Literature DB >> 21152894

Fluid leakage across tracheal tube cuff, effect of different cuff material, shape, and positive expiratory pressure: a bench-top study.

Alberto Zanella1, Vittorio Scaravilli, Stefano Isgrò, Manuela Milan, Massimo Cressoni, Nicolò Patroniti, Roberto Fumagalli, Antonio Pesenti.   

Abstract

PURPOSE: Standard polyvinylchloride (PVC) endotracheal tube (ETT) cuffs do not protect from aspiration across the cuff, a leading cause of ventilator-associated pneumonia (VAP). In a long-lasting in vitro study we compared the effect of different cuff materials (PVC, polyurethane, and guayule latex), shapes (cylindrical, conical), and positive end expiratory pressures (PEEP) in reducing fluid leakage across the cuff.
METHODS: We compared fluid leakage across a cylindrical double-layer guayule latex prototype cuff, three cylindrical PVC cuffs (Mallinckrodt Hi-Lo, Mallinckrodt HighContour, Portex Ivory), one conical PVC cuff (Mallinckrodt TaperGuard), and two polyurethane cuffs (Mallinckrodt SealGuard, conical; Microcuff, cylindrical). Ten centimeters of dyed water was poured above the cuffs inflated (pressure 30 cmH(2)O) in a vertical cylinder (diameter 20 mm). A respiratory circuit connected the bottom of the cylinder to a breathing bag inflated at four pressures (PEEP = 0, 5, 10, 15 cmH(2)O). Pictures were taken every 60 s for 24 h to measure leakage as a reduction in the water column above the cuff. Five new ETTs of each type were tested.
RESULTS: The guayule latex cuffs showed no leakage at all the PEEP levels. Both the cylindrical and conical polyurethane cuffs showed limited leakage (2.1 ± 1.8 cm of water) only for PEEP zero. The PVC cuffs showed reduced leakage with increasing PEEP: 8.4 ± 1.5, 7.8 ± 2.2, 2.2 ± 1.0, and 0 cm of water at 0, 5, 10, and 15 cmH(2)O, respectively. Among all the PVC cuffs, the conical shape ensured higher sealing properties.
CONCLUSIONS: The guayule latex cuffs always prevented fluid leakage; the polyurethane and PVC cuffs required incremental levels of PEEP to prevent fluid leakage ever-present at zero PEEP.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21152894     DOI: 10.1007/s00134-010-2106-z

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  9 in total

1.  A double-layer tracheal tube cuff designed to prevent leakage: a bench-top study.

Authors:  Alberto Zanella; Massimo Cressoni; Myra Epp; Mario Stylianou; Theodor Kolobow
Journal:  Intensive Care Med       Date:  2008-02-08       Impact factor: 17.440

Review 2.  Ventilator-associated pneumonia.

Authors:  Mauricio Valencia; Antoni Torres
Journal:  Curr Opin Crit Care       Date:  2009-02       Impact factor: 3.687

3.  Tracheal fluid leakage in benchtop trials: comparison of static versus dynamic ventilation model with and without lubrication.

Authors:  Mital H Dave; Nelly Koepfer; Caveh Madjdpour; Angela Frotzler; Markus Weiss
Journal:  J Anesth       Date:  2010-02-23       Impact factor: 2.078

4.  Leakage of fluid past the tracheal tube cuff in a benchtop model.

Authors:  P J Young; M Rollinson; G Downward; S Henderson
Journal:  Br J Anaesth       Date:  1997-05       Impact factor: 9.166

5.  Aspiration beyond endotracheal cuffs.

Authors:  R D Seegobin; G L van Hasselt
Journal:  Can Anaesth Soc J       Date:  1986-05

6.  The evolution of tracheal injury due to ventilatory assistance through cuffed tubes: a pathologic study.

Authors:  J D Cooper; H C Grillo
Journal:  Ann Surg       Date:  1969-03       Impact factor: 12.969

7.  Does PEEP reduce the incidence of aspiration around endotracheal tubes?

Authors:  B A Janson; T J Poulton
Journal:  Can Anaesth Soc J       Date:  1986-03

8.  Positive-end expiratory pressure reduces incidence of ventilator-associated pneumonia in nonhypoxemic patients.

Authors:  Francisco Manzano; Enrique Fernández-Mondéjar; Manuel Colmenero; María Eugenia Poyatos; Ricardo Rivera; Juan Machado; Iñaki Catalán; Antonio Artigas
Journal:  Crit Care Med       Date:  2008-08       Impact factor: 7.598

9.  Effect of positive expiratory pressure and type of tracheal cuff on the incidence of aspiration in mechanically ventilated patients in an intensive care unit.

