Efficacy of several digital radiographic imaging systems for laboratory determination of endodontic file length.

M L Oliveira1, G M B Ambrosano, S M Almeida, F Haiter-Neto, G M Tosoni.   

Abstract

AIM: To compare the efficacy of different digital radiographic imaging systems for determining the length of endodontic files.
METHODOLOGY: K-type endodontic files were introduced into the canals of 40 extracted human permanent single-rooted teeth and fixed in place at random lengths. The teeth were radiographed using Digora Optime, CygnusRay MPS and CDR Wireless digital imaging systems. Six observers measured every file length in all the images and repeated this procedure in 50% of the image samples, and assigned a score to the level of difficulty found. Analysis of variance for differences between digital systems and Tukey's test were performed. The level of intraobserver agreement was measured by intraclass correlation. The assigned scores were evaluated by Kruskal-Wallis and Dunn's tests.
RESULTS: The CDR Wireless values did not differ significantly from the actual lengths and the CygnusRay MPS values. The Digora Optime system was significantly different from the others and overestimated the values (P ≤ 0.05). The Digora Optime was significantly easier to use for taking measurements and the CygnusRay MPS the most difficult (P ≤ 0.05). All digital radiographic imaging systems showed excellent agreement with the Intraclass Correlation Coefficient >0.95.
CONCLUSIONS: The three digital radiographic imaging systems were precise. The CDR Wireless system was significantly more accurate in determining endodontic file lengths, and similarly to Digora Optime, was considered the least difficult to use when assessing endodontic file lengths.
© 2011 International Endodontic Journal.

Entities:  

Mesh:

Year:  2011        PMID: 21276021     DOI: 10.1111/j.1365-2591.2011.01860.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  4 in total

1.  Effect of Contrast Inversion Enhancement on the Accuracy of Endodontic File Length Determination in Digital Radiography.

Authors:  Nastaran Farhadi; Ali Shokraneh; Mojdeh Mehdizadeh
Journal:  J Clin Diagn Res       Date:  2015-05-01

2.  Comparison of two methods of digital imaging technology for small diameter K-file length determination.

Authors:  Ehsani Maryam; Abesi Farida; Akbarzade Farhad; Khafri Soraya
Journal:  Int J Health Sci (Qassim)       Date:  2013-11

3.  Image quality evaluation of eight complementary metal-oxide semiconductor intraoral digital X-ray sensors.

Authors:  Sorin Teich; Wisam Al-Rawi; Masahiro Heima; Fady F Faddoul; Gil Goldzweig; Zvi Gutmacher; Dror Aizenbud
Journal:  Int Dent J       Date:  2016-04-22       Impact factor: 2.607

4.  Influence of cone-beam computed tomography image artifacts on the determination of dental arch measurements.

Authors:  Thiago O Gamba; Matheus L Oliveira; Isadora L Flores; Adriana D Cruz; Solange M Almeida; Francisco Haiter-Neto; Sérgio L P C Lopes
Journal:  Angle Orthod       Date:  2013-08-07       Impact factor: 2.079

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.