Diagnostic accuracy of point-of-care testing for diabetic ketoacidosis at emergency-department triage: {beta}-hydroxybutyrate versus the urine dipstick.

Sanjay Arora1, Sean O Henderson, Theodore Long, Michael Menchine.   

Abstract

OBJECTIVE: In the emergency department, hyperglycemic patients are screened for diabetic ketoacidosis (DKA) via a urine dipstick. In this prospective study, we compared the test characteristics of point-of-care β-hydroxybutyrate (β-OHB) analysis with the urine dipstick. RESEARCH DESIGN AND METHODS: Emergency-department patients with blood glucose ≥250 mg/dL had urine dipstick, chemistry panel, venous blood gas, and capillary β-OHB measurements. DKA was diagnosed according to American Diabetes Association criteria.
RESULTS: Of 516 hyperglycemic subjects, 54 had DKA. The urine dipstick had a sensitivity of 98.1% (95% CI 90.1-100), a specificity of 35.1% (30.7-39.6), a positive predictive value of 15% (11.5-19.2), and a negative predictive value of 99.4% (96.6-100) for DKA. Using the manufacturer-suggested cutoff of >1.5 mmol/L, β-OHB had a sensitivity of 98.1% (90.1-100), a specificity of 78.6% (74.5-82.2), a positive predictive value of 34.9% (27.3-43), and a negative predictive value of 99.7% (98.5-100) for DKA.
CONCLUSIONS: Point-of-care β-OHB and the urine dipstick are equally sensitive for detecting DKA (98.1%). However, β-OHB is more specific (78.6 vs. 35.1%), offering the potential to significantly reduce unnecessary DKA work-ups among hyperglycemic patients in the emergency department.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21307381      PMCID: PMC3064039          DOI: 10.2337/dc10-1844

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


Rapid and accurate identification of patients with diabetic ketoacidosis (DKA) is critical but is complicated by the fact that DKA only affects a small percentage of the total number of patients with hyperglycemia (1,2). Current clinical practice in the emergency department is to screen hyperglycemic patients with a urine dipstick for ketones and conduct a laboratory evaluation for DKA on patients whose results screen positive (3). Although the urine dipstick is easily accessible, inexpensive, rapid, and has excellent sensitivity for DKA, its poor specificity (estimated at <50%) results in a large number of false-positive tests and unnecessary work-ups (4). This has led the American Diabetes Association (ADA) to discourage using the urine dipstick and encourage using serum ketones for DKA screening (5). The development of point-of-care β-hydroxybutyrate (β-OHB) devices has made following these ADA recommendations feasible in the emergency department (6). The goal of this prospective study was to compare the test characteristics of a point-of-care β-OHB meter with the urine dipstick for screening for DKA at emergency-department triage.

RESEARCH DESIGN AND METHODS

Setting

This prospective, observational study was conducted at a large, urban emergency department. Institutional review board approval was obtained, and study participants provided written informed consent. The cohort studied represents a convenience sample of emergency-department patients with triage capillary blood glucose ≥250 mg/dL enrolled on 320 individual days over a 2-year period based on research-assistant availability. Subjects were excluded if they were 1) critically ill, 2) in police custody, 3) suffering from acute psychosis, or 4) unable to give informed consent.

Study procedures and data collection

Included patients had a urine dipstick (Siemens Multistix 10SG) for ketones (recorded as positive or negative), a serum chemistry panel, a venous blood gas to determine serum pH, and a point-of-care capillary β-OHB measurement (Precision Xtra meter; Medisense/Abbott Laboratories). DKA was defined according to the following ADA criteria: serum glucose ≥250 mg/dL; anion gap >10 mmol/L; carbon dioxide ≤18 mmol/L; and pH ≤7.30 (7). Subjects’ race, ethnicity, age, sex, previous history of diabetes, and insulin use were recorded.

Sample size considerations

On the basis of previous literature, we assumed that the true sensitivity of β-OHB for detecting DKA was between 98 and 100% and determined that a sample including 54 cases of DKA would confirm this point estimate with a confidence limit no lower than 90% (8–12). The study was stopped after the 54th DKA patient was enrolled.

