Literature DB >> 21116997

Standard practice for cell sorting in a BSL-3 facility.

Stephen P Perfetto1, David R Ambrozak, Richard Nguyen, Mario Roederer, Richard A Koup, Kevin L Holmes.   

Abstract

Over the past decade, there has been a rapid growth in the number of BSL-3 and BSL-4 laboratories in the USA and an increase in demand for infectious cell sorting in BSL-3 laboratories. In 2007, the International Society for Advancement of Cytometry (ISAC) Biosafety Committee published standards for the sorting of unfixed cells and is an important resource for biosafety procedures when performing infectious cell sorting. Following a careful risk assessment, if it is determined that a cell sorter must be located within a BSL-3 laboratory, there are a variety of factors to be considered prior to the establishment of the laboratory. This chapter outlines procedures for infectious cell sorting in a BSL-3 environment to facilitate the establishment and safe operation of a BSL-3 cell sorting laboratory. Subjects covered include containment verification, remote operation, disinfection, personal protective equipment (PPE), and instrument-specific modifications for enhanced aerosol evacuation.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21116997      PMCID: PMC4789760          DOI: 10.1007/978-1-61737-950-5_22

Source DB:  PubMed          Journal:  Methods Mol Biol        ISSN: 1064-3745


  5 in total

Review 1.  Biosafety considerations for flow cytometric analysis of human immunodeficiency virus-infected samples.

Authors:  I Schmid; A Kunkl; J K Nicholson
Journal:  Cytometry       Date:  1999-10-15

Review 2.  Biosafety concerns for shared flow cytometry core facilities.

Authors:  Ingrid Schmid; Steven Merlin; Stephen P Perfetto
Journal:  Cytometry A       Date:  2003-12       Impact factor: 4.355

3.  International Society for Analytical Cytology biosafety standard for sorting of unfixed cells.

Authors:  Ingrid Schmid; Claude Lambert; David Ambrozak; Gerald E Marti; Delynn M Moss; Stephen P Perfetto
Journal:  Cytometry A       Date:  2007-06       Impact factor: 4.355

4.  Introduction to the biosafety guidelines for sorting of unfixed cells.

Authors:  I Schmid; P N Dean
Journal:  Cytometry       Date:  1997-06-01

5.  Biosafety guidelines for sorting of unfixed cells.

Authors:  I Schmid; J K Nicholson; J V Giorgi; G Janossy; A Kunkl; P A Lopez; S Perfetto; L C Seamer; P N Dean
Journal:  Cytometry       Date:  1997-06-01
  5 in total
  2 in total

1.  Novel Impactor and Microsphere-Based Assay Used to Measure Containment of Aerosols Generated in a Flow Cytometer Cell Sorter.

Authors:  Stephen P Perfetto; Phillip J Hogarth; Simon Monard; Ben Fontes; Kristen M Reifel; Brandon K Swan; Jan Baijer; Evan R Jellison; Geoffrey Lyon; Patty Lovelace; Richard Nguyen; David Ambrozak; Kevin L Holmes
Journal:  Cytometry A       Date:  2018-12-18       Impact factor: 4.355

2.  International Society for the Advancement of Cytometry cell sorter biosafety standards.

Authors:  Kevin L Holmes; Benjamin Fontes; Philip Hogarth; Richard Konz; Simon Monard; Charles H Pletcher; Robert B Wadley; Ingrid Schmid; Stephen P Perfetto
Journal:  Cytometry A       Date:  2014-03-13       Impact factor: 4.355

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.