Literature DB >> 21150659

Accuracy and safety assessment of pedicle screw placement using the rapid prototyping technique in severe congenital scoliosis.

Zi-Xiang Wu1, Lu-Yu Huang, Hong-Xun Sang, Zhen-Sheng Ma, Shi-Yong Wan, Geng Cui, Wei Lei.   

Abstract

STUDY
DESIGN: The application of rapid prototyping (RP) technique for improving accuracy of pedicle screw placement in congenital scoliosis is described in this study.
OBJECTIVE: To compare the accuracy and safety of pedicle screw placement in congenital scoliosis using the RP technique versus the conventional fluoroscopy. SUMMARY OF BACKGROUND DATA: Maldeveloped vertebral components in congenital scoliosis leads to prolonged operation time and higher rate of screw misplacement. RP technique can enhance preoperative and perioperative planning. No data are available on the accuracy of pedicle screw fixation using the RP technique.
METHODS: Sixty-two consecutive patients with hemivertebra had undergone posterior-only hemivertebra resection. Pedicle screws were implanted either by the conventional intraoperative fluoroscopy technique (C-arm group; n=28) or the RP technique (RP group; n=34). Accuracy of pedicle screw placement was compared by postoperative computed tomographic scan.
RESULTS: Seventy of 677 inserted screws were found to be misplaced, showing an overall accuracy of 89.7% (90.8% in the thoracic spine and 87.4% in the lumbar spine). In the C-arm group, 86.1% (167 of 194) and 82.0% (82 of 100) of screws were accurately placed in the thoracic and lumbar spine, respectively. While in the RP group, the respective screw placement accuracies were 94.4% (238 of 252) and 91.6% (120 of 131). In the C-arm and the RP groups, 94.8% (279 of 294) and 97.9% (375 of 383) of the screws were within the safety zone, respectively. Compared with the fluoroscopy method, the RP-assisted technique showed a shorter operation time and higher scoliosis correction rate. No neurovascular-related complication was observed with this technique during the study.
CONCLUSION: The application of RP technique in congenital scoliosis can reduce the operation time, the risk of screw misplacement and its consequent complications. The use of RP technique in congenital scoliosis is safe and efficacious.

Entities:  

Mesh:

Year:  2011        PMID: 21150659     DOI: 10.1097/BSD.0b013e318201be2a

Source DB:  PubMed          Journal:  J Spinal Disord Tech        ISSN: 1536-0652


  29 in total

1.  Are computer numerical control (CNC)-manufactured patient-specific metal templates available for posterior thoracic pedicle screw insertion? Feasibility and accuracy evaluation.

Authors:  Xiangxue Kong; Lei Tang; Qiang Ye; Wenhua Huang; Jianyi Li
Journal:  Eur Spine J       Date:  2017-07-17       Impact factor: 3.134

Review 2.  The role of radiography in the study of spinal disorders.

Authors:  Fernando Ruiz Santiago; Antonio Jesús Láinez Ramos-Bossini; Yì Xiáng J Wáng; Daniel López Zúñiga
Journal:  Quant Imaging Med Surg       Date:  2020-12

3.  Design of mulitlevel OLF approach ("V"-shaped decompressive laminoplasty) based on 3D printing technology.

Authors:  Qinjie Ling; Erxing He; Hanbin Ouyang; Jing Guo; Zhixun Yin; Wenhua Huang
Journal:  Eur Spine J       Date:  2017-07-27       Impact factor: 3.134

4.  Application of a novel 3D drill template for cervical pedicle screw tunnel design: a cadaveric study.

Authors:  Zhengxi Yu; Guodong Zhang; Xuanhuang Chen; Xu Chen; Changfu Wu; Yijun Lin; Wenhua Huang; Haibin Lin
Journal:  Eur Spine J       Date:  2017-06-10       Impact factor: 3.134

5.  Rapid-prototype endoprosthesis for palliative reconstruction of an upper extremity after resection of bone metastasis.

Authors:  Dumnoensun Pruksakorn; Nattapon Chantarapanich; Olarn Arpornchayanon; Taninit Leerapun; Kriskrai Sitthiseripratip; Natapoom Vatanapatimakul
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-05-20       Impact factor: 2.924

Review 6.  Methods to determine pedicle screw placement accuracy in spine surgery: a systematic review.

Authors:  Ahmed A Aoude; Maryse Fortin; Rainer Figueiredo; Peter Jarzem; Jean Ouellet; Michael H Weber
Journal:  Eur Spine J       Date:  2015-03-07       Impact factor: 3.134

Review 7.  Systematic review of 3D printing in spinal surgery: the current state of play.

Authors:  Ben Wilcox; Ralph J Mobbs; Ai-Min Wu; Kevin Phan
Journal:  J Spine Surg       Date:  2017-09

8.  Accurate placement of cervical pedicle screws using 3D-printed navigational templates : An improved technique with continuous image registration.

Authors:  Guodong Zhang; Zhengxi Yu; Xuanhuang Chen; Xu Chen; Changfu Wu; Yijun Lin; Wenhua Huang; Haibin Lin
Journal:  Orthopade       Date:  2018-05       Impact factor: 1.087

9.  Lumbar and sacral pedicle screw placement using a template does not improve the midterm pain and disability outcome in comparison with free-hand method.

Authors:  Matjaz Merc; Gregor Recnik; Zmago Krajnc
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-01-19

10.  CLINICAL APPLICATION OF A DRILL GUIDE TEMPLATE FOR PEDICLE SCREW PLACEMENT IN SEVERE SCOLIOSIS.

Authors:  Xin Li; Yaoshen Zhang; Qiang Zhang; Changsong Zhao; Kun Liu
Journal:  Acta Ortop Bras       Date:  2017 Mar-Apr       Impact factor: 0.513

View more