Authors:  Umberto Lucangelo; Walter A Zin; Vittorio Antonaglia; Lara Petrucci; Marino Viviani; Giovanni Buscema; Massimo Borelli; Giorgio Berlot
Journal:  Crit Care Med       Date:  2008-02       Impact factor: 7.598

  9 in total
  31 in total

1.  Endotracheal tube cuff--small important part of a big issue.

Authors:  Shai Efrati; Israel Deutsch; Gabriel M Gurman
Journal:  J Clin Monit Comput       Date:  2012-01-03       Impact factor: 2.502

2.  Randomized Pilot Trial of Two Modified Endotracheal Tubes To Prevent Ventilator-associated Pneumonia.

Authors:  Steven Deem; David Yanez; Laura Sissons-Ross; Jo Ann Elrod Broeckel; Stephen Daniel; Miriam Treggiari
Journal:  Ann Am Thorac Soc       Date:  2016-01

3.  Comparison of fluid leakage across endotracheal tube cuffs using a three-dimensional printed model of the human trachea.

Authors:  Tomohiko Kimijima; Mitsutaka Edanaga; Michiaki Yamakage
Journal:  J Anesth       Date:  2016-01-27       Impact factor: 2.078

4.  Condensation of humidified air in the inflation line of a polyurethane cuff precludes correct continuous pressure monitoring during mechanical ventilation.

Authors:  Herbert Spapen; Walter Moeyersons; Wim Stiers; Geert Desmet; Emiel Suys
Journal:  J Anesth       Date:  2014-05-23       Impact factor: 2.078

5.  Beware the siren's song of novel endotracheal tube designs.

Authors:  Michael Klompas; Lorenzo Berra; Richard Branson
Journal:  Intensive Care Med       Date:  2017-03-31       Impact factor: 17.440

Review 6.  Novel preventive strategies for ventilator-associated pneumonia.

Authors:  Andrea Coppadoro; Edward Bittner; Lorenzo Berra
Journal:  Crit Care       Date:  2012-12-12       Impact factor: 9.097

7.  Pathogenic Link Between Postextubation Pneumonia and Ventilator-Associated Pneumonia: An Experimental Study.

Authors:  Emanuele Rezoagli; Alberto Zanella; Massimo Cressoni; Lorenzo De Marchi; Theodor Kolobow; Lorenzo Berra
Journal:  Anesth Analg       Date:  2017-04       Impact factor: 5.108

8.  Prevention of Lung Bacterial Colonization With a Leak-Proof Endotracheal Tube Cuff: An Experimental Animal Study.

Authors:  Emanuele Rezoagli; Massimo Cressoni; Giacomo Bellani; Giacomo Grasselli; Antonio M Pesenti; Theodor Kolobow; Alberto Zanella
Journal:  Respir Care       Date:  2019-04-23       Impact factor: 2.258

9.  The lowest effective intracuff pressure of the esophagus obstruction tube to prevent reflux of gastric contents: a study on rabbits.

Authors:  LinLi Luo; Huafeng Li; LiangXue Zhou; Jun Zhou; Juan Ni; Rong Zhou
Journal:  J Anesth       Date:  2013-12-04       Impact factor: 2.078

10.  Orotracheal tube as a risk factor for lower respiratory tract infection: preliminary data from a randomised trial.

Authors:  Igor Muzlovic; Janja Perme; David Stubljar
Journal:  Wien Klin Wochenschr       Date:  2018-01-16       Impact factor: 1.704

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.