| Literature DB >> 21314992 |
Philippe Gillet1, Jessica Maltha, Veerle Hermans, Raffaella Ravinetto, Cathrien Bruggeman, Jan Jacobs.
Abstract
BACKGROUND: The present study assessed malaria RDT kits for adequate and correct packaging, design and labelling of boxes and components. Information inserts were studied for readability and accuracy of information.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21314992 PMCID: PMC3045995 DOI: 10.1186/1475-2875-10-39
Source DB: PubMed Journal: Malar J ISSN: 1475-2875 Impact factor: 2.979
Number of RDTs (n = 42) with inadequacies in malaria RDT boxes, device packages, devices, buffer vials and package inserts*
| Items considered to be inadequate | Number (%) | |
|---|---|---|
| Materials: plastic bag or simple cardboard (not humidity-resistant) | 9 (21.4) | |
| No labels, no printed information or labels not humidity-resistant | 6 (14.3) | |
| Differences in name on device packaging, device, buffer and information insert | 27 (64.3) | |
| No EC-REP mentioned on CE labelled RDTs, although required (n = 25) | 3 (12.0) | |
| RDT kit's name nor additional information refer to intended use | 3 (7.1) | |
| RDT kit's name incorrectly refers to | 4 (13.8) | |
| Kit components not displayed | 26 (62%) | |
| Essential information lacking: expiry date, numbers of tests included, storage conditions | 12 (28.6) | |
| Capillary sampling system (lancet and alcohol swap) not included or not optionally included | 24 (57%) | |
| Blood transfer system (capillary, pipette or tube) not included | 3 (7.1%) | |
| Material not humidity-resistant | 4 (9.5) | |
| No desiccant or desiccant without saturation indicator | 18 (42.9) | |
| Essential information lacking: expiry date, lot number, test kit name | 9 (21.4) | |
| No warning label "do not swallow" on desiccant | 6 (14.3) | |
| Space for sample identification too small or not writable with standard pen (felt pen needed) | 40 (95.2) | |
| No or incomplete RDT name on the device | 29 (69.0) | |
| No reading label or simultaneous presence two reading labels consisting of symbols only | 6 (14.3) | |
| Buffer vial not leak proof | 2 (4.8) | |
| Label does not stick well to the vial, prints are not humidity-resistant (n = 40) | 16 (40.0) | |
| Essential information lacking: expiry date, lot number, storage conditions, correct RDT kit's name (n = 41) | 24 (58.5) | |
| No instructions included on how to pierce the buffer vial dropper (n = 15) | 5 (33.3) | |
| Absence of date of release or version number | 20 (47.6) | |
| Identity of target antigens not clearly mentioned | 2 (4.8) | |
| No referral to biosafety precautions (gloves, safe waste disposal, etc.) | 18 (42.9) | |
| Major differences between depicted and real device (n = 40) | 8 (20.0) | |
| Use of figures with unrealistic colours ( | 4 (9.5) | |
| No data on test characteristics (sensitivity, specificity) | 8 (19.0) | |
* Total number of RDT kits = 42 unless otherwise stated.
