Lyda Osorio1, Meleny Ramirez, Anilza Bonelo, Luis A Villar, Beatriz Parra.
Abstract
BACKGROUND: We compared the diagnostic accuracy and reproducibility of commercially available NS1-based dengue tests and explored factors influencing their sensitivities.Entities:
Mesh:
Substances:
Year: 2010 PMID: 21134275 PMCID: PMC3016282 DOI: 10.1186/1743-422X-7-361
Source DB: PubMed Journal: Virol J ISSN: 1743-422X Impact factor: 4.099
Reported sensitivity and specificity of commercially available NS1-based dengue diagnostic tests
| Test | Sensitivity % | Specificity % | place |
|---|---|---|---|
| STRIP™ | 61.6 (55.2-67.8) | 100 (93.8-100) | Vietnam [ |
| PanE™ | 49.4 (38.5-60.4) | 100 (92.1-100) | India [ |
| STRIP™ | 78.9 (70-86.1) | 99 (94.6-99.9) | Singapore [ |
| PanE™ | 67 (57.3-75.7) | 100 (96.4-100) | Singapore [ |
| Platelia™ | 81.7 (73.1-88.4) | 100 (96.4-100) | Singapore [ |
| STRIP™ | 89.6 (84.7-93.2) | 99.1 (96.9-99.9) | Brazil [ |
| PanE™ | 72.3 (65.8-78.1) | 100 (98.4-100) | Brazil [ |
| Platelia™ | 83.6 (78.1-88.2) | 98.7 (96.2-99.7) | Brazil [ |
| Platelia™ | 67.3 (57.1-76.4) | - | Finland [ |
| Platelia™ | 37 | 99.5 | Vietnam [ |
| Platelia™ | 73.6 (63.7-81.6 | - | Australia [ |
| PanE™ | 63.7 (53.5-72). | - | Australia [ |
| Platelia™ | 83.2 (75.5-89.3) | 100 (86.7-100.0) | Vietnam [ |
| Platelia™ | 71.3 (61-80) | 86.1 (70.9-94.4) | Venezuela [ |
| Pan E™ | 60.9 (50.4-70.5) | 94.4 (80.9-99.4) | Venezuela [ |
| STRIP™ | 67.8 (57.4-76.7) | 94.4 (80.9-99.4) | Venezuela [ |
| STRIP™ | 90.4 (86.6-94.4) | 99.5 (97.4-99.9) | Malaysia [ |
| STRIP™ | 98.9 (96.8-100) | 90.6 (85.6-95.7) | Thailand [ |
| STRIP™ | 77.3 (0.54-0.92) | 100 | Taiwan [ |
| Pan E™ | 83.3 (65.2-94.3) | - | China [ |
| Platelia™ | 63.2 (55.7-70.0) | 98.4 (91.7-99.7) | Thailand [ |
| Platelia™ | 83.2 (77.5-87.7) | 100 (92.1-100) | Puerto Rico [ |
| PanE™ | 64.9 (58.2-71.1) | 97.8 (88.4-99.6) | Puerto Rico [ |
| STRIP™ | 77.6 (72.1-82.4) | 100 (92.6-100) | French Guiana [ |
| Platelia™ | 82.4 (77.3-86.7) | 100 (92.6-100) | French Guiana [ |
| Pan E™ | 55.1 (49.0-61.2) | 97.9 (88.9-99.9) | French Guiana [ |
| Platelia™ | 92.3 (64-99.8) | 100 | Thailand [ |
| PanE™ | 63 (53-73) | 100 | Laos [ |
| Platelia™ | 93.4 (89.2-96.3) | 100 (98.9-100) | Singapore [ |
| Platelia™ | 88.7 (94-92.4) | 100 (98.9-100) | French Guiana [ |
Figure 1Study sample selection and laboratory analyses.
