Platelet activation in ovines undergoing sham surgery or implant of the second generation PediaFlow pediatric ventricular assist device.

Carl A Johnson1, Peter D Wearden, Ergin Kocyildirim, Timothy M Maul, Joshua R Woolley, Sang-Ho Ye, Elise M Strickler, Harvey S Borovetz, William R Wagner.   

Abstract

The PediaFlow pediatric ventricular assist device (VAD) is a magnetically levitated turbodynamic pump under development for circulatory support of small children with a targeted flow rate range of 0.3-1.5 L/min. As the design of this device is refined, ensuring high levels of blood biocompatibility is essential. In this study, we characterized platelet activation during the implantation and operation of a second generation prototype of the PediaFlow VAD (PF2) and also performed a series of surgical sham studies to examine purely surgical effects on platelet activation. In addition, a newly available monoclonal antibody was characterized and shown to be capable of quantifying ovine platelet activation. The PF2 was implanted in three chronic ovine experiments of 17, 30, and 70 days, while surgical sham procedures were performed in five ovines with 30-day monitoring. Blood biocompatibility in terms of circulating activated platelets was measured by flow cytometric assays with and without exogenous agonist stimulation. Platelet activation following sham surgery returned to baseline in approximately 2 weeks. Platelets in PF2-implanted ovines returned to baseline activation levels in all three animals and showed an ability to respond to agonist stimulation. Late-term platelet activation was observed in one animal corresponding with unexpected pump stoppages related to a manufacturing defect in the percutaneous cable. The results demonstrated encouraging platelet biocompatibility for the PF2 in that basal platelet activation was achieved early in the pump implant period. Furthermore, this first characterization of the effect of a major cardiothoracic procedure on temporal ovine platelet activation provides comparative data for future cardiovascular device evaluation in the ovine model.
© 2011, Copyright the Authors. Artificial Organs © 2011, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21463346      PMCID: PMC3178345          DOI: 10.1111/j.1525-1594.2010.01124.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  29 in total

1.  Evaluation of platelet function by flow cytometry.

Authors:  A D Michelson; M R Barnard; L A Krueger; A L Frelinger; M I Furman
Journal:  Methods       Date:  2000-07       Impact factor: 3.608

2.  Long-term use of a left ventricular assist device for end-stage heart failure.

Authors:  E A Rose; A C Gelijns; A J Moskowitz; D F Heitjan; L W Stevenson; W Dembitsky; J W Long; D D Ascheim; A R Tierney; R G Levitan; J T Watson; P Meier; N S Ronan; P A Shapiro; R M Lazar; L W Miller; L Gupta; O H Frazier; P Desvigne-Nickens; M C Oz; V L Poirier
Journal:  N Engl J Med       Date:  2001-11-15       Impact factor: 91.245

3.  Heart disease and stroke statistics--2010 update: a report from the American Heart Association.

Authors:  Donald Lloyd-Jones; Robert J Adams; Todd M Brown; Mercedes Carnethon; Shifan Dai; Giovanni De Simone; T Bruce Ferguson; Earl Ford; Karen Furie; Cathleen Gillespie; Alan Go; Kurt Greenlund; Nancy Haase; Susan Hailpern; P Michael Ho; Virginia Howard; Brett Kissela; Steven Kittner; Daniel Lackland; Lynda Lisabeth; Ariane Marelli; Mary M McDermott; James Meigs; Dariush Mozaffarian; Michael Mussolino; Graham Nichol; Véronique L Roger; Wayne Rosamond; Ralph Sacco; Paul Sorlie; Véronique L Roger; Randall Stafford; Thomas Thom; Sylvia Wasserthiel-Smoller; Nathan D Wong; Judith Wylie-Rosett
Journal:  Circulation       Date:  2009-12-17       Impact factor: 29.690

Review 4.  In vivo preclinical anticoagulation regimens after implantation of ventricular assist devices.

Authors:  Diyar Saeed; Kiyotaka Fukamachi
Journal:  Artif Organs       Date:  2009-07       Impact factor: 3.094

5.  Biocompatibility assessment of the first generation PediaFlow pediatric ventricular assist device.

Authors:  Carl A Johnson; Stijn Vandenberghe; Amanda R Daly; Joshua R Woolley; Shaun T Snyder; Josiah E Verkaik; Sang-Ho Ye; Harvey S Borovetz; James F Antaki; Peter D Wearden; Marina V Kameneva; William R Wagner
Journal:  Artif Organs       Date:  2011-01       Impact factor: 3.094

6.  Changes in platelet activation associated with left ventricular assist system placement.

Authors:  H Matsubayashi; D R Fastenau; J A McIntyre
Journal:  J Heart Lung Transplant       Date:  2000-05       Impact factor: 10.247

7.  Second INTERMACS annual report: more than 1,000 primary left ventricular assist device implants.

Authors:  James K Kirklin; David C Naftel; Robert L Kormos; Lynne W Stevenson; Francis D Pagani; Marissa A Miller; Karen L Ulisney; J Timothy Baldwin; James B Young
Journal:  J Heart Lung Transplant       Date:  2010-01       Impact factor: 10.247

