A novel respiratory tracking system for smart-gated PET acquisition.

S A Nehmeh1, A A Haj-Ali, C Qing, C Stearns, H Kalaigian, S Kohlmyer, H Schoder, A Y Ho, S M Larson, J L Humm.   

Abstract

PURPOSE: In this study, the authors validated a novel respiratory tracking device, the multidimensional respiratory tracking (MDRT) system, that was designed to assist in correcting for respiratory motion in PET/CT images. The authors also investigated a novel PET acquisition technique, smart gating (SG), that enables to acquire motion-free PET data prospectively, with minimum user interference and with no additional postprocessing of the PET data.
METHODS: MDRT uses visual tracking techniques to track simultaneously the two-dimensional (in the vertical plane) motion of multiple fiducial markers using a standard video camera. A threshold window is set at the breathing amplitude of interest using the MDRT GUI. A trigger is generated at a rate of 250 Hz as long as the breathing signal is within the threshold window. The triggers are fed into the PET scanner to initialize one single bin of a gated acquisition every 4 ms. No triggers are delivered as the breathing signal drifts outside the threshold window. Consequently, PET data are acquired only whenever the breathing signal is confined within the amplitude threshold window, thus resulting into a motion-free image set. The accuracy of MDRT in tracking the breathing signal was assessed (1) by comparing the period of an oscillating phantom, as measured by MDRT, to that measured with a photogate timer and (2) by comparing the MDRT output to that of the real-time position management (RPM) in ten patients. The SG PET/CT acquisition was validated in phantoms and in two stereotactic body radiosurgery (SBRS) lung DIBH-PET/CT patients.
RESULTS: MDRT was in agreement with the photogate timer in determining the period of motion to less than 2%. The percent errors between MDRT and RPM in the positions of the peaks and troughs of the ten patients' breathing signals were within 10%. In phantoms, SG technique enables to correct for motion-induced artifacts in the PET images and improve the accuracy of PET quantitation. For the SBRS application, in one patient, the patient's CT lesion was not detected in the corresponding clinical PET images, while it exhibited an SUV of 1.8 in the DIBH image set. In the second patient, DIBH-PET images showed an improved PET-to-CT spatial matching and a 52% increase in the lesion SUV.
CONCLUSIONS: MDRT has been shown to be accurate in tracking breathing motion and assisted in implementing a smart-gating PET acquisition technique that allowed to acquire prospectively motion-free PET images.

Entities:  

Mesh:

Year:  2011        PMID: 21361220     DOI: 10.1118/1.3523100

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  The impact of audio-visual biofeedback on 4D PET images: results of a phantom study.

Authors:  Jaewon Yang; Tokihiro Yamamoto; Byungchul Cho; Youngho Seo; Paul J Keall
Journal:  Med Phys       Date:  2012-02       Impact factor: 4.071

2.  Quantification of the thorax-to-abdomen breathing ratio for breathing motion modeling.

Authors:  Benjamin M White; Tianyu Zhao; James Lamb; Jeffrey D Bradley; Daniel A Low
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

3.  The potential of positron emission tomography for intratreatment dynamic lung tumor tracking: a phantom study.

Authors:  Jaewon Yang; Tokihiro Yamamoto; Samuel R Mazin; Edward E Graves; Paul J Keall
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

4.  Respiratory gating in cardiac PET: Effects of adenosine and dipyridamole.

Authors:  Martin Lyngby Lassen; Thomas Rasmussen; Thomas E Christensen; Andreas Kjær; Philip Hasbak
Journal:  J Nucl Cardiol       Date:  2016-09-07       Impact factor: 5.952

5.  Assessing and accounting for the impact of respiratory motion on FDG uptake and viable volume for liver lesions in free-breathing PET using respiration-suspended PET images as reference.

Authors:  Guang Li; C Ross Schmidtlein; Irene A Burger; Carole A Ridge; Stephen B Solomon; John L Humm
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

6.  Imaging and dosimetric errors in 4D PET/CT-guided radiotherapy from patient-specific respiratory patterns: a dynamic motion phantom end-to-end study.

Authors:  S R Bowen; M J Nyflot; C Herrmann; C M Groh; J Meyer; S D Wollenweber; C W Stearns; P E Kinahan; G A Sandison
Journal:  Phys Med Biol       Date:  2015-04-17       Impact factor: 3.609

7.  Assessment of pulmonary melanoma metastases with 18F-FDG PET/CT: which PET-negative patients require additional tests for definitive staging?

Authors:  Marius E Mayerhoefer; Helmut Prosch; Christian J Herold; Michael Weber; Georgios Karanikas
Journal:  Eur Radiol       Date:  2012-06-01       Impact factor: 5.315

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.