Literature DB >> 21375098

Can accurate ultrasonographic measurement of the optic nerve sheath diameter (a non-invasive measure of intracranial pressure) be taught to novice operators in a single training session?

D W Potgieter1, A Kippin, F Ngu, C McKean.   

Abstract

Novices can be quickly and easily taught to ultrasonographically measure optic nerve sheath diameter as a noninvasive surrogate for intracranial pressure. One intensive care unit senior medical officer, one junior intensive care unit registrar, one junior neurosurgery trainee and two registered nurses, all ultrasound novices, were given a four-hour workshop on how to measure the optic nerve sheath diameter with ultrasound. One week later they had the accuracy of their technique assessed by recording their measurements for both eyes of 12 healthy volunteers. Each eye was measured three times and the average and 'best' measurements were determined. The median, 5th and 95th centiles and standard deviation for intra-observer variation were calculated. Median, 5th and 95th centiles for inter-observer variation were obtained by comparing each of the novice operator's measurements with those of an experienced sonographer Median intra-observer difference was < or = 0.27 mm, standard deviation < or = 0.46 mm and 95th centiles < or = 1.19 mm for the five novice operators. Comparing the average of three measurements for each eye, median inter-observer difference for all operators was < or = 0.42 mm and 95th centile values < or = 1.15 mm. Using the single 'best' measurement for each eye, median inter-observer difference was < or = 0.3 mm and 95th centiles < or = 1.43 mm. Ultrasonographic optic nerve sheath diameter measurement can be taught to novice operators in a four-hour workshop with variation comparable to published results collated from 'experts'.

Mesh:

Year:  2011        PMID: 21375098     DOI: 10.1177/0310057X1103900116

Source DB:  PubMed          Journal:  Anaesth Intensive Care        ISSN: 0310-057X            Impact factor:   1.669


  19 in total

1.  Non-invasive methods of estimating intracranial pressure.

Authors:  Jamie B Rosenberg; Ariel L Shiloh; Richard H Savel; Lewis A Eisen
Journal:  Neurocrit Care       Date:  2011-12       Impact factor: 3.210

2.  Ultrasonography Assessments of Optic Nerve Sheath Diameter as a Noninvasive and Dynamic Method of Detecting Changes in Intracranial Pressure.

Authors:  Li-Juan Wang; Li-Min Chen; Ying Chen; Li-Yang Bao; Nan-Nan Zheng; Yu-Zhi Wang; Ying-Qi Xing
Journal:  JAMA Ophthalmol       Date:  2018-03-01       Impact factor: 7.389

3.  Association between optic nerve sheath diameter and mortality in patients with severe traumatic brain injury.

Authors:  Mypinder S Sekhon; Paul McBeth; Jie Zou; Lu Qiao; Leif Kolmodin; William R Henderson; Steve Reynolds; Donald E G Griesdale
Journal:  Neurocrit Care       Date:  2014-10       Impact factor: 3.210

4.  The relationship between transorbital ultrasound measurement of the optic nerve sheath diameter (ONSD) and invasively measured ICP in children : Part I: repeatability, observer variability and general analysis.

Authors:  Llewellyn C Padayachy; Vaishali Padayachy; Ushma Galal; Rebecca Gray; A Graham Fieggen
Journal:  Childs Nerv Syst       Date:  2016-09-20       Impact factor: 1.475

5.  Comparison of the Effects of Propofol and Sevoflurane Anesthesia on Optic Nerve Sheath Diameter in Robot-Assisted Laparoscopic Gynecology Surgery: A Randomized Controlled Trial.

Authors:  Jung Eun Kim; Seong Yoon Koh; In-Jung Jun
Journal:  J Clin Med       Date:  2022-04-12       Impact factor: 4.964

6.  Correlation of measurement of optic nerve sheath diameter with ultrasound and magnetic resonance imaging.

Authors:  Shakti Bedanta Mishra; Afzal Azim; Syed Nabeel Muzaffar
Journal:  Indian J Crit Care Med       Date:  2015-10

7.  Effect of scuba-diving on optic nerve and sheath diameters.

Authors:  Masoud Mehrpour; Narges Sadat Shams-Hosseini; Saeed Rezaali
Journal:  Med J Islam Repub Iran       Date:  2014-09-07

8.  Optimal optic nerve sheath diameter threshold for the identification of elevated opening pressure on lumbar puncture in a Chinese population.

Authors:  Lijuan Wang; Liangshu Feng; Yan Yao; Yuzhi Wang; Ying Chen; Jiachun Feng; Yingqi Xing
Journal:  PLoS One       Date:  2015-02-09       Impact factor: 3.240

9.  Correlation of measurement of optic nerve sheath diameter with ultrasound and magnetic resonance imaging.

Authors:  Munta Kartik; Manimala S Rao
Journal:  Indian J Crit Care Med       Date:  2016-02

10.  Estimating the accuracy of optic nerve sheath diameter measurement using a pocket-sized, handheld ultrasound on a simulation model.

Authors:  Garrett G R J Johnson; Frederick A Zeiler; Bertram Unger; Gregory Hansen; Dimitrios Karakitsos; Lawrence M Gillman
Journal:  Crit Ultrasound J       Date:  2016-11-10
View more