Comparison in the performance of glucose meters in blood glucose monitoring during pregnancy.

Grace Wing Shan Kong1, Wing Hung Tam, Michael Ho Ming Chan, Wing Yee So, Christopher Wai Kei Lam, Ivy Po Chu Yiu, Kit Man Loo, Chi Yin Li.   

Abstract

AIMS: To compare the performance of the four latest models of glucose meters in capillary blood glucose monitoring during pregnancy.
METHODS: 208 pregnant women with gestational diabetes were recruited. Each subject had simultaneous capillary glucose monitoring by two study glucose meters and venous plasma glucose assay. The performance of four glucose meters was compared using error grid analysis (EGA) and the agreement between the meter readings and plasma glucose by Bland-Altman plot analysis.
RESULTS: Elite, Advantage II and CareSens had more than 90% of readings in the acceptable target range of EGA. CareSens had the lowest mean bias by Bland-Altman analysis while Advantage II had the highest proportion of readings within 5% difference from plasma glucose. Readings from all glucose meters except Optium were not influenced by the change in maternal hematocrit levels.
CONCLUSIONS: The performance of four study glucose meters appeared very similar. Copyright (c) 2010 S. Karger AG, Basel.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20090357     DOI: 10.1159/000276572

Source DB:  PubMed          Journal:  Gynecol Obstet Invest        ISSN: 0378-7346            Impact factor:   2.031


  4 in total

1.  How Satisfied Are Patients When Their Choice of Funded Glucose Meter Is Restricted to a Single Brand?

Authors:  Christy Macdonald; Helen Lunt; Michelle Downie; Deborah Kendall
Journal:  J Diabetes Sci Technol       Date:  2017-02-01

2.  Evaluation of CareSens POCT Devices for Glucose Testing in the Routine Hospital Setting.

Authors:  Kagan Huysal; Yasemin U Budak; Hakan Demirci; Mehmet Ali Önelge
Journal:  J Clin Diagn Res       Date:  2015-10-01

3.  Suboptimal performance of blood glucose meters in an antenatal diabetes clinic.

Authors:  Nimalie Jacintha Perera; Lynda Molyneaux; Maria Ines Constantino; Marg McGill; Dennis Koon-See Yue; Stephen Morris Twigg; Glynis Pauline Ross
Journal:  Diabetes Care       Date:  2011-01-07       Impact factor: 19.112

4.  Self-monitoring of blood glucose during pregnancy: indications and limitations.

Authors:  Carlos Antonio Negrato; Lenita Zajdenverg
Journal:  Diabetol Metab Syndr       Date:  2012-12-22       Impact factor: 3.320

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.