Anoop K Shetty1, Simon G Duckett, Julian Bostock, Eric Rosenthal, C Aldo Rinaldi.
Abstract
AIMS: Problems with implanting a left ventricular (LV) lead during cardiac resynchronization therapy (CRT) procedures are not uncommon and may occur for a variety of reasons including phrenic nerve stimulation (PNS) and high capture thresholds. We aimed to perform successful CRT in patients with previous LV lead problems using the multiple pacing configurations available with the St Jude Quartet model 1458Q quadripolar LV lead to overcome PNS or high capture thresholds. METHODS ANDEntities:
Mesh:
Year: 2011 PMID: 21343237 PMCID: PMC3120132 DOI: 10.1093/europace/eur032
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Patient characteristics pre-Quartet lead implant
| Case | Age | Sex | ECG | NYHA | LVEF (%) | Cardiomyopathy aetiology | Patient characteristics |
|---|---|---|---|---|---|---|---|
| 1 | 72 | F | Sinus rhythm; LBBB; QRSd 140 ms | III | 25 | Non-ischaemic | Two previous attempts at CRT with two different bipolar leads but PNS or high capture thresholds in all target veins |
| 2 | 60 | M | Sinus rhythm; LBBB; QRSd 160 ms | III | 20 | Non-ischaemic | Previous attempt at CRT in referring center but implanter unable to pass guide catheter into CS. Only one suitable CS vein found at our center but PNS seen using conventional bipolar lead |
| 3 | 24 | M | AF; LBBB (paced); QRSd 160 ms | III | 20 | Non-ischaemic | Previous AVSD repair and dual-chamber pacemaker for severe sinus and AV node disease. At CRT-P upgrade only two suitable veins seen and PNS found in one of these. LV lead placed in the other CS branch (without PNS at implant) but PNS found at follow-up and LV lead switched off |
| 4 | 64 | F | Sinus rhythm; LBBB; QRSd 170 ms | III | 35 | Non-ischaemic | One previous attempt at CRT with two different bipolar leads but PNS or high capture thresholds in all target veins |
M = male; F = female.
Pacing parameter data
| Pacing parameter | |
|---|---|
| Mean implant LV capture threshold at 0.5 ms pulse width | 2.0 ± 1.6 V |
| Pre-discharge LV capture threshold at 0.5 ms pulse width | 1.4 ± 0.7 V |
| LV capture threshold at latest (8.5 ± 5 weeks) follow-up at 0.5 ms pulse width | 1.3 ± 0.2 V |
| LV lead impedance at implant | 1057 ± 254 Ω |
| Pre-discharge LV lead impedance | 912.5 ± 394 Ω |
| LV lead impedance at latest (8.5 ± 5 weeks) follow-up | 872 ± 387 Ω |