Diagnostic image quality of a comprehensive high-pitch dual-spiral cardiothoracic CT protocol in patients with undifferentiated acute chest pain.

Fabian Bamberg1, Roy Marcus, Wieland Sommer, Florian Schwarz, Konstantin Nikolaou, Christoph R Becker, Maximilian F Reiser, Thorsten R C Johnson.   

Abstract

OBJECTIVE: To evaluate diagnostic image quality of high-pitch dual source comprehensive cardiothoracic CT protocol in patients presenting with acute undifferentiated chest pain.
MATERIALS AND METHODS: Consecutive symptomatic subjects (n=51) with undifferentiated acute chest pain underwent ECG-synchronized high-pitch dual-spiral chest CT angiography (Definition Flash, Siemens Medical Solutions, 2 × 100 kVp or 2 × 120 kV if BMI>30, collimation: 128 × 0.6mm, pitch: 3.2). Independent investigators determined the image quality of each cardiac and pulmonary vessel segment, measured contrast-to-noise-ratio (CNR), and determined radiation exposure. In addition, the prevalence of CT findings (pulmonary embolism (PE), aortic dissection (AD) and significant coronary stenosis (≥ 50%)) was determined. Univariate and multivariate analysis were performed to determine the subpopulation with highest diagnostic quality.
RESULTS: Among 51 subjects (66% male, average age: 63 ± 15.8), the prevalence of positive CT findings was moderate (overall: 11.7%). Overall, image quality of the pulmonary, aortic and coronary vasculature was good (1.26 ± 0.43 and CNR: 2.52) with an average radiation dose of 3.82 mSv and 3.2% of segments rated non-evaluable. The image quality was lowest in the coronary arteries (p=0.02), depending on the heart rate (r=0.52, p<0.001). In subjects with a heart rate of ≤ 65 bpm (n=30) subjective image quality and CNR of the coronary arteries were higher (1.6 ± 0.5 vs. 2.1 ± 0.5, p=0.03 and 1.21 ± 0.3 vs. 1.02 ± 0.3, p=0.05) with only 1.5% segments classified as non-evaluable.
CONCLUSION: High-pitch dual-spiral comprehensive cardiothoracic CT provides low radiation exposure with excellent image quality at heart rates ≤ 65 bpm. In subjects with higher heart rates, image quality of the aortic and pulmonary vasculature remains excellent, while the assessment of the coronary arteries degrades substantially.
Copyright © 2010 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21196093     DOI: 10.1016/j.ejrad.2010.11.032

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  11 in total

1.  Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease.

Authors:  Pei Nie; Ximing Wang; Zhaoping Cheng; Xiaopeng Ji; Yanhua Duan; Jiuhong Chen
Journal:  Eur Radiol       Date:  2012-05-18       Impact factor: 5.315

Review 2.  SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT.

Authors:  Sandra S Halliburton; Suhny Abbara; Marcus Y Chen; Ralph Gentry; Mahadevappa Mahesh; Gilbert L Raff; Leslee J Shaw; Jörg Hausleiter
Journal:  J Cardiovasc Comput Tomogr       Date:  2011 Jul-Aug

Review 3.  [Triple rule-out computed tomography in emergency departments].

Authors:  D Stoevesandt; M Buerke
Journal:  Med Klin Intensivmed Notfmed       Date:  2011-10       Impact factor: 0.840

4.  Cardiac CT: A system architecture study.

Authors:  Paul FitzGerald; James Bennett; Jeffrey Carr; Peter M Edic; Daniel Entrikin; Hewei Gao; Maria Iatrou; Yannan Jin; Baodong Liu; Ge Wang; Jiao Wang; Zhye Yin; Hengyong Yu; Kai Zeng; Bruno De Man
Journal:  J Xray Sci Technol       Date:  2016       Impact factor: 1.535

Review 5.  State-of-the-art in CT hardware and scan modes for cardiovascular CT.

Authors:  Sandra Halliburton; Armin Arbab-Zadeh; Damini Dey; Andrew J Einstein; Ralph Gentry; Richard T George; Thomas Gerber; Mahadevappa Mahesh; Wm Guy Weigold
Journal:  J Cardiovasc Comput Tomogr       Date:  2012-04-07

6.  [Acute jaw-thoracic pain and syncope in a 41-year-old man].

Authors:  S Günther; H Mudra; A Reichelt; N Khaladj; C Hagl; M Pichlmaier
Journal:  Internist (Berl)       Date:  2014-10       Impact factor: 0.743

7.  High-pitch dual-source CT coronary angiography: analysis of the impact on image quality of altered electrocardiography waves during data acquisition.

Authors:  Xiaoling Wang; Jiliang Fang; Haibin Tong; Qing Zhao; Qingqiao Song; Ping Luo; Chao Xue; Min Zhang; Shuhua Yang; Qun Wang; Ping Wang; Fengxiang Shi; Lei Xu; Peijing Rong
Journal:  Int J Cardiovasc Imaging       Date:  2012-06-13       Impact factor: 2.357

8.  Energy Limits in Second Generation High-pitch Dual Source CT - Comparison in an Upper Abdominal Phantom.

Authors:  Martin Beeres; Ralf W Bauer; Josef M Kerl; Thomas J Vogl; Clara Lee
Journal:  J Clin Imaging Sci       Date:  2015-01-30

9.  Comparison of Chest Pain Protocols for Electrocardiography-Gated Dual-Source Cardiothoracic CT in Children and Adults: The Effect of Tube Current Saturation on Radiation Dose Reduction.

Authors:  Hyun Woo Goo
Journal:  Korean J Radiol       Date:  2018-01-02       Impact factor: 3.500

10.  Image Quality and Radiation Dose of High-Pitch Dual-Source Spiral Cardiothoracic Computed Tomography in Young Children with Congenital Heart Disease: Comparison of Non-Electrocardiography Synchronization and Prospective Electrocardiography Triggering.

Authors:  Hyun Woo Goo
Journal:  Korean J Radiol       Date:  2018-10-18       Impact factor: 3.500

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.