Literature DB >> 21245709

Failure of proximal femoral locking compression plate: a case series.

Philip J Glassner1, Nirmal C Tejwani.   

Abstract

OBJECTIVES: The treatment of fractures of the proximal femur has evolved significantly over recent years. The most recent advance is the proximal femoral locking compression plate (PF-LCP; Synthes, West Chester, PA). We present seven failures of the PF-LCP in an attempt to elicit reasons for the failures so that these pitfalls can be avoided in the future.
DESIGN: Retrospective chart review.
SETTING: Level I trauma centers, tertiary referral hospitals. PATIENTS/PARTICIPANTS: Ten patients at five institutions. INTERVENTION: Open reduction and internal fixation with a PF-LCP. MAIN OUTCOME MEASUREMENTS: Failure mode, time until failure, need for further surgery, bony union, or conversion to arthroplasty.
RESULTS: Of the seven cases, two were acute peritrochanteric fractures, one was a periprosthetic fracture at the site of a prior hip fusion, one was an early failure of a compression hip screw, and three were nonunions. The failure mode was implant fracture in four cases and loss of fixation in three cases resulting from varus collapse and implant cutout. Five of seven failures were within the first 3 weeks (average, 12.4 days). The average time to failure for all cases was 37.9 days (range, 5-175 days). The average patient age was 56.7 years (range, 36-72 years).
CONCLUSION: Biomechanical studies have shown the PF-LCP to be stronger or equivalent to other fixation methods for fractures of the femoral neck and subtrochanteric femur fractures. The seven failures in our cases may be partially the result of patient factors as well as technical factors; however, there appears to be a high rate of failure even when surgery is performed by experienced and fellowship-trained traumatologists.

Entities:  

Mesh:

Year:  2011        PMID: 21245709     DOI: 10.1097/BOT.0b013e3181e31ccc

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  24 in total

1.  Proximal femoral nails compared with reverse distal femoral locking plates in intertrochanteric fractures with a compromised lateral wall; a randomised controlled trial.

Authors:  Rehan Ul Haq; Vikrant Manhas; Amite Pankaj; Amit Srivastava; Ish Kumar Dhammi; Anil Kumar Jain
Journal:  Int Orthop       Date:  2014-03-22       Impact factor: 3.075

2.  Management of unstable pertrochanteric fractures with proximal femoral locking compression plates and affect of neck-shaft angle on functional outcomes.

Authors:  Emrah Kovalak; Cenk Ermutlu; Tolga Atay; Özgür Başal
Journal:  J Clin Orthop Trauma       Date:  2017-07-15

3.  Biomechanical design prognosis of two extramedullary fixation devices for subtrochanteric femur fracture: a finite element study.

Authors:  Pratik Nag; Souptick Chanda
Journal:  Med Biol Eng Comput       Date:  2021-01-08       Impact factor: 2.602

4.  Bone healing of critical size defects of the rat femur after the application of bone marrow aspirate and two different rh-BMP7 concentrations.

Authors:  F Högel; S Hoffmann; S Hungerer; E Fleischacker; T Ullamann; O B Betz; P Augat
Journal:  Eur J Trauma Emerg Surg       Date:  2014-11-22       Impact factor: 3.693

5.  Plate failure following plate osteosynthesis in periprosthetic femoral fractures.

Authors:  Sandra Boesmueller; Sebastian F Baumbach; Marcus Hofbauer; Gerald E Wozasek
Journal:  Wien Klin Wochenschr       Date:  2015-06-18       Impact factor: 1.704

6.  Salvage of failed hip osteosynthesis for fractures with modular hip prosthesis.

Authors:  Panagiotis K Karampinas; George Kollias; John Vlamis; Eustratios A Papadelis; Spiros G Pneumaticos
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-03-10

7.  Subtrochanteric Fractures: Comparison of Proximal Femur Locking Plate and Intramedullary Locking Nail Fixation Outcome.

Authors:  Ahmadreza Mirbolook; Babak Siavashi; Adel Ebrahimpour Jafarinezhad; Sina Khajeh Jahromi; Maral Farahmand; Melina Roohi Rad; Mohammad Amin Kahe
Journal:  Indian J Surg       Date:  2013-11-16       Impact factor: 0.656

8.  The use of the 95-degree-angled blade plate in femoral nonunion surgery.

Authors:  Louis F Amorosa; Prem R Jayaram; David S Wellman; Dean G Lorich; David L Helfet
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-07-26

9.  Unstable Intertrochanteric Fracture Fixation - Is Proximal Femoral Locked Compression Plate Better Than Dynamic Hip Screw.

Authors:  Naiyer Asif; Sohail Ahmad; Owais Ahmad Qureshi; Latif Zafar Jilani; Tajdar Hamesh; Tariq Jameel
Journal:  J Clin Diagn Res       Date:  2016-01-01

10.  Novel Treatment Options for the Surgical Management of Young Femoral Neck Fractures.

Authors:  Ashley E Levack; Elizabeth B Gausden; Aleksey Dvorzhinskiy; Dean G Lorich; David L Helfet
Journal:  J Orthop Trauma       Date:  2019-01       Impact factor: 2.512

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.