Literature DB >> 21284867

Digital stereophotogrammetry based on circular markers and zooming cameras: evaluation of a method for 3D analysis of small motions in orthopaedic research.

Evgenij Bobrowitsch1, Christof Hurschler, Gavin Olender, Christian Plaass, Hazibullah Waizy, Heino Arnold, Christina Stukenborg-Colsman.   

Abstract

BACKGROUND: Orthopaedic research projects focusing on small displacements in a small measurement volume require a radiation free, three dimensional motion analysis system. A stereophotogrammetrical motion analysis system can track wireless, small, light-weight markers attached to the objects. Thereby the disturbance of the measured objects through the marker tracking can be kept at minimum. The purpose of this study was to develop and evaluate a non-position fixed compact motion analysis system configured for a small measurement volume and able to zoom while tracking small round flat markers in respect to a fiducial marker which was used for the camera pose estimation.
METHODS: The system consisted of two web cameras and the fiducial marker placed in front of them. The markers to track were black circles on a white background. The algorithm to detect a centre of the projected circle on the image plane was described and applied. In order to evaluate the accuracy (mean measurement error) and precision (standard deviation of the measurement error) of the optical measurement system, two experiments were performed: 1) inter-marker distance measurement and 2) marker displacement measurement.
RESULTS: The first experiment of the 10 mm distances measurement showed a total accuracy of 0.0086 mm and precision of ± 0.1002 mm. In the second experiment, translations from 0.5 mm to 5 mm were measured with total accuracy of 0.0038 mm and precision of ± 0.0461 mm. The rotations of 2.25° amount were measured with the entire accuracy of 0.058° and the precision was of ± 0.172°.
CONCLUSIONS: The description of the non-proprietary measurement device with very good levels of accuracy and precision may provide opportunities for new, cost effective applications of stereophotogrammetrical analysis in musculoskeletal research projects, focusing on kinematics of small displacements in a small measurement volume.

Entities:  

Mesh:

Year:  2011        PMID: 21284867      PMCID: PMC3038981          DOI: 10.1186/1475-925X-10-12

Source DB:  PubMed          Journal:  Biomed Eng Online        ISSN: 1475-925X            Impact factor:   2.819


  15 in total

1.  Total strain fields of the antero-inferior shoulder capsule under subluxation: a stereoradiogrammetric study.

Authors:  D M Malicky; L J Soslowsky; J E Kuhn; M J Bey; C M Mouro; J A Raz; C A Liu
Journal:  J Biomech Eng       Date:  2001-10       Impact factor: 2.097

2.  Neer Award 2001: nonrecoverable strain fields of the anteroinferior glenohumeral capsule under subluxation.

Authors:  David M Malicky; John E Kuhn; Juan C Frisancho; Stephen R Lindholm; Jonathan A Raz; Louis J Soslowsky
Journal:  J Shoulder Elbow Surg       Date:  2002 Nov-Dec       Impact factor: 3.019

3.  Designing highly reliable fiducial markers.

Authors:  Mark Fiala
Journal:  IEEE Trans Pattern Anal Mach Intell       Date:  2010-07       Impact factor: 6.226

4.  Simultaneous measurement of three-dimensional joint kinematics and ligament strains with optical methods.

Authors:  Trevor J Lujan; Spencer P Lake; Timothy A Plaizier; Benjamin J Ellis; Jeffrey A Weiss
Journal:  J Biomech Eng       Date:  2005-02       Impact factor: 2.097

5.  Evaluation of a 3D object registration method for analysis of humeral kinematics.

Authors:  Evgenij Bobrowitsch; Carl Imhauser; Heiko Graichen; Lutz Dürselen
Journal:  J Biomech       Date:  2006-04-19       Impact factor: 2.712

6.  Noninvasive determination of ligament strain with deformable image registration.

Authors:  Nikhil S Phatak; Qunli Sun; Seong-Eun Kim; Dennis L Parker; R Kent Sanders; Alexander I Veress; Benjamin J Ellis; Jeffrey A Weiss
Journal:  Ann Biomed Eng       Date:  2007-03-30       Impact factor: 3.934

7.  Experimental analysis of Model-Based Roentgen Stereophotogrammetric Analysis (MBRSA) on four typical prosthesis components.

Authors:  Frank Seehaus; Judith Emmerich; Bart L Kaptein; Henning Windhagen; Christof Hurschler
Journal:  J Biomech Eng       Date:  2009-04       Impact factor: 2.097

8.  Determining the movements of the skeleton using well-configured markers.

Authors:  I Söderkvist; P A Wedin
Journal:  J Biomech       Date:  1993-12       Impact factor: 2.712

9.  High-precision three-dimensional photogrammetric calibration and object space reconstruction using a modified DLT-approach.

Authors:  H Hatze
Journal:  J Biomech       Date:  1988       Impact factor: 2.712

10.  An investigation on the accuracy of three-dimensional space reconstruction using the direct linear transformation technique.

Authors:  L Chen; C W Armstrong; D D Raftopoulos
Journal:  J Biomech       Date:  1994-04       Impact factor: 2.712

View more
  1 in total

1.  Biomechanical in vitro - stability testing on human specimens of a locking plate system against conventional screw fixation of a proximal first metatarsal lateral displacement osteotomy.

Authors:  Heino Arnold; Christina Stukenborg-Colsman; Christof Hurschler; Frank Seehaus; Evgenij Bobrowitsch; Hazibullah Waizy
Journal:  Open Orthop J       Date:  2012-04-04
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.