Validation of the Medisana MTP Plus upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010.

Emre Erdem1, Türkan Aydogdu, Tekin Akpolat.   

Abstract

OBJECTIVES: Standard validation protocols are objective guides for healthcare providers, physicians, and patients. The purpose of this study was to test validation of the Medisana MTP Plus upper arm blood pressure (BP) measuring monitor for self-measurement according to the European Society of Hypertension International Protocol (ESH-IP2) in adults.
METHODS: The Medisana MTP Plus monitor is an automated and oscillometric upper arm device for home BP monitoring. Nine consecutive measurements were made according to the ESH-IP2. Overseen by an independent supervisor, measurements were recorded by two observers blinded from both each other's readings and from the device readings.
RESULTS: The Medisana MTP Plus device fulfills the validation criteria of the ESH-IP2 for the general population. The mean (standard deviation) of the difference between the observers and the device measurements was 0.6 mmHg (5.1 mmHg) for systolic and 2.7 mmHg (3.4 mmHg) for diastolic pressures, respectively.
CONCLUSION: As the Medisana MTP Plus device has achieved the required standards, it is recommended for home BP monitoring in an adult population.

Entities:  

Mesh:

Year:  2011        PMID: 21200316     DOI: 10.1097/MBP.0b013e3283419623

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  3 in total

1.  MEthods of ASsessing blood pressUre: identifying thReshold and target valuEs (MeasureBP): a review & study protocol.

Authors:  Kimberly C Blom; Sasha Farina; Yessica-Haydee Gomez; Norm R C Campbell; Brenda R Hemmelgarn; Lyne Cloutier; Donald W McKay; Martin Dawes; Sheldon W Tobe; Peter Bolli; Mark Gelfer; Donna McLean; Gillian Bartlett; Lawrence Joseph; Robin Featherstone; Ernesto L Schiffrin; Stella S Daskalopoulou
Journal:  Curr Hypertens Rep       Date:  2015-04       Impact factor: 5.369

2.  Community-based blood pressure measurement by non-health workers using electronic devices: a validation study.

Authors:  Daniel D Reidpath; Mei Lee Ling; Shajahan Yasin; Kanason Rajagobal; Pascale Allotey
Journal:  Glob Health Action       Date:  2012-06-27       Impact factor: 2.640

3.  Validation of the Omron HEM-7201 upper arm blood pressure monitor, for self-measurement in a high-altitude environment, according to the European Society of Hypertension International Protocol revision 2010.

Authors:  K Cho; M Tian; Y Lan; X Zhao; L L Yan
Journal:  J Hum Hypertens       Date:  2013-03-07       Impact factor: 3.012

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.