Beam's-eye-view imaging during non-coplanar lung SBRT.

Stephen S F Yip1, Joerg Rottmann1, Ross I Berbeco1.   

Abstract

PURPOSE: Beam's-eye-view (BEV) imaging with an electronic portal imaging device (EPID) can be performed during lung stereotactic body radiation therapy (SBRT) to monitor the tumor location in real-time. Image quality for each patient and treatment field depends on several factors including the patient anatomy and the gantry and couch angles. The authors investigated the angular dependence of automatic tumor localization during non-coplanar lung SBRT delivery.
METHODS: All images were acquired at a frame rate of 12 Hz with an amorphous silicon EPID. A previously validated markerless lung tumor localization algorithm was employed with manual localization as the reference. From ten SBRT patients, 12 987 image frames of 123 image sequences acquired at 48 different gantry-couch rotations were analyzed. δ was defined by the position difference of the automatic and manual localization.
RESULTS: Regardless of the couch angle, the best tracking performance was found in image sequences with a gantry angle within 20° of 250° (δ = 1.40 mm). Image sequences acquired with gantry angles of 150°, 210°, and 350° also led to good tracking performances with δ = 1.77-2.00 mm. Overall, the couch angle was not correlated with the tracking results. Among all the gantry-couch combinations, image sequences acquired at (θ = 30°, ϕ = 330°), (θ = 210°, ϕ = 10°), and (θ = 250°, ϕ = 30°) led to the best tracking results with δ = 1.19-1.82 mm. The worst performing combinations were (θ = 90° and 230°, ϕ = 10°) and (θ = 270°, ϕ = 30°) with δ > 3.5 mm. However, 35% (17/48) of the gantry-couch rotations demonstrated substantial variability in tracking performances between patients. For example, the field angle (θ = 70°, ϕ = 10°) was acquired for five patients. While the tracking errors were ≤1.98 mm for three patients, poor performance was found for the other two patients with δ ≥ 2.18 mm, leading to average tracking error of 2.70 mm. Only one image sequence was acquired for all other gantry-couch rotations (δ = 1.18-10.29 mm).
CONCLUSIONS: Non-coplanar beams with gantry-couch rotation of (θ = 30°, ϕ = 330°), (θ = 210°, ϕ = 10°), and (θ = 250°, ϕ = 30°) have the highest accuracy for BEV lung tumor localization. Additionally, gantry angles of 150°, 210°, 250°, and 350° also offer good tracking performance. The beam geometries (θ = 90° and 230°, ϕ = 10°) and (θ = 270°, ϕ = 30°) are associated with substantial automatic localization errors. Overall, lung tumor visibility and tracking performance were patient dependent for a substantial number of the gantry-couch angle combinations studied.

Entities:  

Mesh:

Year:  2015        PMID: 26632035      PMCID: PMC4636508          DOI: 10.1118/1.4934824

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  24 in total

1.  Stereotactic body radiation therapy of early-stage non-small-cell lung carcinoma: phase I study.

Authors:  Ronald C McGarry; Lech Papiez; Mark Williams; Tia Whitford; Robert D Timmerman
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-08-22       Impact factor: 7.038

2.  Treatment planning of stereotactic radiotherapy for solitary lung tumor.

Authors:  Kenji Takayama; Yasushi Nagata; Yoshiharu Negoro; Takashi Mizowaki; Takashi Sakamoto; Masato Sakamoto; Tetsuya Aoki; Shinsuke Yano; Sachiko Koga; Masahiro Hiraoka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-04-01       Impact factor: 7.038

Review 3.  Radiotherapy of mobile tumors.

Authors:  Steve B Jiang
Journal:  Semin Radiat Oncol       Date:  2006-10       Impact factor: 5.934

4.  Clinical outcomes of a phase I/II study of 48 Gy of stereotactic body radiotherapy in 4 fractions for primary lung cancer using a stereotactic body frame.

Authors:  Yasushi Nagata; Kenji Takayama; Yukinori Matsuo; Yoshiki Norihisa; Takashi Mizowaki; Takashi Sakamoto; Masato Sakamoto; Michihide Mitsumori; Keiko Shibuya; Norio Araki; Shinsuke Yano; Masahiro Hiraoka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-09-19       Impact factor: 7.038

5.  Optimal number of beams for stereotactic body radiotherapy of lung and liver lesions.

Authors:  Ruiguo Liu; John M Buatti; Terese L Howes; John Dill; Joseph M Modrick; Sanford L Meeks
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-08-14       Impact factor: 7.038

6.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

7.  Small-volume image-guided radiotherapy using hypofractionated, coplanar, and noncoplanar multiple fields for patients with inoperable Stage I nonsmall cell lung carcinomas.

Authors:  Shin-Ichi Fukumoto; Hiroki Shirato; Shinichi Shimzu; Shigeaki Ogura; Rikiya Onimaru; Kei Kitamura; Koichi Yamazaki; Kazuo Miyasaka; Masaharu Nishimura; Hirotoshi Dosaka-Akita
Journal:  Cancer       Date:  2002-10-01       Impact factor: 6.860

8.  Stereotactic single-dose radiotherapy of stage I non-small-cell lung cancer (NSCLC).

Authors:  Holger Hof; Klaus K Herfarth; Marc Münter; Angelika Hoess; Johann Motsch; Michael Wannenmacher; Jürgen Debus J
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-06-01       Impact factor: 7.038

9.  Stereotactic radiotherapy for primary lung cancer and pulmonary metastases: a noninvasive treatment approach in medically inoperable patients.

Authors:  Joern Wulf; Ulrich Haedinger; Ulrich Oppitz; Wibke Thiele; Gerd Mueller; Michael Flentje
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-09-01       Impact factor: 7.038

10.  Geometric uncertainty of 2D projection imaging in monitoring 3D tumor motion.

Authors:  Yelin Suh; Sonja Dieterich; Paul J Keall
Journal:  Phys Med Biol       Date:  2007-05-18       Impact factor: 3.609

View more
  3 in total

1.  Technical Note: 3D localization of lung tumors on cone beam CT projections via a convolutional recurrent neural network.

Authors:  Chuang Wang; Margie Hunt; Lei Zhang; Andreas Rimner; Ellen Yorke; Michael Lovelock; Xiang Li; Tianfang Li; Gig Mageras; Pengpeng Zhang
Journal:  Med Phys       Date:  2020-01-28       Impact factor: 4.071

2.  Design and validation of a MV/kV imaging-based markerless tracking system for assessing real-time lung tumor motion.

Authors:  Pengpeng Zhang; Margie Hunt; Arina B Telles; Hai Pham; Michael Lovelock; Ellen Yorke; Guang Li; Laura Happersett; Andreas Rimner; Gig Mageras
Journal:  Med Phys       Date:  2018-11-13       Impact factor: 4.071

3.  Technical Note: Synthetic treatment beam imaging for motion monitoring during spine SBRT treatments - a phantom study.

Authors:  Tianfang Li; Feifei Li; Weixing Cai; Pengpeng Zhang; Xiang Li
Journal:  Med Phys       Date:  2020-12-07       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.