Literature DB >> 21457494

Pulse pressure variation, stroke volume variation and dynamic arterial elastance.

Raphael Giraud, Nils Siegenthaler, Karim Bendjelid.   

Abstract

Entities:  

Mesh:

Year:  2011        PMID: 21457494      PMCID: PMC3219348          DOI: 10.1186/cc10088

Source DB:  PubMed          Journal:  Crit Care        ISSN: 1364-8535            Impact factor:   9.097


× No keyword cloud information.

We read with interest the recent article by Monge Garcia and colleagues [1]. We have two comments regarding this interesting physiological study. First, although we support their observation [2], their explanations may lead to some confusion. Indeed, the Edwards Vigileo™ system-FloTrac™ sensor calculates stroke volume (SV) using the equation: where σAP is the standard deviation of the arterial pressure (AP) curve and Khi a constant quantifying arterial elastance and vascular resistance [3]. As stroke volume variation (SVV): Which means that SVV (%) ≈ pulse pressure variation (PPV; %) for FloTrac™. With the present mathematical equation, it can be demonstrated that the SVV calculated by FloTrac™ (SVVFT) is not influenced by Khi, which means that SVVFT calculation does not include effective arterial elastance (Ea). The originality of Garcia and colleagues' finding is that they realized that when measuring PPV conventionally [4], the ratio PPV/SVVFT becomes a mirror of a dynamic Ea, as the conventional PPV selected integrates Khi (Figure 1). However, when the present innovative method is used, PPV and SVV should be sampled during the same period of time.
Figure 1

Model of two different arterial elastances (Ea and Ea) on the pulse pressure variation (PVV) relationship curve, with the same stroke volume variation (SVV) value.

Model of two different arterial elastances (Ea and Ea) on the pulse pressure variation (PVV) relationship curve, with the same stroke volume variation (SVV) value. In conclusion, we believe that dynamic Ea predicting arterial pressure response to volume loading in preload-dependent patients is an interesting physiological concept. However, this demonstration cannot be achieved without taking into consideration the impact of PPV/SVV sampling.

Abbreviations

AP: arterial pressure; Ea: arterial elastance; PP: pulse pressure; PPV: pulse pressure variation; SV: stroke volume; SVV: stroke volume variation.

Competing interests

The authors declare that they have no competing interests.
  4 in total

1.  When to recalibrate the PiCCO? From a physiological point of view, the answer is simple.

Authors:  K Bendjelid
Journal:  Acta Anaesthesiol Scand       Date:  2009-05       Impact factor: 2.105

Review 2.  Edwards FloTrac sensor and Vigileo monitor: easy, accurate, reliable cardiac output assessment using the arterial pulse wave.

Authors:  Gerard R Manecke
Journal:  Expert Rev Med Devices       Date:  2005-09       Impact factor: 3.166

3.  Clinical use of respiratory changes in arterial pulse pressure to monitor the hemodynamic effects of PEEP.

Authors:  F Michard; D Chemla; C Richard; M Wysocki; M R Pinsky; Y Lecarpentier; J L Teboul
Journal:  Am J Respir Crit Care Med       Date:  1999-03       Impact factor: 21.405

4.  Dynamic arterial elastance to predict arterial pressure response to volume loading in preload-dependent patients.

Authors:  Manuel Ignacio Monge García; Anselmo Gil Cano; Manuel Gracia Romero
Journal:  Crit Care       Date:  2011-01-12       Impact factor: 9.097

  4 in total
  4 in total

1.  Fact and controversies when assessing fluid responsiveness.

Authors:  Karim Bendjelid
Journal:  J Clin Monit Comput       Date:  2017-04-19       Impact factor: 2.502

2.  Respiratory change in ECG-wave amplitude is a reliable parameter to estimate intravascular volume status.

Authors:  Raphaël Giraud; Nils Siegenthaler; Denis R Morel; Jacques-A Romand; Laurent Brochard; Karim Bendjelid
Journal:  J Clin Monit Comput       Date:  2012-11-02       Impact factor: 2.502

3.  Monitoring dynamic arterial elastance as a means of decreasing the duration of norepinephrine treatment in vasoplegic syndrome following cardiac surgery: a prospective, randomized trial.

Authors:  Pierre-Grégoire Guinot; Osama Abou-Arab; Mathieu Guilbart; Stéphane Bar; Elie Zogheib; Mona Daher; Patricia Besserve; Joseph Nader; Thierry Caus; Said Kamel; Hervé Dupont; Emmanuel Lorne
Journal:  Intensive Care Med       Date:  2017-01-24       Impact factor: 17.440

Review 4.  Functional haemodynamic monitoring.

Authors:  Michael R Pinsky
Journal:  Curr Opin Crit Care       Date:  2014-06       Impact factor: 3.687