Statistical analysis

Data were analyzed using Stata version 10.0 (StataCorp, College Station, TX). Diagnostic test characteristics with 95% CIs were calculated. According to the manufacturer’s guidelines, point-of-care β-OHB ≤1.5 mmol/L was considered negative and >1.5 mmol/L was considered positive. The difference in specificity for β-OHB and urine dipstick ketones was assessed using the McNemar test with a Yates correction (13). We explored the optimal capillary β-OHB cutoff by creating a receiver-operator curve and analyzing diagnostic test characteristics at various cutoffs.

RESULTS

Participants

A total of 859 patients were screened, 616 provided informed consent, and 516 (83.7%) had all data elements necessary for data analysis (excluded patients: 49 with no urine dipstick, 30 with no serum pH via venous blood gas, and 21 with no serum chemistry panel). In our final sample, 54 of 516 subjects (10.5%) met the ADA criteria for DKA. See Table 1 for sample characteristics.
Table 1

Demographic and laboratory characteristics of the study sample

DKANo DKA
n (%)54 (10.5)462 (89.5)
Age (years)41 (30–48)48 (40–57)
Female (%)27.835.3
Race/ethnicity (%)
 Non-Hispanic white03.1
 African American16.713.6
 Hispanic79.280.2
 Asian4.22.6
 Other0.50
Admitted (%)95.235.9
Newly diagnosed diabetes (%)25.914.6
Insulin treated (%)56.644.5
Venous pH7.19 (7.10–7.25)7.39 (7.36–7.41)
β-OHB (mmol/L)0.3 (0.2–1.2)4.9 (3.7–5.6)
HCO3 (mmol/L)8 (6–11)23 (21–26)
Glucose (mg/dL)678 (448–955)409 (328–409)
Anion gap (mmol/L)29 (26–34)13 (11–17)
Positive urine dipstick ketones (%)98.164.9

Data are median (interquartile range), unless otherwise indicated.

Demographic and laboratory characteristics of the study sample Data are median (interquartile range), unless otherwise indicated.

Diagnostic accuracy

At the manufacturer-suggested cutoff of >1.5 mmol/L, the sensitivity of capillary β-OHB and the urine dipstick for DKA were identical at 98.1% (95% CI 90.1–100). However, the specificity of β-OHB was significantly higher than that of the urine dipstick (78.6 [74.5–82.2] vs. 35.1% [30.7–39.6]; P < 0.01). Positive and negative predictive values of capillary β-OHB were 34.9 (27.3–43) and 99.7% (98.5–100), respectively. Positive and negative predictive values of the urine dipstick were 15 (11.5–19.2) and 99.4% (96.6–100), respectively. Test characteristics of β-OHB were stable across a wide range of potential cutoffs (Table 1). The receiver-operator curve suggested that the optimal β-OHB cutoff is >2 mmol/L, where sensitivity remains at 98.1% but specificity improves to 82.3% (Supplementary Data).

