Marco Agrifoglio1, Sara Filippini, Maurizio Roberto, Marco Zanobini, Luisa Gregorini, Francesco Alamanni.
Abstract
We report a case of redo aortic prosthesis replacement for a severe paravalvular leak (PVL) in a man operated with continuous suture technique 7 years earlier. The severe aortic regurgitation was due to the rupture of the suture. In spite of operations to replace malfunctioning heart valves are common procedures and performed all over the world from more than 50 years, there is still an open debate about the most suitable suture technique. In this case report, we'll discuss if the suture technique has a role in preventing or leading complications as severe PVL.Entities:
Mesh:
Year: 2011 PMID: 21513562 PMCID: PMC3095550 DOI: 10.1186/1749-8090-6-60
Source DB: PubMed Journal: J Cardiothorac Surg ISSN: 1749-8090 Impact factor: 1.637
Figure 1The transesophageal echocardiogram shows severe aortic prosthetic regurgitation (white arrow), due to significant paravalvular leakage in the non-coronary aortic sinus.
Figure 2The aortic mechanical prosthesis shows a partial detachment at the level of the non-coronary aortic sinus. The semicontinuous suture was broken (white arrow) and the knot was loose. In the picture only one knot is visible since the last one was cut to make possible the aortic prosthesis explant.