Literature DB >> 21423057

Optimization of radiation exposure and image quality of the cone-beam O-arm intraoperative imaging system in spinal surgery.

Kasim Abul-Kasim1, Marcus Söderberg, Eufrozina Selariu, Mikael Gunnarsson, Mehrsa Kherad, Acke Ohlin.   

Abstract

STUDY
DESIGN: Retrospective study.
OBJECTIVES: To optimize the radiation doses and image quality for the cone-beam O-arm surgical imaging system in spinal surgery. SUMMARY OF
BACKGROUND: Neurovascular compromise has been reported after screw misplacement during thoracic pedicle screw insertion. The use of O-arm with or without navigation system during spinal surgery has been shown to lower the rate of screw misplacement. The main drawback of such imaging surgical systems is the high radiation exposure.
METHODS: Chest phantom and cadaveric pig spine were examined on the O-arm with different scan settings: 2 were recommended by the O-arm manufacturer (120 kV/320 mAs, and 120 kV/128 mAs), and 3 low-dose settings (80 kV/80 mAs, 80 kV/40 mAs, and 60 kV/40 mAs). The radiation doses were estimated by Monte Carlo calculations. Objective evaluation of image quality included interobserver agreement in the measurement of pedicular width in chest phantom and assessment of screw placement in cadaveric pig spine.
RESULTS: The effective dose/cm for 120 kV/320 mAs scan was 13, 26, and 69 times higher than those delivered with 80 kV/80 mAs, 80 kV/40 mAs, and 60 kV/40 mAs scans, respectively. Images with 60 kV/40 mAs were unreliable. Images with 80 kV/80 mAs were considered reliable with good interobserver agreement when measuring the pedicular width (random error 0.38 mm and intraclass correlation coefficient 0.979) and almost perfect agreement when evaluating the screw placement (κ value 0.86).
CONCLUSIONS: The radiation doses of the O-arm system can be reduced 5 to 13 times without negative impact on image quality with regard to information required for spinal surgery.

Entities:  

Mesh:

Year:  2012        PMID: 21423057     DOI: 10.1097/BSD.0b013e318211fdea

Source DB:  PubMed          Journal:  J Spinal Disord Tech        ISSN: 1536-0652


  10 in total

1.  Radiation dose reduction in thoracic and lumbar spine instrumentation using navigation based on an intraoperative cone beam CT imaging system: a prospective randomized clinical trial.

Authors:  Nathalie Pireau; Virginie Cordemans; Xavier Banse; Nadia Irda; Sébastien Lichtherte; Ludovic Kaminski
Journal:  Eur Spine J       Date:  2017-07-22       Impact factor: 3.134

2.  A novel technique of cervical pedicle screw placement with a pilot screw under the guidance of intraoperative 3D imaging from C-arm cone-beam CT without navigation for safe and accurate insertion.

Authors:  Masahiko Takahata; Katsuhisa Yamada; Iwata Akira; Tsutomu Endo; Hideki Sudo; Hidetoki Yokoyama; Norimasa Iwasaki
Journal:  Eur Spine J       Date:  2018-07-23       Impact factor: 3.134

3.  Radiation exposure to the patients in thoracic and lumbar spine fusion using a new intraoperative cone-beam computed tomography imaging technique: a preliminary study.

Authors:  Ludovic Kaminski; V Cordemans; O Cartiaux; M Van Cauter
Journal:  Eur Spine J       Date:  2017-02-06       Impact factor: 3.134

4.  Evaluation of an Experimentally Designed Stereotactic Guidance System for Determining Needle Entry Point during Uniplanar Fluoroscopy-guided Intervention.

Authors:  Jae Heon Lee; Gye Rok Jeon; Jung Hoon Ro; Gyeong Jo Byoen; Tae Kyun Kim; Kyung Hoon Kim
Journal:  Korean J Pain       Date:  2012-04-04

5.  Pedicle violation and Navigational errors in pedicle screw insertion using the intraoperative O-arm: A preliminary report.

Authors:  Jacob E Mathew; Kelvin Mok; Benoit Goulet
Journal:  Int J Spine Surg       Date:  2013-12-01

6.  A Comparison of Cobb Angle: Standing Versus Supine Images of Late-Onset Idiopathic Scoliosis.

Authors:  Ludvig Vavruch; Hans Tropp
Journal:  Pol J Radiol       Date:  2016-06-10

7.  Metal artifacts in intraoperative O-arm CBCT scans.

Authors:  Juha I Peltonen; Touko Kaasalainen; Mika Kortesniemi
Journal:  BMC Med Imaging       Date:  2021-01-06       Impact factor: 1.930

8.  Intraoperative CT-guided navigation versus fluoroscopy for percutaneous pedicle screw placement in 192 patients: a comparative analysis.

Authors:  Giuseppe La Rocca; Edoardo Mazzucchi; Fabrizio Pignotti; Luigi Aurelio Nasto; Gianluca Galieri; Alessandro Olivi; Vincenzo De Santis; Pierluigi Rinaldi; Enrico Pola; Giovanni Sabatino
Journal:  J Orthop Traumatol       Date:  2022-09-01

9.  Intraoperative 2D C-arm and 3D O-arm in children: a comparative phantom study.

Authors:  M Prod'homme; M Sans-Merce; N Pitteloud; J Damet; P Lascombes
Journal:  J Child Orthop       Date:  2018-10-01       Impact factor: 1.548

10.  Clinical application of 3D printing technology to the surgical treatment of atlantoaxial subluxation in small breed dogs.

Authors:  Hiroaki Kamishina; Taku Sugawara; Kohei Nakata; Hidetaka Nishida; Naoko Yada; Toru Fujioka; Yoshihiko Nagata; Akio Doi; Naoyuki Konno; Fujio Uchida; Sadatoshi Maeda
Journal:  PLoS One       Date:  2019-05-03       Impact factor: 3.240

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.