We evaluated the performance of a laboratory-developed multiplex real-time reverse transcription-PCR assay (LDT rRT-PCR), the Centers for Disease Control and Prevention (CDC) 2009 H1N1 rRT-PCR protocol using the LightCycler 480 II, the multiplex xTAG Respiratory Virus Panel (xTAG RVP), and rapid immunodiagnostic testing (RIDT) using the BinaxNOW Influenza A & B to detect 2009 H1N1 with 426 nasopharyngeal swab specimens during the 2009 H1N1 pandemic. The specificity of the methods tested was ≥98%, and the individual test sensitivities were RIDT at 42.3% [95% confidence interval (CI), 31.4-54.0], LDT rRT-PCR at 98.9% (95% CI, 92.9-99.9), CDC 2009 H1N1 rRT-PCR at 78.2% (95% CI, 67.8-86.0), and xTAG RVP at 93.1% (95% CI, 85.0-97.2). A negative RIDT result should not be used to make decisions with respect to treatment or infection prevention. rRT-PCR is the preferred first-line diagnostic test for detecting 2009 H1N1 influenza A.Entities:
Mesh:
Year: 2011 PMID: 21392924 PMCID: PMC7173341 DOI: 10.1016/j.diagmicrobio.2011.01.010
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803
Analytic performance of rRT-PCR, Luminex xTAG RVP, and BinaxNOW Influenza A & B RIDTa
| Influenza A 2009 H1N1 test method | Sensitivity | Specificity | PPV | NPV | Prevalence | TP | TN | FP | FN |
|---|---|---|---|---|---|---|---|---|---|
| BinaxNOW Influenza A & B RIDT | 42.3 (31.4–54.0) | 98.0 (95.7–99.1) | 82.5 (66.6–92.1) | 88.3 (84.6–91.3) | 18.3 (14.8–22.4) | 33 | 341 | 7 | 45 |
| LDT SOIA rRT-PCR | 98.9 (92.9–99.9) | 98.9 (97.0–99.7) | 95.6 (88.4–98.6) | 99.7 (98.2–100) | 19.4 (15.9–23.4) | 86 | 358 | 4 | 1 |
| CDC 2009 H1N1 LC480 rRT-PCR | 78.2 (67.8–86.0) | 99.7 (98.2–100) | 98.6 (91.1–99.9) | 95.0 (92.2–96.9) | 19.4 (15.9–23.5) | 68 | 360 | 1 | 19 |
| Luminex xTAG RVP | 93.1 (85.0–97.2) | 99.5 (97.8–99.9) | 97.6 (90.8–99.6) | 98.4 (96.3–99.3) | 19.4 (15.9–23.5) | 81 | 359 | 2 | 6 |
Positivity is defined as requiring 2 or more positive results obtained by rRT-PCR or xTAG RVP.
95% Confidence intervals.
PPV = positive predictive value; NPV = negative predictive value; TP = true positive; TN = true negative; FP = false positive; FN = false negative.