8.  Extended mechanical circulatory support with a continuous-flow rotary left ventricular assist device.

Authors:  Francis D Pagani; Leslie W Miller; Stuart D Russell; Keith D Aaronson; Ranjit John; Andrew J Boyle; John V Conte; Roberta C Bogaev; Thomas E MacGillivray; Yoshifumi Naka; Donna Mancini; H Todd Massey; Leway Chen; Charles T Klodell; Juan M Aranda; Nader Moazami; Gregory A Ewald; David J Farrar; O Howard Frazier
Journal:  J Am Coll Cardiol       Date:  2009-07-21       Impact factor: 24.094

9.  Advanced heart failure treated with continuous-flow left ventricular assist device.

Authors:  Mark S Slaughter; Joseph G Rogers; Carmelo A Milano; Stuart D Russell; John V Conte; David Feldman; Benjamin Sun; Antone J Tatooles; Reynolds M Delgado; James W Long; Thomas C Wozniak; Waqas Ghumman; David J Farrar; O Howard Frazier
Journal:  N Engl J Med       Date:  2009-11-17       Impact factor: 91.245

10.  Animal model development for the Penn State pediatric ventricular assist device.

Authors:  Elizabeth L Carney; J Brian Clark; John L Myers; Rebecca Peterson; Ronald P Wilson; William J Weiss
Journal:  Artif Organs       Date:  2009-10-22       Impact factor: 3.094

View more
  10 in total

1.  The national heart, lung, and blood institute pediatric circulatory support program: a summary of the 5-year experience.

Authors:  J Timothy Baldwin; Harvey S Borovetz; Brian W Duncan; Mark J Gartner; Robert K Jarvik; William J Weiss
Journal:  Circulation       Date:  2011-03-22       Impact factor: 29.690

2.  Biocompatibility Assessment of the CentriMag-Novalung Adult ECMO Circuit in a Model of Acute Pulmonary Hypertension.

Authors:  Venkat Shankarraman; Ergin Kocyildirim; Salim E Olia; Marina V Kameneva; Ryan J Dzadony; Timothy M Maul; Marc A Simon; Hunter C Champion; William R Wagner; Christian A Bermudez
Journal:  ASAIO J       Date:  2014 Jul-Aug       Impact factor: 2.872

3.  Modern Outcomes of Mechanical Circulatory Support as a Bridge to Pediatric Heart Transplantation.

Authors:  Brody Wehman; Kristen A Stafford; Gregory J Bittle; Zachary N Kon; Charles F Evans; Keshava Rajagopal; Nicholas Pietris; Sunjay Kaushal; Bartley P Griffith
Journal:  Ann Thorac Surg       Date:  2016-02-22       Impact factor: 4.330

4.  Platelet activation after implantation of the Levitronix PediVAS in the ovine model.

Authors:  Carl A Johnson; Venkat Shankarraman; Peter D Wearden; Ergin Kocyildirim; Timothy M Maul; John D Marks; J Scott Richardson; Barry N Gellman; Harvey S Borovetz; Kurt A Dasse; William R Wagner
Journal:  ASAIO J       Date:  2011 Nov-Dec       Impact factor: 2.872

5.  Preclinical performance of a pediatric mechanical circulatory support device: The PediaFlow ventricular assist device.

Authors:  Salim E Olia; Peter D Wearden; Timothy M Maul; Venkat Shankarraman; Ergin Kocyildirim; Shaun T Snyder; Patrick M Callahan; Marina V Kameneva; William R Wagner; Harvey S Borovetz; James F Antaki
Journal:  J Thorac Cardiovasc Surg       Date:  2018-04-21       Impact factor: 5.209

6.  Synthesis, characterization, and paclitaxel release from a biodegradable, elastomeric, poly(ester urethane)urea bearing phosphorylcholine groups for reduced thrombogenicity.

Authors:  Yi Hong; Sang-Ho Ye; Anca L Pelinescu; William R Wagner
Journal:  Biomacromolecules       Date:  2012-10-18       Impact factor: 6.988

7.  In Vivo 5 Day Animal Studies of a Compact, Wearable Pumping Artificial Lung.

Authors:  Shalv P Madhani; Brian J Frankowski; Sang-Ho Ye; Greg W Burgreen; William R Wagner; Robert Kormos; Jonathan D'Cunha; William J Federspiel
Journal:  ASAIO J       Date:  2019-01       Impact factor: 2.872

8.  Month-long Respiratory Support by a Wearable Pumping Artificial Lung in an Ovine Model.

Authors:  Ryan A Orizondo; Katelin S Omecinski; Alexandra G May; Vishaal Dhamotharan; Brian J Frankowski; Greg W Burgreen; Sang-Ho Ye; Ergin Kocyildirim; Pablo G Sanchez; Jonathan D'Cunha; William R Wagner; William J Federspiel
Journal:  Transplantation       Date:  2021-05-01       Impact factor: 5.385

9.  CFD-Based Flow Channel Optimization and Performance Prediction for a Conical Axial Maglev Blood Pump.

Authors:  Weibo Yang; Sijie Peng; Weihu Xiao; Yefa Hu; Huachun Wu; Ming Li
Journal:  Sensors (Basel)       Date:  2022-02-19       Impact factor: 3.576

10.  In vivo testing of the low-flow CO2 removal application of a compact, platform respiratory device.

Authors:  Alexandra G May; Ryan A Orizondo; Brian J Frankowski; Sang-Ho Ye; Ergin Kocyildirim; William R Wagner; Jonathan D'Cunha; William J Federspiel
Journal:  Intensive Care Med Exp       Date:  2020-08-17
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.