Literature DB >> 21233500

Closed-loop coadministration of propofol and remifentanil guided by bispectral index: a randomized multicenter study.

Ngai Liu1, Thierry Chazot, Sophie Hamada, Alain Landais, Nathalie Boichut, Corinne Dussaussoy, Bernard Trillat, Laurent Beydon, Emmanuel Samain, Daniel I Sessler, Marc Fischler.   

Abstract

BACKGROUND: We have developed a proportional-integral-derivative controller allowing the closed-loop coadministration of propofol and remifentanil, guided by a Bispectral Index (BIS) monitor, during induction and maintenance of general anesthesia. The controller was compared with manual target-controlled infusion.
METHODS: In a multicenter study, 196 surgical patients were randomly assigned to dual closed-loop or manual administration of propofol and remifentanil. Comparison between groups was evaluated by calculating a global score that characterized the overall performance of the controller including the percentage of adequate anesthesia, defined as BIS between 40 and 60, the median absolute performance error, and wobble. Secondary outcomes included occurrence of burst suppression ratio, time to tracheal extubation, and drug consumption.
RESULTS: Eighty-three patients assigned to dual-loop control and 84 patients assigned to manual control completed the study. The global score and the percentage of time with BIS between 40 and 60 were better in the dual-loop group (26 ± 11 vs 43 ± 40, P < 0.0001; 82% ± 12% vs 71% ± 19%, P < 0.0001). Overshoot (BIS <40), undershoot (BIS >60), and burst suppression ratio were all significantly less common in the dual-loop group. Modifications to the propofol and remifentanil infusions were more frequent, and adjustments smaller in the dual-loop group. Remifentanil consumption was greater (0.22 ± 0.07 vs 0.16 ± 0.07 μg · kg(-1) · min(-1); P < 0.0001) and the speed to tracheal extubation was shorter (10 ± 4 vs 11 ± 5 minutes; P = 0.02) in the dual-loop group.
CONCLUSION: The controller allows the automated delivery of propofol and remifentanil and maintains BIS values in predetermined boundaries during general anesthesia better than manual administration.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21233500     DOI: 10.1213/ANE.0b013e318205680b

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  31 in total

1.  A beat-by-beat, on-line, cardiovascular index, CARDEAN, to assess circulatory responses to surgery: a randomized clinical trial during spine surgery.

Authors:  M Rossi; A Cividjian; M C Fevre; M E Oddoux; J Carcey; C Halle; M Frost; M Gardellin; J F Payen; L Quintin
Journal:  J Clin Monit Comput       Date:  2012-06-10       Impact factor: 2.502

2.  Lessons learned from closed loops in engineering: towards a multivariable approach regulating depth of anaesthesia.

Authors:  Clara M Ionescu; Ioana Nascu; Robin De Keyser
Journal:  J Clin Monit Comput       Date:  2013-11-23       Impact factor: 2.502

3.  Smart sensors for maintaining physiologic homeostasis.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2011-05-01

4.  Optimizing Robust PID Control of Propofol Anesthesia for Children: Design and Clinical Evaluation.

Authors:  Klaske van Heusden; Kristian Soltesz; Erin Cooke; Sonia Brodie; Nicholas West; Matthias Gorges; J Mark Ansermino; Guy A Dumont
Journal:  IEEE Trans Biomed Eng       Date:  2019-02-08       Impact factor: 4.538

5.  New technology in anaesthesia: friend or foe?

Authors:  R Ross Kennedy
Journal:  J Clin Monit Comput       Date:  2014-04       Impact factor: 2.502

6.  Artificial Intelligence and Machine Learning in Anesthesiology.

Authors:  Christopher W Connor
Journal:  Anesthesiology       Date:  2019-12       Impact factor: 7.892

7.  Feasibility of continuous sedation monitoring in critically ill intensive care unit patients using the NeuroSENSE WAVCNS index.

Authors:  Nicholas West; Paul B McBeth; Sonia M Brodie; Klaske van Heusden; Sarah Sunderland; Guy A Dumont; Donald E G Griesdale; J Mark Ansermino; Matthias Görges
Journal:  J Clin Monit Comput       Date:  2018-02-20       Impact factor: 2.502

8.  Automated sedation outperforms manual administration of propofol and remifentanil in critically ill patients with deep sedation: a randomized phase II trial.

Authors:  Morgan Le Guen; Ngai Liu; Eric Bourgeois; Thierry Chazot; Daniel I Sessler; Jean-Jacques Rouby; Marc Fischler
Journal:  Intensive Care Med       Date:  2012-12-06       Impact factor: 17.440

9.  Automated titration of propofol and remifentanil decreases the anesthesiologist's workload during vascular or thoracic surgery: a randomized prospective study.

Authors:  Corinne Dussaussoy; Marine Peres; Virginie Jaoul; Ngai Liu; Thierry Chazot; Jean Picquet; Marc Fischler; Laurent Beydon
Journal:  J Clin Monit Comput       Date:  2013-03-15       Impact factor: 2.502

10.  A technical description of a novel pharmacological anesthesia robot.

Authors:  Mohamad Wehbe; Erik Arbeid; Shantale Cyr; Pierre A Mathieu; Riccardo Taddei; Joshua Morse; Thomas M Hemmerling
Journal:  J Clin Monit Comput       Date:  2013-06-15       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.