| Literature DB >> 21469190 |
James G Pipe1, Nicholas R Zwart, Eric A Aboussouan, Ryan K Robison, Ajit Devaraj, Kenneth O Johnson.
Abstract
A novel center-out 3D trajectory for sampling magnetic resonance data is presented. The trajectory set is based on a single Fermat spiral waveform, which is substantially undersampled in the center of k-space. Multiple trajectories are combined in a "stacked cone" configuration to give very uniform sampling throughout a "hub," which is very efficient in terms of gradient performance and uniform trajectory spacing. The fermat looped, orthogonally encoded trajectories (FLORET) design produces less gradient-efficient trajectories near the poles, so multiple orthogonal hub designs are shown. These multihub designs oversample k-space twice with orthogonal trajectories, which gives unique properties but also doubles the minimum scan time for critical sampling of k-space. The trajectory is shown to be much more efficient than the conventional stack of cones trajectory, and has nearly the same signal-to-noise ratio efficiency (but twice the minimum scan time) as a stack of spirals trajectory. As a center-out trajectory, it provides a shorter minimum echo time than stack of spirals, and its spherical k-space coverage can dramatically reduce Gibbs ringing.Mesh:
Year: 2011 PMID: 21469190 DOI: 10.1002/mrm.22918
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668