Literature DB >> 21288417

High rate of osteolytic lesions in medium-term followup after the AES total ankle replacement.

Ari Kokkonen1, Mikko Ikävalko, Raine Tiihonen, Hannu Kautiainen, Eero A Belt.   

Abstract

BACKGROUND: Some previous studies have shown a high percentage of early-onset and rapidly progressing osteolysis associated with total ankle arthroplasty (TAA) by the Ankle Evolutive System (AES). The purpose of our study was to analyze medium-term results at our institution.
MATERIALS AND METHODS: Altogether 38 TAAs using AES prostheses were carried out between 2003 and 2007. Diagnoses were rheumatoid arthritis (71%), post-traumatic and idiopathic osteoarthritis (29%). The mean age was 54 years, followup 28 months. Tibial and talar components had hydroxyapatite coating on metal (Co-Cr) components (HA-coated). Since 2005 the design was changed and components were porous coated with titanium and hydroxyapatite (dual-coated).
RESULTS: Two-year survival was 79% (95% CI: 56 to 98). At followup 34 (89%) primary tibial and talar components were preserved. In 19 (50%) TAAs osteolysis (more than or equal to 2 mm) occurred in the periprosthetic bone area and in nine (24%) comprised large "cyst-like osteolysis''. In HA-coated prostheses radiolucent lines (less than or equal to 2 mm) or osteolysis (more than or equal to 2 mm) were detected in 11 (100%) cases and in dual-coated prostheses in 19 (74%) (p = 0.08). On the other hand there was more large "cyst-like osteolysis'' around the dual-coated prosthesis and lesions were larger (p = 0.017). In rheumatoid arthritis osteolysis was detected in 14 (52%) and large "cyst-like osteolysis'' in seven (26%) prostheses and in the group of traumatic and idiopathic osteoarthritis in six (55%) and two (18%), respectively.
CONCLUSION: This study showed a high frequency of osteolysis in medium-term followup after the AES ankle replacement. The outcome was not sufficiently beneficial and we have discontinued use of this prosthesis.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21288417     DOI: 10.3113/FAI.2011.0168

Source DB:  PubMed          Journal:  Foot Ankle Int        ISSN: 1071-1007            Impact factor:   2.827


  11 in total

Review 1.  Total ankle replacement.

Authors:  Alexej Barg; Matthias D Wimmer; Martin Wiewiorski; Dieter C Wirtz; Geert I Pagenstert; Victor Valderrabano
Journal:  Dtsch Arztebl Int       Date:  2015-03-13       Impact factor: 5.594

2.  Total ankle arthroplasty: optimizing computed tomography imaging protocol.

Authors:  Ia Kohonen; Helka Koivu; Tero Vahlberg; Heli Larjava; Kimmo Mattila
Journal:  Skeletal Radiol       Date:  2013-08-03       Impact factor: 2.199

3.  CORR Insights®: Can a Three-component Prosthesis be Used for Conversion of Painful Ankle Arthrodesis to Total Ankle Replacement?

Authors:  Andrew Haskell
Journal:  Clin Orthop Relat Res       Date:  2017-04-28       Impact factor: 4.176

4.  Can a Three-Component Prosthesis be Used for Conversion of Painful Ankle Arthrodesis to Total Ankle Replacement?

Authors:  Markus Preis; Travis Bailey; Lucas S Marchand; Alexej Barg
Journal:  Clin Orthop Relat Res       Date:  2017-04-19       Impact factor: 4.176

5.  The AES total ankle arthroplasty analysis of failures and survivorship at ten years.

Authors:  Alexandre Di Iorio; Anthony Viste; Michel Henry Fessy; Jean Luc Besse
Journal:  Int Orthop       Date:  2017-09-04       Impact factor: 3.075

6.  Thrombembolic complications after total ankle replacement.

Authors:  Alexej Barg; Katharina Barg; Stefan W Schneider; Geert Pagenstert; Marcel Gloyer; Heath B Henninger; Victor Valderrabano
Journal:  Curr Rev Musculoskelet Med       Date:  2013-09-28

7.  Long-term follow-up of Bologna-Oxford (BOX) total ankle arthroplasty.

Authors:  Alberto Bianchi; Nicolò Martinelli; Eleonora Caboni; Giorgio Raggi; Francesca Manfroni; Valerio Sansone
Journal:  Int Orthop       Date:  2021-04-06       Impact factor: 3.075

Review 8.  [Thromboembolic complications following ankle prosthesis implantation].

Authors:  A Barg; S W Schneider; G Pagenstert; B Hintermann; V Valderrabano
Journal:  Orthopade       Date:  2013-11       Impact factor: 1.087

9.  Application of stem cells in orthopedics.

Authors:  Andreas Schmitt; Martijn van Griensven; Andreas B Imhoff; Stefan Buchmann
Journal:  Stem Cells Int       Date:  2012-02-23       Impact factor: 5.443

10.  Incidence of rheumatoid arthritis-related ankle replacement and ankle arthrodesis: a Finnish nationwide register-based study from 1997–2010.

Authors:  Ari Kokkonen; Esa Jämsen; Eero A Belt; Matti U K Lehto
Journal:  Acta Orthop       Date:  2013-08       Impact factor: 3.717

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.