CONCLUSIONS

Approximately one-fourth of emergency-department patients have diabetes, and almost half have poor glycemic control (2). Common clinical emergency-department practice is to consider the diagnosis of DKA in all patients presenting with blood glucose ≥250 mg/dL regardless of the reason for the visit. Because emergency-department overcrowding has stretched resources to the breaking point, there is increasing emphasis on developing, testing, and using the most rapid, accurate, and easy-to-use DKA screening tool available. The results of this study show that the specificity of capillary point-of-care β-OHB is superior to the urine dipstick. Two retrospective emergency-department studies have shown that capillary point-of-care β-OHB is 100% sensitive for DKA (8,9). Charles et al. (10) likewise found that β-OHB was 100% sensitive in a prospective study of seven DKA patients. Naunheim et al. (11) reported a β-OHB sensitivity for DKA of 98% in a prospective sample of 160 patients (57 with DKA). Our study is the first to prospectively compare the test characteristics of point-of-care β-OHB against the urine dipstick in a cohort of emergency-department patients. Using β-OHB in lieu of the urine dipstick to screen our sample could have reduced full laboratory work-ups for DKA by 56.9%. Our study has several limitations. First, we did not provide specific training to the nurses who interpreted the urine dipsticks. We felt this was unnecessary because they 1) are Clinical Laboratory Improvement Amendment certified, 2) are familiar and comfortable with this process, and 3) do so regularly in routine clinical practice. Second, the urine dipstick was considered only negative or positive, and consideration of trace/small ketones versus moderate/large ketones may impact the findings. Third, we did not send blood samples to the laboratory for an enzymatic analysis to correlate the values measured by the point-of-care β-OHB meter because, despite early concerns, current-generation β-OHB meters have been shown to be highly accurate compared with laboratory enzymatic analysis (6,14). Fourth, although there are many ways to define DKA, we chose to use the ADA definition because this is commonly used in DKA literature, allowing for direct comparison between our results and previous reports. In conclusion, point-of-care capillary β-OHB testing at a cutoff of >1.5 mmol/L was equally sensitive (98.1%) but markedly more specific (78.6 vs. 35.1%) than the urine dipstick for the detection of DKA in our prospective sample of emergency-department patients with hyperglycemia. Utilizing point-of-care β-OHB testing has the potential to substantially reduce comprehensive laboratory evaluations for DKA among hyperglycemic emergency-department patients.
  13 in total

Review 1.  Hyperglycemic crises in diabetes.

Authors:  Abbas E Kitabchi; Guillermo E Umpierrez; Mary Beth Murphy; Eugene J Barrett; Robert A Kreisberg; John I Malone; Barry M Wall
Journal:  Diabetes Care       Date:  2004-01       Impact factor: 19.112

2.  Sample size calculation should be performed for design accuracy in diagnostic test studies.

Authors:  Antoine Flahault; Michel Cadilhac; Guy Thomas
Journal:  J Clin Epidemiol       Date:  2005-08       Impact factor: 6.437

3.  National study of U.S. emergency department visits with diabetic ketoacidosis, 1993-2003.

Authors:  Adit A Ginde; Andrea J Pelletier; Carlos A Camargo
Journal:  Diabetes Care       Date:  2006-09       Impact factor: 19.112

4.  Prevalence of undiagnosed and suboptimally controlled diabetes by point-of-care HbA1C in unselected emergency department patients.

Authors:  Michael D Menchine; Sanjay Arora; Carlos A Camargo; Adit A Ginde
Journal:  Acad Emerg Med       Date:  2011-03-01       Impact factor: 3.451

5.  The hospital and home use of a 30-second hand-held blood ketone meter: guidelines for clinical practice.

Authors:  T M Wallace; N M Meston; S G Gardner; D R Matthews
Journal:  Diabet Med       Date:  2001-08       Impact factor: 4.359

6.  Near patient blood ketone measurements and their utility in predicting diabetic ketoacidosis.

Authors:  S Harris; R Ng; H Syed; R Hillson
Journal:  Diabet Med       Date:  2005-02       Impact factor: 4.359

7.  Point-of-care test identifies diabetic ketoacidosis at triage.

Authors:  Rosanne Naunheim; Timothy J Jang; Gerald Banet; Alec Richmond; Janet McGill
Journal:  Acad Emerg Med       Date:  2006-05-11       Impact factor: 3.451

Review 8.  Hyperglycemic crises in adult patients with diabetes.

Authors:  Abbas E Kitabchi; Guillermo E Umpierrez; John M Miles; Joseph N Fisher
Journal:  Diabetes Care       Date:  2009-07       Impact factor: 17.152

9.  Urinary acetoacetate or capillary beta-hydroxybutyrate for the diagnosis of ketoacidosis in the Emergency Department setting.

