Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards.

E De Ponti1, S Morzenti, L Guerra, C Pasquali, M Arosio, V Bettinardi, A Crespi, M C Gilardi, C Messa.   

Abstract

PURPOSE: The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI).
METHODS: Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures.
RESULTS: The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate (k=1) was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%.
CONCLUSIONS: The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE.

Mesh:

Year:  2011        PMID: 21452733     DOI: 10.1118/1.3544655

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Low-dose radiation protocol using 3D mode in a BGO PET/CT.

Authors:  Solange A Nogueira; Renato Dimenstein; Marcelo L Cunha; Jairo Wagner; Marcelo B G Funari; Henrique M Lederman
Journal:  Radiol Med       Date:  2014-06-06       Impact factor: 3.469

2.  New standards for phantom image quality and SUV harmonization range for multicenter oncology PET studies.

Authors:  Go Akamatsu; Naoki Shimada; Keiichi Matsumoto; Hiromitsu Daisaki; Kazufumi Suzuki; Hiroshi Watabe; Keiichi Oda; Michio Senda; Takashi Terauchi; Ukihide Tateishi
Journal:  Ann Nucl Med       Date:  2022-01-14       Impact factor: 2.668

3.  Image quality and variability for routine diagnostic FDG-PET scans in a Japanese community hospital: current status and possibility of improvement.

Authors:  Keiji Shimizu; Seiichi Yamamoto; Keiichi Matsumoto; Megumu Hino; Michio Senda
Journal:  Jpn J Radiol       Date:  2016-04-28       Impact factor: 2.374

4.  An Adaptive Thresholding Method for BTV Estimation Incorporating PET Reconstruction Parameters: A Multicenter Study of the Robustness and the Reliability.

Authors:  M Brambilla; R Matheoud; C Basile; C Bracco; I Castiglioni; C Cavedon; M Cremonesi; S Morzenti; F Fioroni; M Giri; F Botta; F Gallivanone; E Grassi; M Pacilio; E De Ponti; M Stasi; S Pasetto; S Valzano; D Zanni
Journal:  Comput Math Methods Med       Date:  2015-05-19       Impact factor: 2.238

5.  Evaluation of spatial dependence of point spread function-based PET reconstruction using a traceable point-like 22Na source.

Authors:  Taisuke Murata; Kenta Miwa; Noriaki Miyaji; Kei Wagatsuma; Tomoyuki Hasegawa; Keiichi Oda; Takuro Umeda; Takashi Iimori; Yoshitada Masuda; Takashi Terauchi; Mitsuru Koizumi
Journal:  EJNMMI Phys       Date:  2016-10-26

6.  18F-FDG silicon photomultiplier PET/CT: A pilot study comparing semi-quantitative measurements with standard PET/CT.

Authors:  Lucia Baratto; Sonya Young Park; Negin Hatami; Guido Davidzon; Shyam Srinivas; Sanjiv Sam Gambhir; Andrei Iagaru
Journal:  PLoS One       Date:  2017-06-05       Impact factor: 3.240

7.  Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT.

Authors:  Delphine Vandendriessche; Jorge Uribe; Hugo Bertin; Frank De Geeter
Journal:  EJNMMI Phys       Date:  2019-05-10
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.