Characteristics of the evaluated NS1-based dengue diagnostic tests
| Test name | Platelia™ Dengue | Pan-E™ | SD Dengue | Dengue NS1 | SD BIOLINE™ |
|---|---|---|---|---|---|
| Bio-Rad | Inverness | Standard diagnostic Inc | Bio-Rad | Standard diagnostic Inc | |
| France | Australia | South Korea | France | South Korea | |
| ELISA | ELISA | ELISA | Immunochromatography | Immunochromatography | |
| 96 | 96 | 96 | 25 | 1 (of each NS1, IgM and IgG) | |
| 50 μL | 60-75 μL | 50 μL | 50 μL | 85 μL (75 for NS1 plus 10 for IgM/IgG) | |
| 2.5 h | 2.5 h | 2.5 h | 15-30 min | 20 min | |
| Yes | Yes | Yes | No | No |
Measures of performance of commercially available NS1 detection-based dengue diagnostic assays
| Assay | Dengue | Non-dengue | Positive | Negative | Sensitivity % | Specificity % | PPV % | NPV % | LR+ | LR- |
|---|---|---|---|---|---|---|---|---|---|---|
| 212 | 91 | 150 | 84 | 70.8 | 92.3 | 95.5 | 57.5 | 9.2 | 0.31 | |
| (64.1-76.8) | (84.8-96.9) | (91-98.2) | (49.1-65.7) | (4.5-18.8) | (0.25-0.39) | |||||
| 218 | 92 | 155 | 82 | 71.1 | 89.1 | 94 | 56.6 | 6.54 | 0.32 | |
| (64.6-77) | (80.9-94.7) | (89.1-97.1) | (48.1-64.8) | (3.62-11.8) | (0.26-0.4) | |||||
| 218 | 92 | 150 | 87 | 68.8 | 94.6 | 96.8 | 56.1 | 12.7 | 0.33 | |
| (62.2-75) | (87.8-98.2) | (92.6-99) | (48-64.1) | (5.37-29.8) | (0.27-0.4) | |||||
| 218 | 92 | 111 | 89 | 51 | 96.7 | 97.4 | 45.4 | 15.6 | 0.5 | |
| (44.1-57.7) | (90.8-99.3) | (92.5-99.5) | (38.3-52.7) | (5.1-48) | (0.44-0.58) | |||||
| 218 | 92 | 171 | 84 | 78.4 | 91.3 | 95.5 | 64.1 | 9.02 | 0.23 | |
| (72.4-83.7) | (83.6-96.2) | (91.4-98.1) | (55.3-72.3) | (4.64-17.6) | (0.18-0.3) | |||||
| 218 | 92 | 176 | 82 | 80.7 | 89.1 | 94.6 | 66.1 | 7.43 | 0.21 | |
| (75-85.7) | (81-94.7) | (90.3-97.4) | (57.1-74.4) | (4.12-13.8) | (0.16-0.28) | |||||
| 104 | 43 | 60 | 41 | 57.7 | 95.3 | 96.8 | 48.2 | 12.4 | 0.44 | |
| (47.6-67.3) | (84.2-99.4) | (88.8-99.6) | (37.3-59.3) | (3.17-48.5) | (0.35-0.56) | |||||
| 104 | 43 | 64 | 41 | 61.5 | 93.3 | 97 | 50.6 | 13.2 | 0.4 | |
| (51.5-70.9) | (84.2-99.4) | (89.5-99.6) | (39.3-62) | (3.4-51.6) | (0.31-0.5) |
PPV = Positive Predictive Value, NPV = Negative Predictive Value, LR = Likelihood Ratio
Figure 2Sensitivity (with corresponding 95%CI) of NS1-based dengue diagnostic tests according to days of fever onset (A), type of infection (B), viral serotype (C), and (D) severity of disease.
Sensitivity of SD Bioline™ Dengue Duo according to presence of IgG and days of fever.
| Component of SD Bioline™ Dengue Duo analyzed | Negative IgG in acute sera | Positive IgG in acute sera | |||
|---|---|---|---|---|---|
| Days of fever onset | 0 to 3 | 4 to 7 | 0 to 3 | 4 to 7 | |
| Dengue samples | N = 108 | N = 46 | N = 14 | N = 50 | |
| Only NS1 | Positive tests | 71 | 30 | 2 | 8 |
| Sensitivity % | 65.7 | 65.2 | 14.3 | 16 | |
| (95% CI) | (56-74.6) | (49.8-78.6) | (1.78-42.8) | (7.17-29.1) | |
| Any of NS1/IgM | Positive tests | 78 | 34 | 13 | 46 |
| Sensitivity % | 72.2 | 74 | 93 | 92 | |
| (95% CI) | (62.8-80.4) | (59-85.7) | (66.1-99.8) | (80.8-97.8) | |
| Any of NS1/IgM/IgG | Positive tests | 78 | 34 | 14 | 50 |
| Sensitivity % | 72.2 | 74 | 100 | 100 | |
| (95% CI) | (62.8-80.4) | (59-85.7) | (76.8-100) | (92.9-100) | |