Authors:  Pierre Taboulet; Laurent Haas; Raphael Porcher; Jaffar Manamani; Jean-Paul Fontaine; Jean-Paul Feugeas; Jean-Francois Gautier
Journal:  Eur J Emerg Med       Date:  2004-10       Impact factor: 2.799

10.  Point-of-care blood ketone testing: screening for diabetic ketoacidosis at the emergency department.

Authors:  R A Charles; Y M Bee; P H K Eng; S Y Goh
Journal:  Singapore Med J       Date:  2007-11       Impact factor: 1.858

View more
  19 in total

1.  The Practice Guidelines for Primary Care of Acute Abdomen 2015.

Authors:  Toshihiko Mayumi; Masahiro Yoshida; Susumu Tazuma; Akira Furukawa; Osamu Nishii; Kunihiro Shigematsu; Takeo Azuhata; Atsuo Itakura; Seiji Kamei; Hiroshi Kondo; Shigenobu Maeda; Hiroshi Mihara; Masafumi Mizooka; Toshihiko Nishidate; Hideaki Obara; Norio Sato; Yuichi Takayama; Tomoyuki Tsujikawa; Tomoyuki Fujii; Tetsuro Miyata; Izumi Maruyama; Hiroshi Honda; Koichi Hirata
Journal:  Jpn J Radiol       Date:  2016-01       Impact factor: 2.374

Review 2.  Blood Ketones: Measurement, Interpretation, Limitations, and Utility in the Management of Diabetic Ketoacidosis.

Authors:  Ketan Dhatariya
Journal:  Rev Diabet Stud       Date:  2017-02-10

3.  Comparative performance assessment of point-of-care testing devices for measuring glucose and ketones at the patient bedside.

Authors:  Ferruccio Ceriotti; Ewa Kaczmarek; Elena Guerra; Fabrizio Mastrantonio; Fausto Lucarelli; Francesco Valgimigli; Andrea Mosca
Journal:  J Diabetes Sci Technol       Date:  2014-12-17

4.  Serum ketone body measurement in patients with diabetic ketoacidosis.

Authors:  Tomonao Hirobata; Hidefumi Inaba; Yosuke Kaido; Daisuke Kosugi; Saya Itoh; Takaaki Matsuoka; Gen Inoue
Journal:  Diabetol Int       Date:  2022-04-13

Review 5.  Current and emerging trends in point-of-care urinalysis tests.

Authors:  Rongwei Lei; Rannon Huo; Chandra Mohan
Journal:  Expert Rev Mol Diagn       Date:  2019-12-12       Impact factor: 5.225

6.  The Efficacy of Ketogenic Therapies in the Clinical Management of People with Neurodegenerative Disease: A Systematic Review.

Authors:  Lauren S Dewsbury; Chai K Lim; Genevieve Z Steiner
Journal:  Adv Nutr       Date:  2021-07-30       Impact factor: 8.701

7.  Serum 3-Hydroxybutyrate in Patients with Psychogenic Somatoform Symptoms May Be a Predictor of the Effectiveness of Sertraline and Venlafaxine.

Authors:  Norihiro Saito; Masamichi Itoga; Satoko Minakawa; Hiroyuki Kayaba
Journal:  Int J Gen Med       Date:  2021-05-10

8.  Controversies Around the Measurement of Blood Ketones to Diagnose and Manage Diabetic Ketoacidosis.

Authors:  Eric S Kilpatrick; Alexandra E Butler; Linda Ostlundh; Stephen L Atkin; David B Sacks
Journal:  Diabetes Care       Date:  2022-02-01       Impact factor: 19.112

9.  Serum Beta Hydroxybutyrate Concentrations in Cats with Chronic Kidney Disease, Hyperthyroidism, or Hepatic Lipidosis.

Authors:  L Gorman; L C Sharkey; P J Armstrong; K Little; A Rendahl
Journal:  J Vet Intern Med       Date:  2016-01-15       Impact factor: 3.333

10.  Comparing Finger-stick β-Hydroxybutyrate with Dipstick Urine Tests in the Detection of Ketone Bodies.

Authors:  Baris Kuru; Mustafa Sever; Ersin Aksay; Tarik Dogan; Necmiye Yalcin; Ezgi Seker Eren; Fusun Ustuner
Journal:  Turk J Emerg Med       Date:  2016-02-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.