Ramamoorthy Ravichandran1, Johnson Pichy Binukumar1, Cheriyathmanjiyil Antony Davis1.
Abstract
The measured dose in water at reference point in phantom is a primary parameter for planning the treatment monitor units (MU); both in conventional and intensity modulated/image guided treatments. Traceability of dose accuracy therefore still depends mainly on the calibration factor of the ion chamber/dosimeter provided by the accredited Secondary Standard Dosimetry Laboratories (SSDLs), under International Atomic Energy Agency (IAEA) network of laboratories. The data related to Nd,water calibrations, thermoluminescent dosimetry (TLD) postal dose validation, inter-comparison of different dosimeter/electrometers, and validity of Nd,water calibrations obtained from different calibration laboratories were analyzed to find out the extent of accuracy achievable. Nd,w factors in Gray/Coulomb calibrated at IBA, GmBH, Germany showed a mean variation of about 0.2% increase per year in three Farmer chambers, in three subsequent calibrations. Another ion chamber calibrated in different accredited laboratory (PTW, Germany) showed consistent Nd,w for 9 years period. The Strontium-90 beta check source response indicated long-term stability of the ion chambers within 1% for three chambers. Results of IAEA postal TL "dose intercomparison" for three photon beams, 6 MV (two) and 15 MV (one), agreed well within our reported doses, with mean deviation of 0.03% (SD 0.87%) (n = 9). All the chamber/electrometer calibrated by a single SSDL realized absorbed doses in water within 0.13% standard deviations. However, about 1-2% differences in absorbed dose estimates observed when dosimeters calibrated from different calibration laboratories are compared in solid phantoms. Our data therefore imply that the dosimetry level maintained for clinical use of linear accelerator photon beams are within recommended levels of accuracy, and uncertainties are within reported values.Entities:
Keywords: Absorbed dose; calibration factor for reference beam; dose delivery; uncertainties in dosimetry
Year: 2013 PMID: 24672156 PMCID: PMC3959001 DOI: 10.4103/0971-6203.121199
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Calibration factors Nd,w for thimble (FC 65) chambers from different laboratories
Strontium-90 check source, chamber response for reading ∫500 mGy
Absorbed dose estimates by IAEA postal TLD intercomparison
Dose estimate in cylindrical perspex phantom with Nd,w
Corrected dose value from electrometers on Gy display for different ion chambers
Estimate of absorbed doses with dosimeters traceable to different laboratories