cheson
Journal:  N Engl J Med       Date:  2006-02-02       Impact factor: 91.245

2.  Study of the performance of a novel 1 mm resolution dual-panel PET camera design dedicated to breast cancer imaging using Monte Carlo simulation.

Authors:  Jin Zhang; Peter D Olcott; Garry Chinn; Angela M K Foudray; Craig S Levine
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

3.  New Directions for dMiCE - a Depth-of-Interaction Detector Design for PET Scanners.

Authors:  T K Lewellen; L R Macdonald; R S Miyaoka; W McDougald; K Champley
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2007

4.  Exact and approximate rebinning algorithms for 3-D PET data.

Authors:  M Defrise; P E Kinahan; D W Townsend; C Michel; M Sibomana; D F Newport
Journal:  IEEE Trans Med Imaging       Date:  1997-04       Impact factor: 10.048

5.  Quantitative fluoroestradiol positron emission tomography imaging predicts response to endocrine treatment in breast cancer.

Authors:  Hannah M Linden; Svetlana A Stekhova; Jeanne M Link; Julie R Gralow; Robert B Livingston; Georgiana K Ellis; Philip H Petra; Lanell M Peterson; Erin K Schubert; Lisa K Dunnwald; Kenneth A Krohn; David A Mankoff
Journal:  J Clin Oncol       Date:  2006-05-08       Impact factor: 44.544

6.  Advancements to the planogram frequency-distance rebinning algorithm.

Authors:  Kyle M Champley; Raymond R Raylman; Paul E Kinahan
Journal:  Inverse Probl       Date:  2010-03-25       Impact factor: 2.407

7.  Quantification of radiotracer uptake with a dedicated breast PET imaging system.

Authors:  Raymond R Raylmana; Mark F Smith; Paul E Kinahan; Stan Majewski
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

8.  Clinical imaging characteristics of the positron emission mammography camera: PEM Flex Solo II.

Authors:  Lawrence MacDonald; John Edwards; Thomas Lewellen; David Haseley; James Rogers; Paul Kinahan
Journal:  J Nucl Med       Date:  2009-09-16       Impact factor: 10.057

9.  Bayesian reconstruction of photon interaction sequences for high-resolution PET detectors.

Authors:  Guillem Pratx; Craig S Levin
Journal:  Phys Med Biol       Date:  2009-08-04       Impact factor: 3.609

10.  Statistical LOR estimation for a high-resolution dMiCE PET detector.

Authors:  Kyle M Champley; Thomas K Lewellen; Lawrence R MacDonald; Robert S Miyaoka; Paul E Kinahan
Journal:  Phys Med Biol       Date:  2009-10-07       Impact factor: 3.609

View more
  2 in total

1.  Optimization of the Energy Window for PETbox4, a Preclinical PET Tomograph With a Small Inner Diameter.

Authors:  Z Gu; Q Bao; R Taschereau; H Wang; B Bai; A F Chatziioannou
Journal:  IEEE Trans Nucl Sci       Date:  2014-06-01       Impact factor: 1.679

2.  Analytical study of the effect of the system geometry on photon sensitivity and depth of interaction of positron emission mammography.

Authors:  Pablo Aguiar; Cristina Lois
Journal:  J Oncol       Date:  2012-09-20       Impact factor: 4.375

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.