Joint correction of respiratory motion artifact and partial volume effect in lung/thoracic PET/CT imaging.

Guoping Chang1, Tingting Chang, Tinsu Pan, John W Clark, Osama R Mawlawi.   

Abstract

PURPOSE: Respiratory motion artifacts and partial volume effects (PVEs) are two degrading factors that affect the accuracy of image quantification in PET/CT imaging. In this article, the authors propose a joint motion and PVE correction approach (JMPC) to improve PET quantification by simultaneously correcting for respiratory motion artifacts and PVE in patients with lung/thoracic cancer. The objective of this article is to describe this approach and evaluate its performance using phantom and patient studies.
METHODS: The proposed joint correction approach incorporates a model of motion blurring, PVE, and object size/shape. A motion blurring kernel (MBK) is then estimated from the deconvolution of the joint model, while the activity concentration (AC) of the tumor is estimated from the normalization of the derived MBK. To evaluate the performance of this approach, two phantom studies and eight patient studies were performed. In the phantom studies, two motion waveforms-a linear sinusoidal and a circular motion-were used to control the motion of a sphere, while in the patient studies, all participants were instructed to breathe regularly. For the phantom studies, the resultant MBK was compared to the true MBK by measuring a correlation coefficient between the two kernels. The measured sphere AC derived from the proposed method was compared to the true AC as well as the ACs in images exhibiting PVE only and images exhibiting both PVE and motion blurring. For the patient studies, the resultant MBK was compared to the motion extent derived from a 4D-CT study, while the measured tumor AC was compared to the AC in images exhibiting both PVE and motion blurring.
RESULTS: For the phantom studies, the estimated MBK approximated the true MBK with an average correlation coefficient of 0.91. The tumor ACs following the joint correction technique were similar to the true AC with an average difference of 2%. Furthermore, the tumor ACs on the PVE only images and images with both motion blur and PVE effects were, on average, 75% and 47.5% (10%) of the true AC, respectively, for the linear (circular) motion phantom study. For the patient studies, the maximum and mean AC/SUV on the PET images following the joint correction are, on average, increased by 125.9% and 371.6%, respectively, when compared to the PET images with both PVE and motion. The motion extents measured from the derived MBK and 4D-CT exhibited an average difference of 1.9 mm.
CONCLUSIONS: The proposed joint correction approach can improve the accuracy of PET quantification by simultaneously compensating for the respiratory motion artifacts and PVE in lung/thoracic PET/CT imaging.

Entities:  

Mesh:

Year:  2010        PMID: 21302779      PMCID: PMC3820620          DOI: 10.1118/1.3512780

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

1.  Reducing respiratory motion artifacts in positron emission tomography through retrospective stacking.

Authors:  Brian Thorndyke; Eduard Schreibmann; Albert Koong; Lei Xing
Journal:  Med Phys       Date:  2006-07       Impact factor: 4.071

2.  A motion-incorporated reconstruction method for gated PET studies.

Authors:  Feng Qiao; Tinsu Pan; John W Clark; Osama R Mawlawi
Journal:  Phys Med Biol       Date:  2006-07-20       Impact factor: 3.609

3.  Deep-inspiration breath-hold PET/CT of the thorax.

Authors:  Sadek A Nehmeh; Yusuf E Erdi; Gustavo S P Meirelles; Olivia Squire; Steven M Larson; John L Humm; Heiko Schöder
Journal:  J Nucl Med       Date:  2007-01       Impact factor: 10.057

Review 4.  Partial-volume effect in PET tumor imaging.

Authors:  Marine Soret; Stephen L Bacharach; Irène Buvat
Journal:  J Nucl Med       Date:  2007-05-15       Impact factor: 10.057

5.  A practical, automated quality assurance method for measuring spatial resolution in PET.

Authors:  Martin A Lodge; Arman Rahmim; Richard L Wahl
Journal:  J Nucl Med       Date:  2009-07-17       Impact factor: 10.057

6.  Simultaneous recovery of size and radioactivity concentration of small spheroids with PET data.

Authors:  C H Chen; R F Muzic; A D Nelson; L P Adler
Journal:  J Nucl Med       Date:  1999-01       Impact factor: 10.057

7.  The application of the sinusoidal model to lung cancer patient respiratory motion.

Authors:  R George; S S Vedam; T D Chung; V Ramakrishnan; P J Keall
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

8.  Deblurring of breathing motion artifacts in thoracic PET images by deconvolution methods.

Authors:  Issam El Naqa; Daniel A Low; Jeffrey D Bradley; Milos Vicic; Joseph O Deasy
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

9.  Partial-volume correction in PET: validation of an iterative postreconstruction method with phantom and patient data.

Authors:  Boon-Keng Teo; Youngho Seo; Stephen L Bacharach; Jorge A Carrasquillo; Steven K Libutti; Himanshu Shukla; Bruce H Hasegawa; Randall A Hawkins; Benjamin L Franc
Journal:  J Nucl Med       Date:  2007-05       Impact factor: 10.057

10.  Incorporation of wavelet-based denoising in iterative deconvolution for partial volume correction in whole-body PET imaging.

Authors:  N Boussion; C Cheze Le Rest; M Hatt; D Visvikis
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-02-18       Impact factor: 9.236

View more
  5 in total

Review 1.  Advances in 4D radiation therapy for managing respiration: part I - 4D imaging.

Authors:  Geoffrey D Hugo; Mihaela Rosu
Journal:  Z Med Phys       Date:  2012-07-10       Impact factor: 4.820

2.  Application of partial volume effect correction and 4D PET in the quantification of FDG avid lung lesions.

Authors:  Ali Salavati; Samuel Borofsky; Teo K Boon-Keng; Sina Houshmand; Benjapa Khiewvan; Babak Saboury; Ion Codreanu; Drew A Torigian; Habib Zaidi; Abass Alavi
Journal:  Mol Imaging Biol       Date:  2015-02       Impact factor: 3.488

3.  Diagnostic accuracy of 18F-FDG PET/CT for detecting recurrence in patients with primary skeletal Ewing sarcoma.

Authors:  Punit Sharma; Bangkim Chandra Khangembam; K C Sudhir Suman; Harmandeep Singh; Sishir Rastogi; Shah Alam Khan; Sameer Bakhshi; Sanjay Thulkar; Chandrasekhar Bal; Arun Malhotra; Rakesh Kumar
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-04-05       Impact factor: 9.236

4.  Application of Image Fusion Algorithm Combined with Visual Saliency in Target Extraction of Reflective Tomography Lidar Image.

Authors:  Xinyuan Zhang; Yihua Hu; Shilong Xu; Fei Han; Yicheng Wang
Journal:  Comput Intell Neurosci       Date:  2022-02-27

5.  Impact of low injected activity on data driven respiratory gating for PET/CT imaging with continuous bed motion.

Authors:  Joseph G Meier; Radwan H Diab; Trevor M Connor; Osama R Mawlawi
Journal:  J Appl Clin Med Phys       Date:  2022-04-28       Impact factor: 2.243

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.