Icare rebound tonometry in children with known and suspected glaucoma.

Meghan S Flemmons1, Ya-Chuan Hsiao, Jacqueline Dzau, Sanjay Asrani, Sarah Jones, Sharon F Freedman.   

Abstract

BACKGROUND: Accurate intraocular pressure (IOP) measurement, important in managing pediatric glaucoma, often presents challenges. The Icare rebound tonometer shows promise for screening healthy children and has been reported comparable with Goldmann applanation in adults with glaucoma. The purpose of this study was to evaluate the Icare tonometer against Goldmann applanation for clinic IOP measurement in pediatric glaucoma.
METHODS: This was a prospective study comparing Icare versus Goldmann tonometry in pediatric glaucoma. Children with known or suspected glaucoma were recruited from scheduled clinic visits. IOP was measured with the Icare tonometer by a clinician and subsequently measured with Goldmann applanation tonometry (GAT) by a different single masked clinician.
RESULTS: A total of 71 eyes of 71 children with known or suspected glaucoma were included. IOP by GAT ranged from 9 to 36 mm Hg. Icare readings ranged from 11 to 44 mm Hg. Mean difference between Icare and GAT was 2.3 ± SD 3.7 mm Hg, p < 0.0001. Icare IOPs were within ± 3 mm Hg of GAT in 63%. Icare IOPs were ≥GAT IOPs in 75%. The following factors were not associated with Icare IOPs greater than GAT: child's age, glaucoma diagnosis, strabismus, nystagmus, central corneal thickness, Icare instrument-reported reliability, number of glaucoma surgeries or medications, corneal abnormalities, and visual acuity.
CONCLUSIONS: IOP by Icare tonometry was within 3 mm Hg of IOP by GAT in 63% and greater than GAT in 75%. This device may be reasonable to estimate IOP in selected children with known or suspected glaucoma whose IOP cannot otherwise be obtained in clinic; however, correlation of Icare IOPs with clinical findings must continue to be considered in each case.
Copyright © 2011 American Association for Pediatric Ophthalmology and Strabismus. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21419676     DOI: 10.1016/j.jaapos.2010.11.022

Source DB:  PubMed          Journal:  J AAPOS        ISSN: 1091-8531            Impact factor:   1.220


  13 in total

1.  Pediatric ptosis as a sign of treatable autonomic dysfunction.

Authors:  Lara Phillips; David Robertson; Mark R Melson; Emily M Garland; Karen M Joos
Journal:  Am J Ophthalmol       Date:  2013-04-24       Impact factor: 5.258

2.  Obesity, Blood Pressure, and Intraocular Pressure: A Cross-Sectional Study in Italian Children.

Authors:  Claudia Pileggi; Rosa Papadopoli; Caterina De Sarro; Carmelo Giuseppe Angelo Nobile; Maria Pavia
Journal:  Obes Facts       Date:  2021-04-01       Impact factor: 3.942

3.  Intraocular Pressure Measurement in Childhood Glaucoma under Standardized General Anaesthesia: The Prospective EyeBIS Study.

Authors:  Alicja Strzalkowska; Nina Pirlich; Julia V Stingl; Alexander K Schuster; Jasmin Rezapour; Felix M Wagner; Justus Buse; Esther M Hoffmann
Journal:  J Clin Med       Date:  2022-05-18       Impact factor: 4.964

4.  Finite element method for estimation of applanation force and to study the influence of intraocular pressure of eye on tonometry.

Authors:  R B Bharathi; Rakshath G Poojary; Gopalakrishna K Prabhu; Ramesh S Ve
Journal:  Int Ophthalmol       Date:  2022-06-04       Impact factor: 2.029

5.  The utility of rebound tonometer in late elderly subjects.

Authors:  Rei Sakata; Jiro Numaga
Journal:  Clin Interv Aging       Date:  2014-01-22       Impact factor: 4.458

6.  Comprasion of ICare rebound tonometer and Goldmann applanation tonometer in high myopia.

Authors:  Halil Huseyin Cagatay; Metin Ekinci; Zeliha Yazar; Gokcen Gokce; Erdinc Ceylan
Journal:  ScientificWorldJournal       Date:  2014-10-14

7.  A Comparison of Rebound to Indentation Tonometry in Supine Sedated Children with Glaucoma.

Authors:  Dora H AlHarkan; Fatemah T Al-Shamlan; Deepak P Edward; Arif O Khan
Journal:  Middle East Afr J Ophthalmol       Date:  2016 Apr-Jun

8.  Intraocular pressure in children after congenital heart surgery: A single-center study.

Authors:  Sunali Goyal; Paul H Phillips; Lamonda A Corder; Michael J Robertson; Xiomara Garcia; Michael L Schmitz; Punkaj Gupta
Journal:  Ann Pediatr Cardiol       Date:  2017 Sep-Dec

9.  Comparison of handheld rebound tonometry with Goldmann applanation tonometry in children with glaucoma: a cohort study.

Authors:  Annegret Hella Dahlmann-Noor; Renata Puertas; Shenille Tabasa-Lim; Ahmed El-Karmouty; Mustafa Kadhim; Nicholas Kloster Wride; Amanda Lewis; Dawn Grosvenor; Poornima Rai; Maria Papadopoulos; John Brookes; Catey Bunce; Peng Tee Khaw
Journal:  BMJ Open       Date:  2013-04-02       Impact factor: 2.692

10.  The measurement of intraocular pressure over positive soft contact lenses by rebound tonometry.

Authors:  Fabrizio Zeri; Mario De Cusatis; Luigi Lupelli; Peter Graham Swann
Journal:  J Optom       Date:  2015-10-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.