Literature DB >> 21361218

EPID-guided 3D dose verification of lung SBRT.

M Aristophanous1, J Rottmann, L E Court, R I Berbeco.   

Abstract

PURPOSE: To investigate the feasibility of utilizing tumor tracks from electronic portal imaging device (EPID) images taken during treatment to verify the delivered dose.
METHODS: The proposed method is based on a computation of the delivered fluence by utilizing the planned fluence and the tumor motion track for each field. A phantom study was designed to assess the feasibility of the method. The CIRS dynamic thorax phantom was utilized with a realistic soft resin tumor, modeled after a real patient tumor. The dose calculated with the proposed method was compared to direct measurements taken with 15 metal oxide semiconductor field effect transistors (MOSFETs) inserted in small fissures made in the tumor model. The phantom was irradiated with the tumor static and moved with different range of motions and setup errors. EPID images were recorded throughout all deliveries and the tumor model was tracked post-treatment with in-house developed software. The planned fluence for each field was convolved with the tumor motion tracks to obtain the delivered fluence. Utilizing the delivered fluence from each field, the delivered dose was calculated. The estimated delivered dose was compared to the dose directly measured with the MOSFETs. The feasibility of the proposed method was also demonstrated on a real lung cancer patient, treated with stereotactic body radiotherapy.
RESULTS: The calculation of delivered dose with the delivered fluence method was in good agreement with the MOSFET measurements, with average differences ranging from 0.8% to 8.3% depending on the proximity of a dose gradient. For the patient treatment, the planned and delivered dose volume histograms were compared and verified the overall good coverage of the target volume.
CONCLUSIONS: The delivered fluence method was applied successfully on phantom and clinical data and its accuracy was evaluated. Verifying each treatment fraction may enable correction strategies that can be applied during the course of treatment to ensure the desired dose coverage.

Entities:  

Mesh:

Year:  2011        PMID: 21361218     DOI: 10.1118/1.3532821

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Registration of clinical volumes to beams-eye-view images for real-time tracking.

Authors:  Jonathan H Bryant; Joerg Rottmann; John H Lewis; Pankaj Mishra; Paul J Keall; Ross I Berbeco
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

2.  3D delivered dose assessment using a 4DCT-based motion model.

Authors:  Weixing Cai; Martina H Hurwitz; Christopher L Williams; Salam Dhou; Ross I Berbeco; Joao Seco; Pankaj Mishra; John H Lewis
Journal:  Med Phys       Date:  2015-06       Impact factor: 4.071

Review 3.  Review of Real-Time 3-Dimensional Image Guided Radiation Therapy on Standard-Equipped Cancer Radiation Therapy Systems: Are We at the Tipping Point for the Era of Real-Time Radiation Therapy?

Authors:  Paul J Keall; Doan Trang Nguyen; Ricky O'Brien; Pengpeng Zhang; Laura Happersett; Jenny Bertholet; Per R Poulsen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2018-04-14       Impact factor: 7.038

4.  The markerless lung target tracking AAPM Grand Challenge (MATCH) results.

Authors:  Marco Mueller; Per Poulsen; Rune Hansen; Wilko Verbakel; Ross Berbeco; Dianne Ferguson; Shinichiro Mori; Lei Ren; John C Roeske; Lei Wang; Pengpeng Zhang; Paul Keall
Journal:  Med Phys       Date:  2021-12-29       Impact factor: 4.071

5.  A method of dose reconstruction for moving targets compatible with dynamic treatments.

Authors:  Per Rugaard Poulsen; Mai Lykkegaard Schmidt; Paul Keall; Esben Schjodt Worm; Walther Fledelius; Lone Hoffmann
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

6.  Beam's-eye-view imaging during non-coplanar lung SBRT.

Authors:  Stephen S F Yip; Joerg Rottmann; Ross I Berbeco
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

7.  Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs.

Authors:  Yinghui Li; Jinhan Zhu; Jinping Shi; Lixin Chen; Xiaowei Liu
Journal:  PLoS One       Date:  2019-06-28       Impact factor: 3.240