Proof of principle study of the use of a CMOS active pixel sensor for proton radiography.

Joao Seco1, Nicolas Depauw.   

Abstract

PURPOSE: Proof of principle study of the use of a CMOS active pixel sensor (APS) in producing proton radiographic images using the proton beam at the Massachusetts General Hospital (MGH).
METHODS: A CMOS APS, previously tested for use in s-ray radiation therapy applications, was used for proton beam radiographic imaging at the MGH. Two different setups were used as a proof of principle that CMOS can be used as proton imaging device: (i) a pen with two metal screws to assess spatial resolution of the CMOS and (ii) a phantom with lung tissue, bone tissue, and water to assess tissue contrast of the CMOS. The sensor was then traversed by a double scattered monoenergetic proton beam at 117 MeV, and the energy deposition inside the detector was recorded to assess its energy response. Conventional x-ray images with similar setup at voltages of 70 kVp and proton images using commercial Gafchromic EBT 2 and Kodak X-Omat V films were also taken for comparison purposes.
RESULTS: Images were successfully acquired and compared to x-ray kVp and proton EBT2/X-Omat film images. The spatial resolution of the CMOS detector image is subjectively comparable to the EBT2 and Kodak X-Omat V film images obtained at the same object-detector distance. X-rays have apparent higher spatial resolution than the CMOS. However, further studies with different commercial films using proton beam irradiation demonstrate that the distance of the detector to the object is important to the amount of proton scatter contributing to the proton image. Proton images obtained with films at different distances from the source indicate that proton scatter significantly affects the CMOS image quality.
CONCLUSION: Proton radiographic images were successfully acquired at MGH using a CMOS active pixel sensor detector. The CMOS demonstrated spatial resolution subjectively comparable to films at the same object-detector distance. Further work will be done in order to establish the spatial and energy resolution of the CMOS detector for protons. The development and use of CMOS in proton radiography could allow in vivo proton range checks, patient setup QA, and real-time tumor tracking.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21452699     DOI: 10.1118/1.3496327

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Proton-counting radiography for proton therapy: a proof of principle using CMOS APS technology.

Authors:  G Poludniowski; N M Allinson; T Anaxagoras; M Esposito; S Green; S Manolopoulos; J Nieto-Camero; D J Parker; T Price; P M Evans
Journal:  Phys Med Biol       Date:  2014-05-01       Impact factor: 3.609

2.  Water-equivalent path length calibration of a prototype proton CT scanner.

Authors:  R F Hurley; R W Schulte; V A Bashkirov; A J Wroe; A Ghebremedhin; H F-W Sadrozinski; V Rykalin; G Coutrakon; P Koss; B Patyal
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

3.  A novel proton-integrating radiography system design using a monolithic scintillator detector: experimental studies.

Authors:  Chinmay D Darne; Daniel G Robertson; Fahed Alsanea; Charles-Antoine Collins-Fekete; Sam Beddar
Journal:  Nucl Instrum Methods Phys Res A       Date:  2021-12-16       Impact factor: 1.455

4.  A proton imaging system using a volumetric liquid scintillator: a preliminary study.

Authors:  Chinmay D Darne; Fahed Alsanea; Daniel G Robertson; Fada Guan; Tinsu Pan; David Grosshans; Sam Beddar
Journal:  Biomed Phys Eng Express       Date:  2019-07-12

5.  The impact of secondary fragments on the image quality of helium ion imaging.

Authors:  Lennart Volz; Pierluigi Piersimoni; Vladimir A Bashkirov; Stephan Brons; Charles-Antoine Collins-Fekete; Robert P Johnson; Reinhard W Schulte; Joao Seco
Journal:  Phys Med Biol       Date:  2018-10-02       Impact factor: 3.609

Review 6.  Proton radiography and tomography with application to proton therapy.

Authors:  G Poludniowski; N M Allinson; P M Evans
Journal:  Br J Radiol       Date:  2015-06-04       Impact factor: 3.039

7.  Helium ion beam imaging for image guided ion radiotherapy.

Authors:  M Martišíková; T Gehrke; S Berke; G Aricò; O Jäkel
Journal:  Radiat Oncol       Date:  2018-06-14       Impact factor: 3.481

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.