Literature DB >> 17142419

Treatment of unstable distal radial fractures with the volar locking plating system.

Kevin C Chung1, Andrew J Watt, Sandra V Kotsis, Zvi Margaliot, Steven C Haase, H Myra Kim.   

Abstract

BACKGROUND: The best treatment for an inadequately reduced fracture of the distal part of the radius is not well established. We collected prospective outcomes data for patients undergoing open reduction and internal fixation of an inadequately reduced distal radial fracture with use of the volar locking plating system.
METHODS: Over a two-year period, 161 patients underwent open reduction and internal fixation of an inadequately reduced distal radial fracture with use of the volar locking plating system. Patients were enrolled in the present study three months after the fracture on the basis of strict entry criteria and were evaluated three, six, and twelve months after surgery. Outcome measures included radiographic parameters, grip strength, lateral pinch strength, the Jebsen-Taylor test, wrist range of motion, and the Michigan Hand Outcomes Questionnaire.
RESULTS: Eighty-seven patients with a distal radial fracture were enrolled. The mean age at the time of enrollment was 48.9 years. Forty percent (thirty-five) of the eighty-seven fractures were classified as AO type A, 9% (eight) were classified as type B, and 51% (forty-four) were classified as type C. Radiographic assessment showed that the plating system maintained anatomic reduction at the follow-up periods. At the time of the twelve-month follow-up, the mean grip strength on the injured side was worse than that on the contralateral side (18 compared with 21 kg; p<0.01), the mean pinch strength on the injured side was not significantly different from that on the contralateral side (8.7 compared with 8.9 kg; p=0.27), and the mean flexion of the wrist on the injured side was 86% of that on the contralateral side. All Michigan Hand Outcomes Questionnaire domains approached normal scores at six months, with small continued improvement to one year.
CONCLUSIONS: The volar locking plating system appears to provide effective fixation when used for the treatment of initially inadequately reduced distal radial fractures.

Entities:  

Mesh:

Year:  2006        PMID: 17142419     DOI: 10.2106/JBJS.E.01298

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  58 in total

1.  The cadaveric anatomy of the distal radius: implications for the use of volar plates.

Authors:  P A McCann; D Clarke; R Amirfeyz; R Bhatia
Journal:  Ann R Coll Surg Engl       Date:  2012-03       Impact factor: 1.891

2.  Clinical research in hand surgery.

Authors:  Kevin C Chung
Journal:  J Hand Surg Am       Date:  2010-01       Impact factor: 2.230

Review 3.  Systematic review of outcomes following fixed angle intramedullary fixation of distal radius fractures.

Authors:  John Hardman; Nawfal Al-Hadithy; Thomas Hester; Raymond Anakwe
Journal:  Int Orthop       Date:  2015-04-15       Impact factor: 3.075

4.  The Cost-Effectiveness of Surgical Fixation of Distal Radial Fractures: A Computer Model-Based Evaluation of Three Operative Modalities.

Authors:  Prashant V Rajan; Rameez A Qudsi; George S M Dyer; Elena Losina
Journal:  J Bone Joint Surg Am       Date:  2018-02-07       Impact factor: 5.284

5.  Relationship between patient satisfaction and objective functional outcome after surgical treatment for distal radius fractures.

Authors:  Kevin C Chung; Ann Haas
Journal:  J Hand Ther       Date:  2009-06-26       Impact factor: 1.950

6.  Measuring quality in health care and its implications for pay-for-performance initiatives.

Authors:  Kevin C Chung; Melissa J Shauver
Journal:  Hand Clin       Date:  2009-02       Impact factor: 1.907

7.  Validity and responsiveness of the Jebsen-Taylor Hand Function Test.

Authors:  Erika Davis Sears; Kevin C Chung
Journal:  J Hand Surg Am       Date:  2009-12-02       Impact factor: 2.230

Review 8.  Considerations in the Treatment of Osteoporotic Distal Radius Fractures in Elderly Patients.

Authors:  Peter J Ostergaard; Matthew J Hall; Tamara D Rozental
Journal:  Curr Rev Musculoskelet Med       Date:  2019-03

Review 9.  Measuring quality of surgical care: is it attainable?

Authors:  Kevin C Chung; Rod J Rohrich
Journal:  Plast Reconstr Surg       Date:  2009-02       Impact factor: 4.730

10.  External fixation of fractures of the distal radius: a randomized comparison of the Hoffman compact II non-bridging fixator and the Dynawrist fixator in 75 patients followed for 1 year.

Authors:  Yngvar Krukhaug; Stein Ugland; Stein A Lie; Leiv M Hove
Journal:  Acta Orthop       Date:  2009-02       Impact factor: 3.717

View more