Literature DB >> 21221728

Effect of the environment on wear ranking and corrosion of biomedical CoCrMo alloys.

A Igual Muñoz1, S Mischler.   

Abstract

The corrosion behaviour and the wear ranking of biomedical high carbon (HC) and low carbon (LC) CoCrMo alloys sliding against an alumina ball in four different simulated body fluids [NaCl and phosphate buffered solutions (PBS) with and without albumin] has been analyzed by tribocorrosion and electrochemical techniques. The effects of alloy and of albumin on corrosion depend on the base electrolyte: differences between LC and HC alloy were only observed in NaCl solutions but not in PBS. Albumin increased significantly corrosion of both alloys in PBS solutions while its effect in NaCl was smaller. The wear ranking of the HC and LC alloys also depends on the environment. In the present study, HC CoCrMo alloy had lower wear resistance in NaCl and PBS + albumin than the LC alloy, while no differences between both alloys were found in the other solutions. This was attributed to surface chemical effects affecting third body behaviour.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21221728     DOI: 10.1007/s10856-010-4224-0

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  8 in total

1.  Quantitative analysis of the wear and wear debris from low and high carbon content cobalt chrome alloys used in metal on metal total hip replacements.

Authors:  J L Tipper; P J Firkins; E Ingham; J Fisher; M H Stone; R Farrar
Journal:  J Mater Sci Mater Med       Date:  1999-06       Impact factor: 3.896

2.  Microstructure and corrosion behaviour in biological environments of the new forged low-Ni Co-Cr-Mo alloys.

Authors:  Sachiko Hiromoto; Emi Onodera; Akihiko Chiba; Katsuhiko Asami; Takao Hanawa
Journal:  Biomaterials       Date:  2005-08       Impact factor: 12.479

3.  Effect of microstructure on the dry sliding friction behavior of CoCrMo alloys used in metal-on-metal hip implants.

Authors:  R Varano; J D Bobyn; J B Medley; S Yue
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2006-02       Impact factor: 3.368

4.  The effect of microstructure on the wear of cobalt-based alloys used in metal-on-metal hip implants.

Authors:  R Varano; J D Bobyn; J B Medley; S Yue
Journal:  Proc Inst Mech Eng H       Date:  2006-02       Impact factor: 1.617

5.  A novel low wearing differential hardness, ceramic-on-metal hip joint prosthesis.

Authors:  P J Firkins; J L Tipper; E Ingham; M H Stone; R Farrar; J Fisher
Journal:  J Biomech       Date:  2001-10       Impact factor: 2.712

Review 6.  Development and validation of a second-generation metal-on-metal bearing: laboratory studies and analysis of retrievals.

Authors:  Claude B Rieker; Rolf Schön; Petra Köttig
Journal:  J Arthroplasty       Date:  2004-12       Impact factor: 4.757

7.  Wear evaluation of cobalt-chromium alloy for use in a metal-on-metal hip prosthesis.

Authors:  Kenneth R St John; Lyle D Zardiackas; Robert A Poggie
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2004-01-15       Impact factor: 3.368

8.  Pin-on-plate studies on the effect of rotation on the wear of metal-on-metal samples.

Authors:  S C Scholes; A Unsworth
Journal:  J Mater Sci Mater Med       Date:  2001-04       Impact factor: 3.896

  8 in total
  7 in total

1.  The effect of contact load on CoCrMo wear and the formation and retention of tribofilms.

Authors:  M A Wimmer; M P Laurent; M T Mathew; C Nagelli; Y Liao; L D Marks; J J Jacobs; A Fischer
Journal:  Wear       Date:  2015 May-Jun       Impact factor: 3.892

2.  Wear-corrosion synergism in a CoCrMo hip bearing alloy is influenced by proteins.

Authors:  Mathew T Mathew; Joshua J Jacobs; Markus A Wimmer
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

3.  CoCrMo metal-on-metal hip replacements.

Authors:  Yifeng Liao; Emily Hoffman; Markus Wimmer; Alfons Fischer; Joshua Jacobs; Laurence Marks
Journal:  Phys Chem Chem Phys       Date:  2013-01-21       Impact factor: 3.676

Review 4.  Tribochemical Characterization and Tribocorrosive Behavior of CoCrMo Alloys: A Review.

Authors:  Wei Quan Toh; Xipeng Tan; Ayan Bhowmik; Erjia Liu; Shu Beng Tor
Journal:  Materials (Basel)       Date:  2017-12-26       Impact factor: 3.623

5.  Macrophage Biocompatibility of CoCr Wear Particles Produced under Polarization in Hyaluronic Acid Aqueous Solution.

Authors:  Blanca Teresa Perez-Maceda; María Encarnación López-Fernández; Iván Díaz; Aaron Kavanaugh; Fabrizio Billi; María Lorenza Escudero; María Cristina García-Alonso; Rosa María Lozano
Journal:  Materials (Basel)       Date:  2018-05-08       Impact factor: 3.623

6.  Evaluation of corrosion resistance of Co-Cr alloys fabricated with different metal laser sintering systems.

Authors:  Süleyman Hakan Tuna; Erhan Karaca; İsmail Aslan; Gürel Pekkan; Nuran Özçiçek Pekmez
Journal:  J Adv Prosthodont       Date:  2020-06-18       Impact factor: 1.904

7.  Titanium Implant Impairment and Surrounding Muscle Cell Death Following High-Salt Diet: An In Vivo Study.

Authors:  Mathieu Lecocq; Marie-Solenne Felix; Jean-Marc Linares; Julien Chaves-Jacob; Patrick Decherchi; Erick Dousset
Journal:  PLoS One       Date:  2016-01-13       Impact factor: 3.240