Number of RDT information inserts (n = 40) addressing critical steps in procedure and interpretation
| Bring the RDT device and buffer to room temperature | 32 (80.0) |
| Check the integrity of the device package | 9 (22.5) |
| Check expiry date | 27 (67.5) |
| Use the device immediately after opening | 28 (70.0) |
| Place the device on a level surface | 0 (0.0) |
| Check the desiccant for signs of exposure to humidity | 11 (27.5) |
| Write down sample identification | 3 (7.5) |
| Wipe finger with alcohol | 26 (65.0) |
| Allow the finger to dry before pricking | 12 (30.0) |
| Hold the transfer device (loop, straw) vertical | 8 (20.0) |
| Hold the buffer vial vertical | 12 (30.0) |
| Do not to use another buffer than the one provided with the kit | 9 (22.5) |
| Use an adequate light source for reading | 3 (7.5) |
| All possible line combinations for invalid test results are mentioned | 12 (30.0) |
| All possible test line combinations for positive test results are mentioned | 31 (77.5) |
| Interpretation of a faint test line as positive is mentioned | 8 (20.0) |
| Causes of false negative results are mentioned, in particular low parasite densities | 11 (27.5) |
| Causes of false positive results are mentioned, | 3 (7.5) |
| Persistence of HRP-2 is mentioned | 19 (47.5) |
| To repeat the test in case of a negative RDT result and persistent suspicion of malaria is mentioned | 1 (2.5) |
Overview of the RDT kits evaluated in the present study
| HRP-2 | 7 | 4 | 5 | 7 | |
| pan-pLDH | 1 | 1 | 0 | 1 | |
| Pv-pLDH | 1 | 1 | 0 | 1 | |
| HRP-2, pan-pLDH | 11 | 5 | 7 | 9 | |
| HRP-2, aldolase‡ | 5 | 5* | 5 | 5 | |
| HRP-2, Pv-pLDH | 4 | 1 | 1 | 2 | |
| Pf-pLDH, pan-pLDH | 6 | 5 | 6 | 6 | |
| HRP-2, Pv-pLDH, pan-pLDH | 7 | 3 | 5 | 6 | |
* One three-band test (HRP-2, aldolase) is FDA approved.
† WHO list of ISO:13485:2003 certified manufacturers and their RDT products (list of known commercially available antigen detecting malaria RDTs) [12].
‡ One of these RDTs used both aldolase and pan-pLDH as pan-malarial antigen.
Figure 1Four-band RDT. The allocated place for writing sample identification is too small. The grid at the left hand may be confused with a sample well. There are two different reading labels at each side of the reading window, of which the lower one is printed on a label that is not well fixed.
Figure 2Example of RDT cassettes. Most of the cassettes have separated wells for sample and buffer application. There is no uniform labelling of the wells: different characters (e.g. "S", "A") are used randomly for the sample well, buffer well and combined sample/buffer well. The reading labels are indicated with acronyms, characters or numbers.
Figure 3Interpretation section of the information insert and cassette of a . The real device has a single sample/buffer whereas the depicted one displays separate wells. The characters used for the reading label on the illustration are inverted compared to the real device.
Figure 4Interpretation section of the information insert and cassette of a . Shape and labels of wells and reading window are different between the real and the depicted device. Characters are embedded in the plastic housing and poorly discernable. The text is correct and complete (even the prozone effect and how to deal with it) but less readable (Flesch-Kincaid grade level 9.1).
Figure 5Information insert of a three-band RDT, procedure section. There is a discrepancy between the reading time mentioned in the text compared to that showed on the illustration.
Figure 6Typographic features of RDT information inserts (n = 40): font sizes and line spacing.
Figure 7Example of typography used in RDT information inserts: the package insert on the top is user-unfriendly (font size 6, line spacing 0,5, close letter type, average number of words per line 21, Flesch-Kincaid grade 9,5). The package insert on the background, from the same company but for another RDT uses a better typography (font size 8, line spacing 2, open letter type, average number of words per line 14), but the readability is still elevated (Flesch-Kincaid grade 9,8).
Figure 8Readability of the RDT information inserts (n = 40) and job aids (n = 7) expressed as Flesch-Kincaid grade level.
Figure 9Interpretation section of the information insert of a three-band . For invalid tests, only the absence of all lines is mentioned, not the possibility of a visible test line without a control line. The invalid results are not depicted. The print quality is poor, no colours are used for the control and test lines.
Figure 10Illustration depicting sampling of capillary blood. The health care worker's hand is depicted without gloves. The simultaneous presence of English and French text may be difficult for a non-experienced reader.
Figure 11CE-label displayed on a RDT kit box. The shape and relative dimensions of the characters do not comply with the requirements as mentioned in the EC Directive 98/79 (depicted in the insert, upper right corner). There is no authorized representative (EC-REP) affixed, although required.
Figure 12Example of the lateral side of a RDT box: all essential information is printed on a single side of the box, contributing to clear storage.