Use of a continuous glucose sensor in an extracorporeal life support circuit.

Garry M Steil1, Jamin Alexander, Alexandra Papas, Langer Monica, Biren P Modi, Hannah Piper, Tom Jaksic, Rebecca Gottlieb, Michael S D Agus.   

Abstract

BACKGROUND: Standard care for infants on extracorporeal life support (ECLS) relies on intermittent measurement of blood glucose (BG); however, this can lead to significant changes in BG that go unrecognized for several hours. The present study was designed to assess performance and clinical applicability of a subcutaneous glucose sensor technology modified for use as a blood-contacting sensor within the ECLS circuit.
METHODS: Twelve children, aged 3 years or less, requiring ECLS support were studied. Three continuous glucose sensors (Medtronic MiniMed) were inserted into hubs placed in line with the ECLS circuit. Blood glucose was assessed with a laboratory analyzer (BG(LAB); Bayer Rapidlab 860) approximately every 5 h (mean 4.9 ± 3.3 h) with more frequent samples obtained with a bedside monitor (HemoCue) as needed. Sensor current (I(SIG)) was transmitted to a laptop computer and retrospectively calibrated using BGLAB. Sensor performance was assessed by mean absolute relative difference (MARD), linear regression slope and intercept, and correlation, all with BGLAB as reference.
RESULTS: The BGLAB averaged 107.6 ± 36.4 mg/dl (mean ± standard deviation) ranging from 58 to 366 mg/dl. The MARD was 11.4%, with linear regression slope (0.86 ± 0.030) and intercept (9.0 ± 3.2 mg/dl) different from 1 and 0, respectively (p < .05), and correlation (r² = 0.76; p < .001). The system was not associated with any adverse events, and placement and removal into the hubs was easily accomplished. Instances in which more frequent BG values were obtained using a bedside HemoCue (BGHEMO) monitor showed the sensor to respond rapidly to changes.
CONCLUSIONS: We conclude that continuous sensors can be adapted for use in an ECLS circuit with accuracy similar to or better than that achieved with the subcutaneous site. Continuous glucose monitoring in this population can rapidly detect changes in BG that would not otherwise be observed. Further studies will be needed to assess the benefit of continuous glucose monitoring in this population.
© 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21303630      PMCID: PMC3045226          DOI: 10.1177/193229681100500113

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  28 in total

1.  Defining the period of recovery of the glucose concentration after its local perturbation by the implantation of a miniature sensor.

Authors:  Ting Chen; David W Schmidtke; Adam Heller
Journal:  Clin Chem Lab Med       Date:  2002-08       Impact factor: 3.694

2.  Accuracy of a new real-time continuous glucose monitoring algorithm.

Authors:  D Barry Keenan; Raymond Cartaya; John J Mastrototaro
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

3.  The first use of live continuous glucose monitoring in patients on extracorporeal life support.

Authors:  Patrick J Javid; David R Halwick; Peter Betit; John E Thompson; Kenny Long; Yanan Zhang; Tom Jaksic; Michael S D Agus
Journal:  Diabetes Technol Ther       Date:  2005-06       Impact factor: 6.118

4.  Intraoperative glucose control in diabetic and nondiabetic patients during cardiac surgery.

Authors:  Charles E Smith; Nicholas R Styn; Satish Kalhan; Alfred C Pinchak; Inderjit S Gill; Richard P Kramer; Tejbir Sidhu
Journal:  J Cardiothorac Vasc Anesth       Date:  2005-04       Impact factor: 2.628

5.  Clinical evaluation of hormonal stress state in medical ICU patients: a prospective blinded observational study.

Authors:  Ursula G Kyle; Philippe Jolliet; Laurence Genton; Christoph A Meier; Nouri Mensi; Jean-Daniel Graf; Jean-Claude Chevrolet; Claude Pichard
Journal:  Intensive Care Med       Date:  2005-10-25       Impact factor: 17.440

6.  Feasibility of automating insulin delivery for the treatment of type 1 diabetes.

Authors:  Garry M Steil; Kerstin Rebrin; Christine Darwin; Farzam Hariri; Mohammed F Saad
Journal:  Diabetes       Date:  2006-12       Impact factor: 9.461

7.  Extracorporeal Life Support Registry Report 2004.

Authors:  Steven A Conrad; Peter T Rycus; Heidi Dalton
Journal:  ASAIO J       Date:  2005 Jan-Feb       Impact factor: 2.872

8.  Neurodevelopmental status at age five years of neonates treated with extracorporeal membrane oxygenation.

Authors:  P Glass; A E Wagner; P H Papero; S R Rajasingham; L A Civitello; M S Kjaer; C E Coffman; P R Getson; B L Short
Journal:  J Pediatr       Date:  1995-09       Impact factor: 4.406

9.  Tissue response to subcutaneous implantation of glucose-oxidase-based glucose sensors in rats.

Authors:  Nadja Henninger; Stefanie Woderer; Hans-Martin Kloetzer; Arnulf Staib; Ralph Gillen; Li Li; Xiaolei Yu; Norbert Gretz; Bettina Kraenzlin; Johannes Pill
Journal:  Biosens Bioelectron       Date:  2007-03-30       Impact factor: 10.618

10.  Intensive insulin therapy and pentastarch resuscitation in severe sepsis.

Authors:  Frank M Brunkhorst; Christoph Engel; Frank Bloos; Andreas Meier-Hellmann; Max Ragaller; Norbert Weiler; Onnen Moerer; Matthias Gruendling; Michael Oppert; Stefan Grond; Derk Olthoff; Ulrich Jaschinski; Stefan John; Rolf Rossaint; Tobias Welte; Martin Schaefer; Peter Kern; Evelyn Kuhnt; Michael Kiehntopf; Christiane Hartog; Charles Natanson; Markus Loeffler; Konrad Reinhart
Journal:  N Engl J Med       Date:  2008-01-10       Impact factor: 91.245

View more
  3 in total

Review 1.  Hypoglycemia in critically ill children.

Authors:  E Vincent S Faustino; Eliotte L Hirshberg; Clifford W Bogue
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

2.  The artificial pancreas: evaluating risk of hypoglycaemia following errors that can be expected with prolonged at-home use.

Authors:  H Wolpert; M Kavanagh; A Atakov-Castillo; G M Steil
Journal:  Diabet Med       Date:  2015-07-04       Impact factor: 4.359

3.  Fear of Self-Injecting and Self-Testing and the Related Risk Factors in Adolescents with Type 1 Diabetes: A Cross-Sectional Study.

Authors:  Ayman A Al Hayek; Asirvatham A Robert; Saleha Babli; Khuloud Almonea; Mohamed A Al Dawish
Journal:  Diabetes Ther       Date:  2016-12-19       Impact factor: 2.945

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.