Literature DB >> 16275007

Importance of implant technique on risk of major paravalvular leak (PVL) after St. Jude mechanical heart valve replacement: a report from the Artificial Valve Endocarditis Reduction Trial (AVERT).

Lars Englberger1, Hartzell V Schaff, W R Eric Jamieson, Elizabeth D Kennard, Kyung A Im, Richard Holubkov, Thierry P Carrel.   

Abstract

OBJECTIVE: To examine risk factors for major paravalvular leak (PVL) events after mechanical heart valve replacement.
METHODS: We analyzed outcome of 807 patients randomized into the Artificial Valve Endocarditis Reduction Trial (AVERT). The mean follow-up time was 30.6 months and 21 major PVL events were reported. Three additional major PVL events associated with endocarditis were excluded from analysis. All baseline medical history variables, as well as operative parameters (including use of pledgets and suture technique) were examined using Cox regression.
RESULTS: Major PVL was reported after 11 aortic, 9 mitral, and 1 double valve replacement. 6/404 (1.5%) patients with conventional valves experienced a major PVL event versus 15/403 (3.7%) in the Silzone group. 10/172 (5.8%) patients with valve suture technique without pledgets experienced a major PVL event versus 11/635 (1.7%) patients with pledgets. Final multivariable model showed that only suture technique without pledgets (p=0.005) was an independent significant risk factor for major PVL events. Silzone cuff showed a strong trend (p=0.055).
CONCLUSIONS: Suture technique without pledgets is an independent significant risk factor for major PVL events. In this study, use of pledgets during valve replacement had a protective effect against subsequent paravalvular leak, supporting the use of buttress reinforcement for valve suture. The use of Silzone cuff, although not statistically significant, showed a strong trend as a risk factor.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16275007     DOI: 10.1016/j.ejcts.2005.09.014

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  13 in total

1.  Safety and Feasibility of a Nonpledgeted Suture Technique for Heart Valve Replacement.

Authors:  Patrick G Chan; Ernest G Chan; Laura Seese; Ibrahim Sultan; Arman Kilic; Thomas G Gleason; Danny Chu
Journal:  JAMA Surg       Date:  2019-03-01       Impact factor: 14.766

2.  Use of a nonpledgeted suture technique is safe and efficient for aortic valve replacement.

Authors:  Damien J LaPar; Gorav Ailawadi; Castigliano M Bhamidipati; Mansher Singh; David Dare; John A Kern; Irving L Kron
Journal:  J Thorac Cardiovasc Surg       Date:  2010-05-20       Impact factor: 5.209

3.  Multicentre, propensity-matched study to evaluate long-term impact of implantation technique in isolated aortic valve replacement on mortality and incidence of redo surgery.

Authors:  Yama Haqzad; Mahmoud Loubani; Mubarak Chaudhry; Priyadharshanan Ariyaratnam; Norman Briffa
Journal:  Interact Cardiovasc Thorac Surg       Date:  2016-02-17

Review 4.  Update in Paravalvular Leak Closure.

Authors:  Aken Desai; John C Messenger; Robert Quaife; John Carroll
Journal:  Curr Cardiol Rep       Date:  2021-07-16       Impact factor: 2.931

Review 5.  Paravalvular Leaks-From Diagnosis to Management.

Authors:  Samuel Bernard; Evin Yucel
Journal:  Curr Treat Options Cardiovasc Med       Date:  2019-11-14

6.  Pathogenesis of paravalvular leakage as a complication occurring in the late phase after surgery.

Authors:  Fumiaki Kuwabara; Akihiko Usui; Yoshimori Araki; Yuji Narita; Shinichi Mizutani; Hideki Oshima; Yuichi Ueda
Journal:  J Artif Organs       Date:  2011-04-20       Impact factor: 1.731

7.  Very-long-term outcomes of mechanical valves in mitral position focusing on valve-related complications.

Authors:  Gaku Uchino; Hirohisa Murakami; Nobuhiko Mukohara; Hiroshi Tanaka; Yoshikatsu Nomura; Shunsuke Miyahara; Motoharu Kawashima; Jun Fujisue; Shuto Tonoki
Journal:  Interact Cardiovasc Thorac Surg       Date:  2022-07-09

8.  Non infective severe aortic paravalvular leakage 7 years after surgery: the role of suture technique.

Authors:  Marco Agrifoglio; Sara Filippini; Maurizio Roberto; Marco Zanobini; Luisa Gregorini; Francesco Alamanni
Journal:  J Cardiothorac Surg       Date:  2011-04-23       Impact factor: 1.637

9.  Outcomes of nonpledgeted horizontal mattress suture technique for mitral valve replacement.

Authors:  Gun Jik Kim; Jong Tae Lee; Young Ok Lee; Joon Young Cho; Tak-Hyuk Oh
Journal:  Korean J Thorac Cardiovasc Surg       Date:  2014-12-05

10.  Comparative survival and role of STS score in aortic paravalvular leak after SAVR or TAVR: a retrospective study from the USA.

Authors:  Ratnasari Padang; Mahmoud Ali; Kevin L Greason; Christopher G Scott; Manasawee Indrabhinduwat; Charanjit S Rihal; Mackram F Eleid; Vuyisile T Nkomo; Patricia A Pellikka; Sorin V Pislaru
Journal:  BMJ Open       Date:  2018-12-09       Impact factor: 2.692