A double-layer tracheal tube cuff designed to prevent leakage: a bench-top study.

Alberto Zanella1, Massimo Cressoni, Myra Epp, Mario Stylianou, Theodor Kolobow.   

Abstract

OBJECTIVE: We designed a new endotracheal tube (ETT) cuff that does not form the folds that cause leakage of colonized subglottic secretions upon inflation within the trachea: a standard high-volume low-pressure cuff was draped with a second, highly elastic cuff made of a low-protein guayule natural latex rubber with 0.5 ml gel between the cuffs. We compared this prototype ETT cuff with four commercially available ETTs for efficacy in the prevention of fluid leakage across the cuff.
DESIGN: In vitro study. MEASUREMENTS AND
RESULTS: We compared fluid leakage in our prototype cuff with that in four commercially available ETTs. Three cylindrical glass tubes 16, 20, and 22 mm in diameter were used as model tracheas, and five different intracuff pressures (20, 25, 30, 40, and 50 cmH(2)O) were tested. Each test was repeated three times with new ETTs. The guayule latex ETT cuff showed an average fluid leakage of 0.0007 +/- 0.002 ml/min which was significantly lower than that in any of the other ETTs (Microcuff 0.07 +/- 0.09, Mallinckrodt/Hi-Lo Evac 5 +/- 5, Euromedical 7 +/- 4, Sheridan/CF 41 +/- 69).
CONCLUSIONS: Our prototype ETT cuff significantly reduced fluid leakage in this bench-top study. In vivo testing and evaluation is to follow.

Entities:  

Mesh:

Year:  2008        PMID: 18259729     DOI: 10.1007/s00134-008-1016-9

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  12 in total

1.  Inflation of a pressure-limited cuff inside a model trachea.

Authors:  P J Young; W H Young
Journal:  Med Eng Phys       Date:  2003-07       Impact factor: 2.242

2.  Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube.

Authors:  Alexander Dullenkopf; Andreas Gerber; Markus Weiss
Journal:  Intensive Care Med       Date:  2003-08-16       Impact factor: 17.440

3.  Absence of cross-reactivity of IgE antibodies from subjects allergic to Hevea brasiliensis latex with a new source of natural rubber latex from guayule (Parthenium argentatum).

Authors:  D J Siler; K Cornish; R G Hamilton
Journal:  J Allergy Clin Immunol       Date:  1996-11       Impact factor: 10.793

4.  High-volume, low-pressure cuffs. Are they always low pressure?

Authors:  D Guyton; M J Banner; R R Kirby
Journal:  Chest       Date:  1991-10       Impact factor: 9.410

5.  Pneumonia in intubated patients: role of respiratory airway care.

Authors:  J Rello; R Soñora; P Jubert; A Artigas; M Rué; J Vallés
Journal:  Am J Respir Crit Care Med       Date:  1996-07       Impact factor: 21.405

6.  Aspiration pneumonia; incidence of aspiration with endotracheal tubes.

Authors:  S B Spray; G D Zuidema; J L Cameron
Journal:  Am J Surg       Date:  1976-06       Impact factor: 2.565

7.  The evolution of tracheal injury due to ventilatory assistance through cuffed tubes: a pathologic study.

Authors:  J D Cooper; H C Grillo
Journal:  Ann Surg       Date:  1969-03       Impact factor: 12.969

8.  A low-volume, low-pressure tracheal tube cuff reduces pulmonary aspiration.

Authors:  Peter J Young; Sivarajasingham Pakeerathan; Mark C Blunt; Sriharsha Subramanya
Journal:  Crit Care Med       Date:  2006-03       Impact factor: 7.598

9.  Viral impermeability of hypoallergenic, low protein, guayule latex films.

Authors:  K Cornish; C D Lytle
Journal:  J Biomed Mater Res       Date:  1999-12-05

Review 10.  The pathogenesis of ventilator-associated pneumonia: its relevance to developing effective strategies for prevention.

Authors:  Nasia Safdar; Christopher J Crnich; Dennis G Maki
Journal:  Respir Care       Date:  2005-06       Impact factor: 2.258

View more
  14 in total

1.  Endotracheal tube cuff--small important part of a big issue.

Authors:  Shai Efrati; Israel Deutsch; Gabriel M Gurman
Journal:  J Clin Monit Comput       Date:  2012-01-03       Impact factor: 2.502

2.  Variables affecting leakage past endotracheal tube cuffs: a bench study.

Authors:  Renée Pitts; Daniel Fisher; Demet Sulemanji; Joseph Kratohvil; Yandong Jiang; Robert Kacmarek
Journal:  Intensive Care Med       Date:  2010-09-18       Impact factor: 17.440

3.  Fluid leakage across tracheal tube cuff, effect of different cuff material, shape, and positive expiratory pressure: a bench-top study.

Authors:  Alberto Zanella; Vittorio Scaravilli; Stefano Isgrò; Manuela Milan; Massimo Cressoni; Nicolò Patroniti; Roberto Fumagalli; Antonio Pesenti
Journal:  Intensive Care Med       Date:  2010-12-09       Impact factor: 17.440

4.  Tracheal fluid leakage in benchtop trials: comparison of static versus dynamic ventilation model with and without lubrication.

Authors:  Mital H Dave; Nelly Koepfer; Caveh Madjdpour; Angela Frotzler; Markus Weiss
Journal:  J Anesth       Date:  2010-02-23       Impact factor: 2.078

5.  Beware the siren's song of novel endotracheal tube designs.

Authors:  Michael Klompas; Lorenzo Berra; Richard Branson
Journal:  Intensive Care Med       Date:  2017-03-31       Impact factor: 17.440

6.  Effects of tracheal orientation on development of ventilator-associated pneumonia: an experimental study.

Authors:  Alberto Zanella; Massimo Cressoni; Myra Epp; Viktoria Hoffmann; Mario Stylianou; Theodor Kolobow
Journal:  Intensive Care Med       Date:  2012-02-18       Impact factor: 17.440

7.  A polyurethane cuffed endotracheal tube is associated with decreased rates of ventilator-associated pneumonia.

Authors:  Melissa A Miller; Jennifer L Arndt; Mark A Konkle; Carol E Chenoweth; Theodore J Iwashyna; Kevin R Flaherty; Robert C Hyzy
Journal:  J Crit Care       Date:  2010-07-23       Impact factor: 3.425

8.  Pathogenic Link Between Postextubation Pneumonia and Ventilator-Associated Pneumonia: An Experimental Study.

Authors:  Emanuele Rezoagli; Alberto Zanella; Massimo Cressoni; Lorenzo De Marchi; Theodor Kolobow; Lorenzo Berra
Journal:  Anesth Analg       Date:  2017-04       Impact factor: 5.108

9.  Prevention of Lung Bacterial Colonization With a Leak-Proof Endotracheal Tube Cuff: An Experimental Animal Study.

Authors:  Emanuele Rezoagli; Massimo Cressoni; Giacomo Bellani; Giacomo Grasselli; Antonio M Pesenti; Theodor Kolobow; Alberto Zanella
Journal:  Respir Care       Date:  2019-04-23       Impact factor: 2.258

10.  Orotracheal tube as a risk factor for lower respiratory tract infection: preliminary data from a randomised trial.

Authors:  Igor Muzlovic; Janja Perme; David Stubljar
Journal:  Wien Klin Wochenschr       Date:  2018-01-16       Impact factor: 1.704

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.