| Literature DB >> 21319350 |
Sun Young Cho1, Min Jin Kim, Jin-Tae Suh, Hee Joo Lee.
Abstract
PURPOSE: PCR is widely used for rapidly and accurately detecting Mycobacterium species. The purpose of this study was to assess the diagnostic performance of three real-time PCR kits and evaluate the concordance with two older PCR methods.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21319350 PMCID: PMC3051208 DOI: 10.3349/ymj.2011.52.2.301
Source DB: PubMed Journal: Yonsei Med J ISSN: 0513-5796 Impact factor: 2.759
The Kinds of Mycobacterium Species in NTM DNA Specimens
NTM, non-tuberculous mycobacterium.
The Composition of Clinical Specimens
MTB, Mycobacterium tuberculosis; SP, sputum; BW, bronchial washing.
*Blood, body fluid, urine and tissue.
†5 MTB PCR results show complete concordant results; in-house PCR protocol, the COBAS Amplicor MTB, the COBAS TaqMan MTB, the AdvanSure TB/NTM real-time PCR, the Real-Q M. tuberculosis kit.
Fig. 1In-house MTB PCR. Lane 1, size marker; Lane 2, internal control; Lane 7, positive; Lane 3-6, 8, negative. MTB, Mycobacterium tuberculosis.
The MTB PCR Results for 91 Clinical Specimens
MTB, Mycobacterium tuberculosis; AFB, acid fast bacilli.
*COBAS Amplicor MTB (Roche, Indianapolis, IN, USA).
†COBAS TaqMan MTB (Roche, USA).
‡AdvanSure TB/NTM real-time PCR kit (LG Lifescience, Seoul, Korea).
§Real-Q M. tuberculosis kit (Biosewoom, Seoul, Korea).
∥One of the 91 clinical samples had no culture and AFB stain results.
Discrepant Cases for Both the Clinical Specimens and the DNA Samples Provided
AFB, acid fast bacilli; TP, true positive; FP, false positive; NTM, non-tuberculous mycobacterium; MIC, M. intracellulare; MLF, M. lentiflavum.
*It was confirmed to M. abscessus.
†AdvanSure Mycobacteria Genotyping Chip (LG Lifescience, Seoul, Korea).
Fig. 2Two discrepant cases reaffirmed by sequencing and using the AdvanSure Mycobacteria Genotyping Chip (LG Life science, Seoul, Korea). (A) For 37 DNA samples provided, the AdvanSure kit detected all mycobacterial species and Real-Q kit showed a negative result for M. lentiflavum (target sequence; IS6110 region). (B) In contrast with the one false positive result from the AdvanSure kit, the Real-Q missed one true positive M. intracellulare in the clinical samples (target sequence; 16S-23S internal transcribed spacer region).