Phantom experiments on a PSAPD-based compact gamma camera with submillimeter spatial resolution for small animal SPECT.

Sangtaek Kim1, Mickel McClish, Fares Alhassen, Youngho Seo, Kanai S Shah, Robert G Gould.   

Abstract

We demonstrate a position sensitive avalanche photodiode (PSAPD) based compact gamma camera for the application of small animal single photon emission computed tomography (SPECT). The silicon PSAPD with a two-dimensional resistive layer and four readout channels is implemented as a gamma ray detector to record the energy and position of radiation events from a radionuclide source. A 2 mm thick monolithic CsI:Tl scintillator is optically coupled to a PSAPD with a 8mm×8mm active area, providing submillimeter intrinsic spatial resolution, high energy resolution (16% full-width half maximum at 140 keV) and high gain. A mouse heart phantom filled with an aqueous solution of 370 MBq (99m)Tc-pertechnetate (140 keV) was imaged using the PSAPD detector module and a tungsten knife-edge pinhole collimator with a 0.5 mm diameter aperture. The PSAPD detector module was cooled with cold nitrogen gas to suppress dark current shot noise. For each projection image of the mouse heart phantom, a rotated diagonal readout algorithm was used to calculate the position of radiation events and correct for pincushion distortion. The reconstructed image of the mouse heart phantom demonstrated reproducible image quality with submillimeter spatial resolution (0.7 mm), showing the feasibility of using the compact PSAPD-based gamma camera for a small animal SPECT system.

Entities:  

Year:  2010        PMID: 21278833      PMCID: PMC3026481          DOI: 10.1109/TNS.2010.2056386

Source DB:  PubMed          Journal:  IEEE Trans Nucl Sci        ISSN: 0018-9499            Impact factor:   1.679


  5 in total

1.  SCINTILLATION CAMERA WITH MULTICHANNEL COLLIMATORS.

Authors:  H O ANGER
Journal:  J Nucl Med       Date:  1964-07       Impact factor: 10.057

2.  Attenuation correction for small animal SPECT imaging using x-ray CT data.

Authors:  Andrew B Hwang; Bruce H Hasegawa
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

3.  Pinhole SPECT: an approach to in vivo high resolution SPECT imaging in small laboratory animals.

Authors:  D A Weber; M Ivanovic; D Franceschi; S E Strand; K Erlandsson; M Franceschi; H L Atkins; J A Coderre; H Susskind; T Button
Journal:  J Nucl Med       Date:  1994-02       Impact factor: 10.057

4.  U-SPECT-II: An Ultra-High-Resolution Device for Molecular Small-Animal Imaging.

Authors:  Frans van der Have; Brendan Vastenhouw; Ruud M Ramakers; Woutjan Branderhorst; Jens O Krah; Changguo Ji; Steven G Staelens; Freek J Beekman
Journal:  J Nucl Med       Date:  2009-03-16       Impact factor: 10.057

5.  Monte Carlo simulations of compact gamma cameras based on avalanche photodiodes.

Authors:  Philippe Després; Tobias Funk; Kanai S Shah; Bruce H Hasegawa
Journal:  Phys Med Biol       Date:  2007-05-08       Impact factor: 3.609

  5 in total
  2 in total

1.  Signal and noise properties of position-sensitive avalanche photodiodes.

Authors:  Yongfeng Yang; Yibao Wu; Richard Farrell; Purushottam A Dokhale; Kanai S Shah; Simon R Cherry
Journal:  Phys Med Biol       Date:  2011-09-06       Impact factor: 3.609

2.  Temperature dependent operation of PSAPD-based compact gamma camera for SPECT imaging.

Authors:  Sangtaek Kim; Mickel McClish; Fares Alhassen; Youngho Seo; Kanai S Shah; Robert G Gould
Journal:  IEEE Trans Nucl Sci       Date:  2011-10-10       Impact factor: 1.679

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.