Abstract
BACKGROUND: Malaria rapid diagnostic tests (RDTs) are kits that generally include 20 to 25 test strips or cassettes, but only a single buffer vial. In field settings, laboratory staff occasionally uses saline, distilled water (liquids for parenteral drugs dilution) or tap water as substitutes for the RDT kit's buffer to compensate for the loss of a diluent bottle. The present study assessed the effect of buffer substitution on the RDT results.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20650003 PMCID: PMC3224932 DOI: 10.1186/1475-2875-9-215
Source DB: PubMed Journal: Malar J ISSN: 1475-2875 Impact factor: 2.979
Numbers of RDT brands showing false positive test lines when run with blood of malaria free subjects and substitute liquids
| RDT format | Numbers of different RDT brands assessed | Numbers of RDTs brands showing false positive test lines when run with substitute liquid* | |||
|---|---|---|---|---|---|
| Saline | Distilled water | Tap water | Any substitute liquid | ||
| Two band | 6 | 3 (1) | 3 (1) | 4 (1) | 4 (1) |
| Three band | 17 | 12 (9) | 11 (10) | 12 (9) | 12 (10) |
| Four band | 4 | 1 (0) | 3 (2) | 4 (2) | 4 (2) |
| Total | 27 | 16 (10) | 17 (13) | 20 (12) | 20 (13) |
*Numbers refer to the different RDT brands for which false positive test lines were visible in at least one sample. Between brackets: numbers of RDT brands for which false positive lines were visible in the samples of at least three subjects.
Numbers of false-positive test lines for the different Plasmodium antigens of the RDTs when run with blood of malaria free subjects and substitute liquids
| RDT target antigen | Numbers of RDT brands (n = 27) detecting the target | Total numbers of tests performed with each substitute* | Numbers of false positive test lines when run with substitute liquid | ||
|---|---|---|---|---|---|
| Saline | Distilled water | Tap water | |||
| HRP-2 | 21 | 98 | 22 | 28 | 36 |
| Pf-pLDH | 5 | 25 | 8 | 9 | 9 |
| Pv-pLDH | 9 | 45 | 12 | 21 | 20 |
| Pan-pLDH | 14 | 70 | 30 | 25 | 33 |
| Aldolase | 3 | 11 | 1 | 5 | 5 |
*Each RDT brand was assessed with blood of five malaria-free subjects except one HRP-2-based kit and one HRP-2/aldolase-based kit that were assessed with two and one samples respectively.
Figure 1Example of buffer substitution's effect on a three-band RDT. Three-band RDT cassettes run with blood of a malaria free subject and the RDT kit's dedicated buffer (1), injection water (2), saline (3), tap water (4) and when run with injection water in absence of blood sample (5). Cassettes "Pf" and "Pv" refer to Plasmodium falciparum or Plasmodium vivax positive samples run with the RDT kit's dedicated buffer. All cassettes show regular control lines, cassette 1 shows the expected result (no test line visible), cassettes 2, 3 and 5 show false positive pan-pLDH and Pf-HRP2 lines, cassette 4 shows a false positive pan-pLDH line, and cassettes "Pf" and "Pv" show the expected results (positive for pan-pLDH and Pf-HRP-2 lines or positive for pan-pLDH line respectively). In the cassette 3, the false positive reaction is partially masked by the background.
Figure 2Example of buffer substitution's effect on a two-band RDT. Cassettes of a two-band RDT run with blood of a malaria free subject and the RDT kit's dedicated buffer (1), injection water (2), saline (3), tap water (4) and when run with injection water in absence of blood sample (5). Cassettes "Pf" and "Pv" refer to Plasmodium falciparum or Plasmodium vivax positive samples run with the RDT kit's dedicated buffer. All cassettes show regular control lines, cassette 1 shows the expected result (no test line visible), cassettes 2, 3, 4 and 5 show false positive Pf-HRP-2 line, cassettes "Pf" and "Pv" show the expected results (positive for Pf-HRP-2 line or no test line visible respectively).
Numbers of false-positive test lines for the different Plasmodium antigens of the RDTs when run with distilled water in absence of blood sample
| RDT target antigen | Numbers of RDT brands (n = 27) detecting the target | Numbers of false positive test lines | Numbers of false positive test lines according to test line intensity | |||
|---|---|---|---|---|---|---|
| Faint | Weak | Medium | Strong | |||
| HRP-2 | 21 | 11 | 2 | 5 | 3 | 1 |
| Pf-pLDH | 5 | 2 | 1 | 1 | 0 | 0 |
| Pv-pLDH | 9 | 1 | 1 | 0 | 0 | 0 |
| Pan-pLDH | 14 | 6 | 1 | 4 | 1 | 0 |
| Aldolase | 3 | 2 | 1 | 1 | 0 | 0 |