| Literature DB >> 21157527 |
R Ravichandran1, J P Binukumar, C A Davis, K Krishnamurthy, S S Sivakumar.
Abstract
There is need for simple methods for checking consistency of beam outputs and energy in linear accelerators used for radiotherapy. A method was designed by the department using perspex phantom with which the dosimetric data of two medical linear accelerators (Clinac 600 CD, Clinac 2300 CD) were evaluated over a period of 30 months. The efficacy of methods followed was checked. Routine beam consistency checks were designed for photon beams with 15 cm/ 5 cm depth ionizations in perspex phantom and variable depth combinations for electron beams. Calculated ionization ratios were compared with measured values to show their significance. The dose/MU for all radiation beams was maintained within 2% accuracy over the period of 30 months. Clinac 600 CD machine showed decreasing trend of cGy/MU, while Clinac 2300 CD showed increasing trend of cGy/MU over a period, which needed tuning of monitor chamber two times each. Tuning of output to achieve standard value was carried out once, for all electron energies when the output dose/MU exceeded 3%. During one week (June 2005), there were slight changes in electron energy detected using the ratio method, which did not recur anytime afterwards. The methods designed are adequate to find the consistency in the beam output and energies in the radiotherapy linacs.Entities:
Keywords: Calibration of linacs; perspex phantoms; quality assurance methods
Year: 2007 PMID: 21157527 PMCID: PMC3000505 DOI: 10.4103/0971-6203.35720
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Geometry of perspex phantom measurements for weekly check of photon dose output, photon beam energy (a), electron beam dose output and beam energy (b). The chamber location and thickness details are seen
Comparison of response of beam level dosimeters with reference instrument
| 839/8763 | ||||
| July 2005 | Perspex | 600CD 6MV | 1.0090 | — |
| 2300CD 6MV | 1.0020 | 1.0016 | ||
| 2300CD 15MV | 1.0020 | 1.0000 | ||
| Dec 2005 | Water | 2300CD 6MV | 1.0050 | 1.0008 |
| 854/8763 | ||||
| Mar 2006 | S.Water | 2300CD 6MV | 0.9968 | 0.9963 |
| 2300CD 15MV | 0.9967 | 0.9969 | ||
| Dec 2006 | S.Water | 600CD 6MV | 1.0050 | 1.0030 |
| 600CD 6MV | 1.0050 | 1.0022 |
Figure 2Measured percentage variations in dose output measured in perspex phantoms for a) 6 MV X-rays from Clinac 600 CD b) 6 MV X-rays from Clinac 2300 CD c) 15 MV X-rays from Clinac 2300 CD and d) 15 MeV electrons from Clinac 2300 CD
Comparison of measured and calculated ratios of ionizations at 15 cm/5 cm depths
| 6 MV | 0.5460 | 0.5731 | 0.5879 | 0.5060 | 0.5469 | −0.2 |
| 15 MV | 0.6220 | 0.6442 | 0.6526 | 0.5714 | 0.6120 | +1.6 |
Comparison of measured and calculated ratios of ionizations for electrons
| 6 MeV | 20/10 | 0.8062 | 0.8163 | −1.3 |
| 9 MeV | 30/10 | 0.8293 | 0.8511 | −2.6 |
| 12 MeV | 40/10 | 0.8632 | 0.8842 | −2.4 |
| 15 MeV | 50/10 | 0.8409 | 0.8484 | −0.9 |
| 18 MeV | 60/10 | 0.8149 | 0.8300 | −1.9 |
| 22 MeV | 70/10 | 0.7887 | 0.7939 | −0.7 |
Figure 3Measured percentage variations in measured ratios of ionizations at two reference depths in perspex phantom a) 6 MV X-rays from Clinac 600 CD b) 6 MV X-rays from Clinac 2300 CD c) 15 MV X-rays from Clinac 2300 CD and d) 15 MeV electrons from Clinac 2300 CD
Changed values of ratios in perspex for electron energies
| 6 MeV | 20/10 | 0.8078 ± 0.00154 (n=98) | 0.8368 ± 0.00573 (n=6) | <0.001 |
| 9 MeV | 30/10 | 0.8291 ± 0.00142 (n=98) | 0.8515 ± 0.00428 (n=6) | <0.001 |
| 12 MeV | 40/10 | 0.8624 ± 0.00117 (n=98) | 0.8779 ± 0.00295 (n=6) | <0.001 |
| 15 MeV | 50/10 | 0.8394 ± 0.00125 (n=98) | 0.8514 ± 0.00234 (n=6) | <0.001 |
| 18 MeV | 60/10 | 0.8136 ± 0.00095 (n=98) | 0.8228 ± 0.00018 (n=6) | <0.001 |
| 22 MeV | 70/10 | 0.7874 ± 0.00113 (n=98) | 0.7948 ± 0.00220 (n=6) | <0.001 |
Figure 4Relative dosimeter readings for various electron energies in perspex for different overlying perspex thickness above chamber. FCD 105 cm is maintained constant
Figure 5Cross plot of ratios of ionizations from measurements of Figure 4 shown against corresponding electron energies. Straight line portion of the ratio curves for different depth combinations corresponds to rapid fall of region of ionization curves of Figure 4
Comparison of electron energies after changes observed in ratios of ionizations
| 6 MeV | 2.36 | 2.99 | 5.50 | 6.16 | 2.37 | 3.09 | 5.52 | 6.36 |
| 9 MeV | 3.58 | 4.42 | 8.33 | 9.03 | 3.64 | 4.60 | 8.48 | 9.38 |
| 12 MeV | 4.92 | 6.06 | 11.46 | 12.32 | 5.02 | 6.30 | 11.70 | 12.79 |
| 15 MeV | 6.18 | 7.54 | 14.39 | 15.29 | 6.24 | 7.76 | 14.77 | 15.74 |
| 18 MeV | 7.47 | 9.22 | 17.41 | 18.69 | 7.61 | 9.42 | 17.73 | 19.09 |
| 22 MeV | 8.71 | 10.90 | 20.28 | 22.09 | 8.91 | 10.95 | 20.76 | 22.24 |